| Literature DB >> 35950259 |
Mark W Geurts1, Dustin J Jacqmin2, Lindsay E Jones3, Stephen F Kry4, Dimitris N Mihailidis5, Jared D Ohrt4, Timothy Ritter6, Jennifer B Smilowitz2, Nicholai E Wingreen7.
Abstract
The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education, and professional practice of medical physics. The AAPM has more than 8000 members and is the principal organization of medical physicists in the United States. The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for the purpose of revision or renewal, as appropriate, on their fifth anniversary or sooner. Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized. The following terms are used in the AAPM practice guidelines: Must and Must Not: Used to indicate that adherence to the recommendation is considered necessary to conform to this practice guideline. While must is the term to be used in the guidelines, if an entity that adopts the guideline has shall as the preferred term, the AAPM considers that must and shall have the same meaning. Should and Should Not: Used to indicate a prudent practice to which exceptions may occasionally be made in appropriate circumstances.Entities:
Keywords: Dose calculations; MPPG; practice guideline; treatment planning
Mesh:
Year: 2022 PMID: 35950259 PMCID: PMC9512346 DOI: 10.1002/acm2.13641
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
FIGURE 1Workflow of TPS dose algorithm commissioning, validation, and routine QA. The numbers refer to sections of this report
Detectors suitable for TPS commissioning and validation of photon and electron beams
| Detectors | Uses | Comments | References |
|---|---|---|---|
| Scanning ion chambers | Beam scanning for photons and electrons | Typical scanning chambers have an air cavity of 4–6 mm diameter | TG‐106 |
| Electron diodes and film | Beam scanning for electrons, output factors | Diodes are recommended over ion chambers with electrons to reduce stopping power dependency | TG‐25, |
| Small field detectors | Small field scanning & output factors, | Carefully select the detector type and size to fit the application. When scanning for penumbra, diodes are recommended. | TG‐106, |
| Large ion chamber | Aggregate MLC transmission factors | Interleaf transmission | LoSasso |
| Film and/or array detector | 2D dose distributions, including dynamic/virtual wedge and planar fluence maps, intraleaf measurements |
Absolute dosimetry preferred, relative dosimetry adequate. Desirable if the device can be mounted on the gantry and/or in a phantom at different geometries | TG‐106, |
If a diode detector is used for small field measurements, a “daisy chain” approach is recommended to minimize the energy‐dependence effects; the diode is first cross‐compared with an ion chamber for a larger field and then is used to measure the smaller fields.
Using film for intraleaf transmission is usually less precise than interleaf transmission.
Equipment required for TPS commissioning of photon and electron beams
| Equipment | Uses | Comments | References |
|---|---|---|---|
| 3D water phantom | Beam scanning | Must have sufficient scanning range and lateral/depth scatter | TG‐106, |
| Electrometers and cables | Beam scanning, output calibration, relative and absolute dosimetry | ADCL calibration, low noise and leakage with wide dynamic range and linear response | TG‐51, |
| Buildup cap or mini phantom | In‐air output factor measurement | Measurements required for some planning systems, some second check systems | TG‐74, |
| Water‐equivalent phantom material in slab form | Buildup and backscatter for measurements | >20 cm of total thickness in varying increments, width and length >30 cm, cavity for detector(s) | TG‐106, |
| CT density phantom | CT number to electron or mass density calibration | Should include tissue‐equivalent materials spanning the clinical range of low‐density lung to high‐density bone. | TG‐66 |
| Array detector | Nonphysical wedge measurement and other 2D dose distributions | The array detector should be calibrated for each energy it is used for. | TG‐120 |
| Heterogeneity phantom with lung‐equivalent material | End‐to‐end testing | Include cavities for detectors, useful for annual QA reference test | TG‐65, |
| Anthropomorphic phantom | Anatomic model testing, end‐to‐end testing, use testing | Include cavities for detectors | IAEA TRS‐430 |
| Software for data processing | Processing, comparing, and analyzing profiles, depth‐dose curves, and other beam data | May be included with the 3D water tank scanning software | TG‐106 |
| IMRT/VMAT or arc therapy phantom | VMAT or arc therapy | Options include a solid phantom holding a planar array, 3D detector arrays, film inside a phantom, other | TG‐120 |
Approximate values of rLCPE for typical clinical photon energies. The FWHM values represent the smallest field size that can be confidently measured for two representative detector sizes without special small field dosimetry considerations or verifications
| %dd(10,10)x | Representative Photon Energy (MV) |
| Minimum FWHM for representative 7 mm largest external dimension (cm) | Minimum FWHM for representative 3 mm largest external dimension (cm) |
|---|---|---|---|---|
| 66 | 6 | 1.0 | 2.7 | 2.3 |
| 73 | 10 | 1.6 | 3.9 | 3.5 |
| 77 | 15 | 1.9 | 4.5 | 4.1 |
| 80 | 18 | 2.1 | 4.9 | 4.5 |
| 63 | 6FFF | 0.8 | 2.3 | 1.9 |
| 71 | 10FFF | 1.4 | 3.5 | 3.1 |
FF refers to a flattening filter free beam.
TPS modelling checks and tolerances
| # | Objective | Examples | Tolerances |
|---|---|---|---|
| 5.1 | Verify that TPS reproduces the absolute dose under reference calibration conditions | Calculate the absolute dose per MU to the reference point using the same SSD and field size as calibration protocol | 0.5% |
| 5.2 | Compare calculated dose in clinical planning module to calculated dose in physics module | Compare dose profiles computed in the modeling tool to profiles computed in the clinical planning module for a large (> 30 × 30 cm2) field | Within statistical, grid size, and material definition uncertainty |
| 5.3 | Perform spot checks comparing calculated dose in clinical planning module to data collected during commissioning | Calculate and compare PDD, output factors, and/or off axis factors for nonreference depths, off axis positions, and field sizes | 2% |
Applicable if the treatment planning system has a physics module for beam modeling.
When comparing profiles such as PDD, this tolerance is the local difference.
Basic photon beam validation checks. See Table 6 for tolerances
| # | Dose comparison objective | Examples |
|---|---|---|
| 5.4 | Small static (non‐SRS) MLC‐shaped fields | IAEA TRS 430 |
| 5.5 | MLC transmission, leaf overtravel, and output factor effects | Large field with extensive blocking such as mantle, IAEA TRS 430 Photon Test 3 |
| 5.6 | Off‐axis modeling including primary collimator | Off‐axis field, IAEA TRS 430 Photon Test 2, measure with diagonal profiles or collimator rotation |
| 5.7 | Field divergence and depth dose changes with SSD | Asymmetric field at minimum SSD, IAEA TRS 430 Photon Test 6 |
| 5.8 | Oblique surface incidence | 10 × 10 cm2 field at 20° incidence, IAEA TRS 430 Photon Test 10 |
| 5.9 | Nonphysical wedge fluence modifiers | Large (>15 cm) field for each nonphysical wedge angle |
Multiple objectives can be consolidated into one measurement field, such as 5.5 and 5.6.
5.4–5.8 are intended for each open and (hard) wedged field. Nonphysical wedges are considered an extension of the corresponding open field in terms of spectra and only require the addition of 5.9.
Basic TPS photon beam evaluation methods and tolerances
| Region | Evaluation methods | Tolerances |
|---|---|---|
| High dose | Relative dose with one parameter change from reference conditions | 2% |
| Relative dose with multiple parameter changes | 5% | |
| Penumbra | Distance to agreement | 2 mm |
| Low‐dose tail | Up to 5 cm from field edge | 3% of max field dose |
Tolerances are relative to local dose unless otherwise noted, and are consistent with those used by IROC Houston.
For example, off‐axis with physical wedge.
2 mm aligns with evaluation tolerances used in TG‐53 1 and TG‐218.41.
Heterogeneous TPS photon beam validation checks
| # | Objective | Examples | Tolerances | References |
|---|---|---|---|---|
| 6.1 | Validate planning system reported electron (or mass) densities against known values | CT‐density calibration for air, lung, water, dense bone, and possibly additional tissue types | – | TG‐65, |
| 6.2 | Heterogeneity correction distal to lung tissue | 5 × 5 cm2, measure and calculate dose ratio above and below heterogeneity, outside of the buildup region | 3% | IAEA TRS‐430, |
Tolerances are relative to local dose unless otherwise noted.
VMAT/IMRT checks. See Table 9 for tolerances
| # | Objective | Examples | Detectors | References |
|---|---|---|---|---|
| 7.1 | Verify small field PDD | < 2 × 2 cm2 MLC shaped field, with PDD acquired at a clinically relevant SSD | Diode or plastic scintillator | TG‐155 |
| 7.2 | Verify output for small MLC‐defined fields | Use small square and rectangular MLC‐defined segments, measuring output at a clinically relevant depth for each | Diode, plastic scintillator, mini chamber or microion chamber | TG‐155, |
| 7.3 | TG‐119 tests | Plan, measure, and compare planning and QA results to the TG‐119 report for both the Head and Neck and C‐shape cases | Ion chamber, film, and/or array | TG‐119 |
| 7.4 | Clinical tests | Choose at least two relevant clinical cases; plan, measure, and perform an in‐depth analysis of the results | Ion chamber, film, and/or array | TG‐218 |
| 7.5 | External review | Simulate, plan, and treat an anthropomorphic phantom with embedded dosimeters. | Various options exist | Kry et al. |
A bar pattern scanned with a diode can be used to obtain additional absolute dose profile comparison in the direction perpendicular to MLC movement 64.
If IROC Houston service is used, they typically employ TLDs and radiochromic film. Certain commercial phantoms can accommodate ion chambers for point dose measurements.
VMAT/IMRT evaluation methods and criteria
| Measurement method | Regions | Evaluation criteria |
|---|---|---|
| Ion Chamber | Low‐gradient target region | 2% of prescribed dose |
| High‐gradient (OAR) region | 3% of prescribed dose | |
| Planar/Volumetric Array | All regions | No minimum pass rate, |
| End‐to‐End | Low‐gradient target region | 5% of prescribed dose |
The QMP should follow TG‐218 41 to develop tolerances for pretreatment QA that are appropriate at their clinic.
Application of a local 2%/2 mm gamma criterion can result in the discovery of easily correctable problems with IMRT commissioning.
Basic TPS validation tests for electron beams and minimum tolerance values
| # | Dose comparison objective | Examples | Tolerances |
|---|---|---|---|
| 8.1 | Verify that TPS reproduces the absolute dose under reference calibration conditions | Calculate the absolute dose per MU to the reference point using the same SSD and cone size as the calibration protocol | 0.5% |
| 8.2 | Basic model verification with shaped fields | PDD and profile comparison of custom cutouts at standard and extended SSDs | 3%/3 mm |
| 8.3 | Surface irregularities/obliquity | PDD and/or profile comparison of obliquely incident fields using reference cone and nominal clinical SSD | 5% |
| 8.4 | Inhomogeneity test | Compare manual dose calculation or chamber measurement to TPS with reference cone and nominal clinical SSD | 7% |
|
|
|
|
|
| 1‐2 | QMP understands algorithms has received proper training | ||
| 2 | Develop schedule to acquire data, model and verify the dose algorithms | ||
| 3 | Consult manufacturer's guidance for data acquisition | ||
| 3.A | Verify operation, calibration of measurement equipment | ||
| 3.B | Acquire appropriate CT calibration data | ||
| 3.C | Review detector and corrections needs for small field measurements | ||
| 3.D | Review raw data (compare with published data, check for error, confirm import into TPS) | ||
| 4 | Model per manufacturer's instructions, compare to published data for same machine/TPS | ||
| 5.1 | Verify that TPS reproduces the absolute dose under reference calibration conditions | 0.5% | |
| 5.2 | Compare calculated dose in clinical planning module to calculated dose in physics module | Within statistical, grid size, and material definition uncertainty | |
| 5.3 | Perform spot checks comparing calculated dose in clinical planning module to commissioning data | 2% | |
| 5.4 | Small static (non‐SRS) MLC‐shaped fields |
2% in high dose region (one parameter change) 5% in high dose region (multiple parameters) 2 mm in penumbra 3% of max dose in low dose tail (up to 5 cm from field edge)
| |
| 5.5 | MLC transmission, leaf overtravel, output factor effects | ||
| 5.6 | Off axis modelling including primary collimator | ||
| 5.7 | Field divergence and depth dose changes with SSD | ||
| 5.8 | Oblique surface incidence | ||
| 6.1 | Validate planning system reported electron (or mass) densities against known values | ||
| 6.2 | Heterogeneity correction distal to lung tissue | 3% | |
| 7.1 | Verify small field PDD |
2% of prescribed dose with ion chamber in low‐gradient region 3% of prescribed dose with ion chamber in high‐gradient region Evaluate planar/volumetric results at 2% local/2mm, pre‐treatment QA should pass >95% using 3% global/2mm and 10% threshold 5% of prescribed dose with end‐to‐end tests in low‐gradient region | |
| 7.2 | Verify output for small MLC‐defined fields | ||
| 7.3 | TG‐119 tests | ||
| 7.4 | Clinical tests | ||
| 7.5 | External review | ||
| 8.1 | Verify the TPS reproduces the absolute dose under reference calibration conditions for electron energies | 0.5% | |
| 8.2 | Basic electron model verification with shaped fields | 3%/3 mm | |
| 8.3 | Electron surface irregularities/obliquity | 5% | |
| 8.4 | Electron inhomogeneity test | 7% | |
| 9 | Routine QA established, including baseline QA plan(s) identified for each configured beam |