Jinghua Tang1, Liudi Jiang1, Michael McGrath1,2, Dan Bader3, Piotr Laszczak1, David Moser1, Saeed Zahedi2. 1. School of Engineering, Faculty of Engineering and Physical Sciences, University of Southampton, Southampton, UK. 2. Blatchford Products Ltd., Basingstoke, UK. 3. Skin Health Research Group, Faculty of Environmental and Life Sciences, University of Southampton, Southampton, UK.
Abstract
The study was designed to establish a biomechanical assessment platform for the lower limb residuum/socket interface as a function of duration and speed of movement. The approach exploits an interface sensor which measures multi-directional stresses at the interface. The corresponding interface coupling motion was assessed using a 3D motion capture system. A longitudinal study, involving a trans-femoral amputee, was conducted with nine repeated level walking sessions over a 12-month period. The effect of walking speed on interface biomechanics was also assessed. Interface peak pressures and shear stresses in the range of 55-59 kPa and 12-19 kPa were measured, respectively, over all sessions in the 12 months study period at the posterior-proximal location of the residuum. The peak pressure and longitudinal shear values were found to fluctuate approximately 11% and 40% as against its maximum value, respectively, over 12 months. In addition, up to 12° of angular coupling and up to 28 mm of pistoning were recorded over a gait cycle, which was found to change by 29% and 45% respectively over the study period. The variation in walking speed, by altering self-selected cadence, resulted in changes of pressure and shear stresses at mid-stance of the gait cycle. In particular, as compared with self-selected cadence, for fast speed, peak pressure and peak longitudinal shear stress decreased by 5% and 33%, respectively. For slow speed, peak pressure and peak longitudinal shear stress increased by 7% and 17%, respectively. The corresponding angular and pistoning revealed a variation of up to 29% and 45%, respectively. This biomechanical assessment approach shows promise in the quantitative assessment of interface kinematics and kinetics for lower limb prosthetics, the usage of which could assist the clinical assessment of prosthetic socket fit.
The study was designed to establish a biomechanical assessment platform for the lower limb residuum/socket interface as a function of duration and speed of movement. The approach exploits an interface sensor which measures multi-directional stresses at the interface. The corresponding interface coupling motion was assessed using a 3D motion capture system. A longitudinal study, involving a trans-femoral amputee, was conducted with nine repeated level walking sessions over a 12-month period. The effect of walking speed on interface biomechanics was also assessed. Interface peak pressures and shear stresses in the range of 55-59 kPa and 12-19 kPa were measured, respectively, over all sessions in the 12 months study period at the posterior-proximal location of the residuum. The peak pressure and longitudinal shear values were found to fluctuate approximately 11% and 40% as against its maximum value, respectively, over 12 months. In addition, up to 12° of angular coupling and up to 28 mm of pistoning were recorded over a gait cycle, which was found to change by 29% and 45% respectively over the study period. The variation in walking speed, by altering self-selected cadence, resulted in changes of pressure and shear stresses at mid-stance of the gait cycle. In particular, as compared with self-selected cadence, for fast speed, peak pressure and peak longitudinal shear stress decreased by 5% and 33%, respectively. For slow speed, peak pressure and peak longitudinal shear stress increased by 7% and 17%, respectively. The corresponding angular and pistoning revealed a variation of up to 29% and 45%, respectively. This biomechanical assessment approach shows promise in the quantitative assessment of interface kinematics and kinetics for lower limb prosthetics, the usage of which could assist the clinical assessment of prosthetic socket fit.
Since the inception of modular designs for lower limb prostheses in 1950,
there has been a rapid surge of a range of advanced limb designs,
the use of which has led to improved amputee care. However, many amputees
still report issues related to socket fit and stump pain induced by prosthetic
limbs, leading to poor satisfaction rate.
This has become a challenge to UK National Health Service for lower limb
amputee care.Similar to the gait cycle (GC) of able-bodied subjects, an amputee GC can be broadly
divided into stance and swing phases. As the foot contacts the ground in stance
phase, a ground reaction force is generated, acting upon the prosthetic foot.
Subsequently, the load is transferred through the prosthetic components, for example
ankle and knee of the trans-femoral amputees, to the socket interface.
Load transfer from the ground to the socket can reach up to 270% of body
weight and such load can occur up to 2 h/day. These multi-directional loads will be
re-distributed over the residuum, via the socket. However, many amputees suffer from
diabetes, vascular disease and/or peripheral neuropathy, thus the soft tissues
covering their residuum are less tolerant to loading and highly susceptible to breakdown,
leading to the formation of stump ulcers.
Thus, the need to measure pressure and shear at the residuum/socket interface
has been well recognised in this field.Indeed, residuum movement has been previously reported to be of magnitude of up to
40 mm in the axial direction
and up to 7° in the sagittal plane.
Excessive motions at the interface could also compromise gait stability and
the effective safety of the amputee.
It is thus important to evaluate the real-time interface biomechanics during
a normal functional activity. We have developed such a comprehensive assessment
platform combining the novel interface coupling model for kinematic measurements and
a tri-axial pressure and shear sensing system for kinetic measurements.This longitudinal study has been designed to evaluate the characteristics of gait for
a single participant on nine separate sessions over a 12 month period with an
additional focus on the effect of walking speed on interface biomechanics.
Materials and methods
The participant
To ensure consistency and comparability across multiple sessions, a unilateral
trans-femoral amputee (male, 29 years, body mass of 80 kg, height of 178 cm,
post amputation of more than 20 years) participated in the study. He had a
stable residual limb, free from infection and was able to walk without
assistance. The participant was fitted with his habitual prosthesis throughout
the study, including a supra-condylar suspension socket, KX06 knee and Echelon
VT foot (Blatchford Products Ltd., Basingstoke, UK). A senior prosthetist
verified the socket fit and the prosthetic alignment. The study was approved by
the University Research Ethics Committee (ID: 12058 and ID: 6008) of the
University of Southampton.
The experiment setup and theory
Figure 1 shows the
overall experimental setup, consisting of a tri-axial stress sensor system for
the interface kinetic measurement and a conventional two-camera CODA 3D motion
capture system (Charlwood Dynamics Ltd., Leicestershire, UK) to provide data for
interface kinematics. CODA marker placement protocol has been followed according
to CDL Gait-Cluster & Pelvic Slider version 1.06. On the prosthetic side,
four reflective markers secured on a rigid cluster frame (outlined in white
lines in Figure 1),
were placed on top of the shorts, which was firmly strapped to the lower part of
the socket that was not covered by the short. This was to ensure no relative
movement between the reflective makers and the socket.
Figure 1.
Schematic to depict the experimental test protocol in a gait laboratory,
indicating the interface sensor and the marker placement on the
participant, as well as the synchronisation between the sensor system
and the 3D motion capture system.
Schematic to depict the experimental test protocol in a gait laboratory,
indicating the interface sensor and the marker placement on the
participant, as well as the synchronisation between the sensor system
and the 3D motion capture system.The tri-axial pressure and shear sensor system are detailed in a previous paper
by the authors.
It consists of three sensor units and a wireless data acquisition unit
(DAQ). Individual sensors were located at the anterior-proximal (AP),
anterior-distal (AD) and posterior-proximal (PP) sites, providing a real-time
output of interface pressure and shear during walking (Figure 2(a)). All sensor units are thin
(1 mm thick) and very flexible. The suitability of its usage within the socket
without causing any discomfort, nor the need to alter socket has been assured in
our studies[11,12] based on the amputee’s feedback and assessment of a
senior prosthetist who were present throughout the tests.
Figure 2.
(a) Sensor placement on the inner socket wall and the corresponding
locations on the residuum (b) an illustration to define virtual residuum
segment (VRS), socket segment (SS), angular coupling (α) and pistoning
(L).
(a) Sensor placement on the inner socket wall and the corresponding
locations on the residuum (b) an illustration to define virtual residuum
segment (VRS), socket segment (SS), angular coupling (α) and pistoning
(L).The residuum/socket interface kinematics, involving pistoning and angular
coupling, was based on marker data extracted from the 3D motion capture system,
as detailed in our previous work.
To review briefly, two separate segments are constructed at the socket
interface, namely the Virtual Residuum Segment (VRS) and Socket Segment (SS), as
shown in Figure 2(b).
Local coordinate systems are defined for both the VRS and SS to calculate the
relative angular movement between VRS and SS in real time. Angular coupling in
sagittal plane (α) was chosen for analysis as it represents the plane with
greatest movement based on the previous study.
The dynamic pistoning (L), calculated as a dynamic displacement between
the hip joint centre and prosthetic knee pivot centre, characterised the
pistoning between the VRS and SS.A synchronisation device was designed and implemented to allow simultaneous
measurements of the interface stresses and couplings during a single walking
test. Effectively, when acquisition of the marker position started, a 5 volt
trigger pulse was transmitted from the motion capture system to register the
start of the data acquisition.
Walking tests
Physical makers for 3D motion capture and sensors for stress measurement were
placed on the participant by the same investigator. The participant was
subsequently asked to walk at self-selected speed along the 8 m level walkway
with a force plate embedded approximately at its half-way point. At least eight
clean trials were collected in a single evaluation session. A clean trial is
defined as trials in which all markers are captured by the cameras and there was
a complete single foot in contact with the force plate. In total, nine
evaluation sessions were conducted over a 12 month period, out of which four
were for assessing interface stress using the stress sensing system, four were
for assessing interface coupling using the 3D motion capture system, and the
final one was designed to combine both assessments simultaneously. For the first
eight sessions, the stress and coupling measurement sessions were alternated
with at least one month separation between the two consecutive sessions, over
the 12 month period.In the last session, the walking cadence, in steps per minute, was calculated
based on the self-selected walking speed of the participant. The calculated
walking cadence was then used to approximate slow and fast walking, represented
by −20% and +20% of self-selected cadence. The participant was subsequently
instructed to repeat the level walking tests at these slow and fast walking
speeds, as aided by a metronome. A minimum of eight clean trials were conducted
for both slow and fast walking tests.The initial stump shape and mass distribution was assessed by a senior
prosthetist. At the end of each session, body mass was recorded as an indication
of any possible change in stump mass distribution, over the 12 month study
period.
Data processing and participant feedback
During each session over the 12 months, the peak values for each of the kinematic
and/or kinetic measurements were extracted from each clean trial. Median values
of measurement parameters, such as peak pressure and peak shear stress, were
obtained from each session and used for analysis. In particular, the fluctuation
of the medians, over the study period, were reported as percentages of the
observed maximum out of all trials. In addition, the percentage change in peak
interface kinematics and kinetics, associated with fast and slow walking speeds,
were estimated with respect to those evident during level walking.After each session, participant feedback was recorded on the health state of the
residuum, change in socket fit due to body mass fluctuation and satisfaction
level of the prosthetic limb.
Results
Longitudinal interface kinetic assessment over 12 months
Figure 3 illustrates the
pressure and the shear stresses in the circumferential (Sc) and
longitudinal (SL) directions measured at the three residuum
locations, taken from a single measurement session. Pressure profiles reveal a
double-hump profile during the stance phase at each location (Figure 3(a)). Peak
pressures of up to 55, 30 and 30 kPa were recorded, in that particular session
at PP, AP and AD locations, respectively. In the swing phase, the pressure was
restored to the original value seen at initial contact (IC), that is, 0% GC.
With respect to Sc, values of up to −5 kPa were measured at the two proximal
locations with a corresponding value of up to −11 kPa measured at AD (Figure 3(b)). Up to 13
and 12 kPa of SL were measured at PP and AD locations, respectively,
with values of < 2 kPa at AP (Figure 3(c)).
Figure 3.
Mean and SD of (a) pressure, (b) circumferential shear stress
(SC) and (c) longitudinal shear stress (SL)
obtained at PP, AD, and AP locations in Session 2.
Mean and SD of (a) pressure, (b) circumferential shear stress
(SC) and (c) longitudinal shear stress (SL)
obtained at PP, AD, and AP locations in Session 2.It is evident that the peak values for pressure and both shear stresses are
higher for PP and AD locations when compared to the AP location. Figure 4 illustrates the
median values for peak pressure and longitudinal stress across the five
measurement sessions at the PP and AD locations. Over the 12 month period, there
were up to 11% and 40% change in the median values of the peak PP pressure and
SL, respectively. The corresponding changes at the AD location
were 24% and 23%, respectively. The mean peak P and SL values in
Figure 4 showed no
clear chronological trend, from Session 1 to Session 5.
Figure 4.
Peak pressure and longitudinal stress, SL, at (a) PP location
and (b) AD location of the residuum, over five measurement sessions. In
each session, results were calculated from 10 walking tests. The boxes
are bounded by inter quartile range (IQR) and divided by a line,
representing median value. Circles inside the box represent the mean
value. Solid diamond symbol represents the outlier.
Peak pressure and longitudinal stress, SL, at (a) PP location
and (b) AD location of the residuum, over five measurement sessions. In
each session, results were calculated from 10 walking tests. The boxes
are bounded by inter quartile range (IQR) and divided by a line,
representing median value. Circles inside the box represent the mean
value. Solid diamond symbol represents the outlier.
Longitudinal interface kinematic assessment over 12 months
Results indicated that the majority of the lower limb movement (α < 12°) is in
the sagittal plane during the gait cycle (Figure 5(a)) with correspondingly small
values in both coronal and transverse planes. In the stance phase, there was a
general increase of α from IC to toe-off (TO), whereas in the swing phase, a
decrease in α was evident such that it was restored to the original value at IC.
Up to 29 mm of pistoning was measured over a GC (Figure 5(b)). There was a general
decrease of L from IC to approximately 20% of the GC, suggesting distal movement
of the residuum. Beyond this point, there was a slight increase of L until
mid-stance phase. A subsequent decrease of L was evident between 30% and 50% of
GC. During the remainder of the GC, L increases and is restored to the original
value at IC.
Figure 5.
(a) Mean and SD of the interface angular coupling in sagittal plane, α
and (b) pistoning, L obtained over a GC in Session 3.
(a) Mean and SD of the interface angular coupling in sagittal plane, α
and (b) pistoning, L obtained over a GC in Session 3.Figure 6 illustrates the
median values for peak α and L across the five measurement sessions. These
results reveal changes in the kinematic parameters of up to 29% and 45%,
respectively, over the 12 month period.
Figure 6.
Absolute values of peak angular coupling in sagittal plane and pistoning,
over five measurement sessions. In each session, results were calculated
from eight walking trials. The boxes are bounded by inter quartile range
(IQR) and divided by a line, representing median value. Circles inside
the box represents the mean value.
Absolute values of peak angular coupling in sagittal plane and pistoning,
over five measurement sessions. In each session, results were calculated
from eight walking trials. The boxes are bounded by inter quartile range
(IQR) and divided by a line, representing median value. Circles inside
the box represents the mean value.Approximately ±2% of body mass variation has been recorded over the 12 month
study period. The senior prosthetist confirmed that the body mass variation
measured over the study period is insignificant to result in a notable change in
stump volume or shape.
Effect of walking speed on interface biomechanics
In the mid-stance trough, a pressure of up to 43 kPa was measured at the PP
location at the self-selected walking speed (Figure 7(a)). The change in walking
speeds resulted in a corresponding change of up to 12%, with respect to the
values obtained in self-selected condition. In addition, SL values of
up to 6 kPa were obtained at PP location at self-selected walking speed in the
mid-stance phase (Figure
7(b)). The changes in walking speeds resulted in a corresponding
change of up to 67%.
Figure 7.
(a) Pressure and (b) longitudinal shear, SL at PP location,
obtained at slow, self-selected and fast walking speed, over a GC. (c)
Pistoning, L and (d) angular coupling in sagittal plane, α obtained,
obtained at slow, self-selected and fast walking speed, over a GC.
(a) Pressure and (b) longitudinal shear, SL at PP location,
obtained at slow, self-selected and fast walking speed, over a GC. (c)
Pistoning, L and (d) angular coupling in sagittal plane, α obtained,
obtained at slow, self-selected and fast walking speed, over a GC.With respect to L, values of up to −17 mm were estimated when walking at
self-selected walking speed in the mid-stance (Figure 7(c)). The changes in walking
speed resulted in a corresponding change of approximately 36%. In swing phase,
up to 12 mm of residuum motion was measured at self-selected speed. The changes
in walking speed resulted in a corresponding change of up to 42%. With respect
to α, peak values of up to 10° were estimated when walking at the self-selected
walking speed (Figure
7(d)).
Discussion
This study investigates the change in residuum/socket interface kinematics and
kinetics, over a period of 12 months and at different walking speed, using the
techniques report previously.[11-13] Interface stresses and
coupling profiles were repeatedly measured over this time. Findings revealed that
the magnitude of pistoning and angular couplings were similar to those in the
seminal study by Convery and Murray.
Residuum/socket interface biomechanics over 12 months
Stresses obtained at PP and AD locations were chosen to analyse the fluctuation
over the 12 month period, as they represent the key load bearing
locations.[14,15] The relatively low interface stresses at the AP
location as compared with PP and AD is also evidenced in Figure 3. In general, we observed up to
11% (pressure), and 40% (SL) changes for the interface stresses, as
well as 29% (α) and 45% (L) changes for the relative motion in the five
measurement sessions over the 12 month study. The percentage values represent
the absolute change in stress and pistoning of only 8 kPa and 12 mm,
respectively, so any conclusions must be treated cautiously. Indeed, the authors
propose further investigations of this type to evaluate the clinical
significance of these findings.Over the course of the study, there was no significant change in participant’s
body mass and the mature stump was considered to be in a healthy condition. In
addition, the participant used the same prothesis in each session with alignment
checked by a single certified prosthetist and each measurement session was
started at a similar time of day. In addition, feedback was recorded after each
measurement session and neither discomfort nor change in socket fit were
reported. The change in pressure and shear stresses observed in this study can
be a direct result of several factors. These include the effect of donning and
doffing of the prothesis which may influence the forces exerted on the residuum
tissues, causing subsequent change in peak values of stress. In addition,
residuum volume fluctuation
may result in a change in socket fit condition leading to a
re-distribution of stresses over the residuum. From an experimental perspective,
the placement of each sensor may not have been reproducible, although effort was
made to mark the sensor location on the inner socket wall as a guidance for
sensor placement.To the best of our knowledge, there is few reported studies focussing on
interface stress variation over time, especially for that of trans-femoral
amputees. Sanders et al.
reported 30 kPa pressure variation (equivalent to 36% of mean peak
pressure) and 5 kPa shear variation (equivalent to 43% of mean peak shear) over
a 6-month period for trans-tibial amputees. We discuss here our trans-femoral
amputee results alongside with this previously reported trans-tibial result with
a view to providing preliminary comparison of longitudinal changes across these
two main amputee populations. It is noted that all these reported changes for
trans-tibial amputees
are higher than the 11% measured at the PP location in this present
study. This difference might be attributed to the bulk tissue mass on a
trans-femoral residuum, which dominate the pressure distribution profile at
proximal location of the residuum. By contrast, the change in pressure
(approximately 24%) at the AD location in the present study was comparable to
that previously reported. This could be predicted given that the present sensor
was placed at the cut end of the femur, representing a bony prominence, and the
pressures were dominated by the interaction between the bone and the socket,
similar to the case on trans-tibial amputees.In this study, change in SL ranges in a range of 6%–23% was obtained
ta AD location, which was lower than the value previously reported.
This may be explained by the differences in the sockets in trans-tibial
and trans-femoral designs. Indeed, trans-tibial sockets typically involve a
tighter fit allowing more shear stress to be transmitted between the residuum
and the socket.
Effect of walking speed on socket interface biomechanics
It was evident from Figure
7(a) that an increase in walking speed resulted in the decrease in
pressure in the mid-region of the stance phase (approximately 30% of GC). This
can be potentially explained by the increased movement of the body centre of
mass in a vertical direction, associated with an increase in walking speed.
The participant in the present study, who has been using the prothesis
for over 10 years, has optimised his walking strategy to minimise the metabolic
energy consumption at different walking speeds. Indeed, the increase in centre
of mass movement in vertical direction is one of the strategies to conserve
metabolic energy.The increase in walking speed resulted in a reduction in pistoning (Figure 7(d)) relative to
the socket, at approximately 30% of GC. This reduction may be potentially
explained by the efficient use of musculoskeletal work to maintain the residuum
in a stable position and achieve foot-flat during mid-stance phase.The increase in walking speed also resulted in an increased value of peak angular
coupling in the sagittal plane. This may be predicted given the greater force at
the PP location when compared to that at AP location, resulting in a pressure
gradient as walking speed increases (Figure 3(a)). The pressure gradient
will, in turn, translate the residuum movement towards the posterior location of
the socket.It is also worth noting that the variation of speed alone resulted in approximate
changes of 12% and 67% in pressure and SL, with corresponding changes
of 36% (pistoning) and 30% (angular coupling) in the relative motion at the
interface. The changes, resulted from the variation in walking speed, were in
similar ranges as compared with the changes observed over the 12 month period.
This further implies that, although the interface kinetic and kinematic
variations have been observed over a 12 month period, the degree of variation is
equivalent to those induced by change in daily activities, for example, walking
speeds. This was supported by the feedback from the participant during the
various test sessions, namely, there was no change apparent over the study
period. This may also imply that if quantified socket fit assessment is to be
adapted in future clinics, we may require larger variations of interface
biomechanics to evaluate socket fit levels. The combined kinetic-kinematic
strategy adopted in the present study could represent a promising approach to
assist such a quantitative assessment.
Conclusions
In this study, a platform for the assessment of socket interface biomechanics was
evaluated, using an interface stress sensor system (kinetics) and 3D motion capture
system (kinematics). Each method was experimentally evaluated on a trans-femoral
amputee, across five level walking sessions over a period of 12 months.Preliminary results showed changes of up to 40% and 45% in interface kinetics and
kinematics, respectively. In addition, the assessment platform was also found to be
sensitive to changes in walking speed. Such a combined biomechanical assessment
platform for the residuum/socket interface can potentially be used to aid the
current socket fitting process and design of the patient-specific adjustable sockets
and fully integrated limb systems, based on the socket movement and corresponding
interface stresses.
Authors: Arezoo Eshraghi; Noor Azuan Abu Osman; Hossein Gholizadeh; Mohammad Karimi; Sadeeq Ali Journal: Prosthet Orthot Int Date: 2012-01-22 Impact factor: 1.895
Authors: Jinghua Tang; Michael McGrath; Piotr Laszczak; Liudi Jiang; Dan L Bader; David Moser; Saeed Zahedi Journal: Med Eng Phys Date: 2015-10-30 Impact factor: 2.242