A pressure and shear sensor system for stress measurement at lower limb residuum/socket interface.

P Laszczak1, M McGrath2, J Tang2, J Gao2, L Jiang2, D L Bader3, D Moser4, S Zahedi4.   

Abstract

A sensor system for measurement of pressure and shear at the lower limb residuum/socket interface is described. The system comprises of a flexible sensor unit and a data acquisition unit with wireless data transmission capability. Static and dynamic performance of the sensor system was characterised using a mechanical test machine. The static calibration results suggest that the developed sensor system presents high linearity (linearity error ≤ 3.8%) and resolution (0.9 kPa for pressure and 0.2 kPa for shear). Dynamic characterisation of the sensor system shows hysteresis error of approximately 15% for pressure and 8% for shear. Subsequently, a pilot amputee walking test was conducted. Three sensors were placed at the residuum/socket interface of a knee disarticulation amputee and simultaneous measurements were obtained during pilot amputee walking test. The pressure and shear peak values as well as their temporal profiles are presented and discussed. In particular, peak pressure and shear of approximately 58 kPa and 27 kPa, respectively, were recorded. Their temporal profiles also provide dynamic coupling information at this critical residuum/socket interface. These preliminary amputee test results suggest strong potential of the developed sensor system for exploitation as an assistive technology to facilitate socket design, socket fit and effective monitoring of lower limb residuum health.
Copyright © 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Lower-limb amputees; Pressure and shear sensor; Residuum/socket interface

Mesh:

Year:  2016        PMID: 27118308     DOI: 10.1016/j.medengphy.2016.04.007

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  12 in total

Review 1.  Implantable biosensors for musculoskeletal health.

Authors:  Kylie E Nash; Keat Ghee Ong; Robert E Guldberg
Journal:  Connect Tissue Res       Date:  2022-02-17       Impact factor: 3.417

2.  A simulation-based analysis of the effects of variable prosthesis stiffness on interface dynamics between the prosthetic socket and residual limb.

Authors:  Michael A McGeehan; Peter G Adamczyk; Kieran M Nichols; Michael E Hahn
Journal:  J Rehabil Assist Technol Eng       Date:  2022-07-15

3.  Stakeholder perspectives for possible residual limb monitoring system for persons with lower-limb amputation.

Authors:  Lilly Tran; Ryan Caldwell; Matthew Quigley; Stefania Fatone
Journal:  Disabil Rehabil       Date:  2018-09-05       Impact factor: 3.033

4.  Interfacial pressure and shear sensor system for fingertip contact applications.

Authors:  Maria Valero; Nick Hale; Jing Tang; Liudi Jiang; Mike McGrath; Jianliang Gao; Piotr Laszczak; David Moser
Journal:  Healthc Technol Lett       Date:  2016-12-13

5.  Textile-based Pressure Sensors for Monitoring Prosthetic-Socket Interfaces.

Authors:  Jordan Tabor; Talha Agcayazi; Aaron Fleming; Brendan Thompson; Ashish Kapoor; Ming Liu; Michael Lee; He Helen Huang; Alper Bozkurt; Tushar Ghosh
Journal:  IEEE Sens J       Date:  2021-01-21       Impact factor: 3.301

Review 6.  Technologies to monitor the health of loaded skin tissues.

Authors:  Dan L Bader; Peter R Worsley
Journal:  Biomed Eng Online       Date:  2018-04-12       Impact factor: 2.819

7.  An Inductive Sensing System to Measure In-Socket Residual Limb Displacements for People Using Lower-Limb Prostheses.

Authors:  Katrina M Henrikson; Ethan J Weathersby; Brian G Larsen; John C Cagle; Jake B McLean; Joan E Sanders
Journal:  Sensors (Basel)       Date:  2018-11-09       Impact factor: 3.576

8.  A Deep Learning Approach to Non-linearity in Wearable Stretch Sensors.

Authors:  Ben Oldfrey; Richard Jackson; Peter Smitham; Mark Miodownik
Journal:  Front Robot AI       Date:  2019-05-08

9.  A Novel Multi-Axial Pressure Sensor Probe for Measuring Triaxial Stress States Inside Soft Materials.

Authors:  Giuseppe Zullo; Anna Leidy Silvestroni; Gianluca Candiotto; Andrey Koptyug; Nicola Petrone
Journal:  Sensors (Basel)       Date:  2021-05-17       Impact factor: 3.576

10.  A personalised prosthetic liner with embedded sensor technology: a case study.

Authors:  Linda Paternò; Vimal Dhokia; Arianna Menciassi; James Bilzon; Elena Seminati
Journal:  Biomed Eng Online       Date:  2020-09-14       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.