In vivo dosimetry in gynecological applications-A feasibility study.

Marion P R Van Gellekom1, Richard A M Canters2, Frank J W M Dankers2, Agnes Loopstra3, Elzbieta M van der Steen-Banasik3, Marie A D Haverkort3.   

Abstract

PURPOSE: To investigate the feasibility of in vivo dosimetry using microMOSFET dosimeters in patients treated with brachytherapy using two types of gynecological applicators. METHODS AND MATERIALS: In this study, a microMOSFET was placed in an empty needle of an Utrecht Interstitial Fletcher applicator or MUPIT (Martinez Universal Perineal Interstitial Template) applicator for independent verification of treatment delivery. Measurements were performed in 10 patients, with one to three microMOSFETs per applicator and repeated for one to four fractions, resulting in 50 in vivo measurements. Phantom measurements were used to determine characteristics of the microMOSFETs.
RESULTS: Phantom measurements showed a linear relationship between dose and microMOSFET threshold voltage, and a calibration coefficient (mV/cGy) was determined. Reproducibility of repeated 50 cGy irradiations was 2% (1 standard deviation). Distance and angle dependencies were measured and correction factors were determined. Subsequently, three microMOSFETs were placed in a phantom to measure a validation plan. The difference between predicted and measured dose was less than the measurement uncertainty (±9%, 2 standard deviations). In vivo measurements were corrected for distance and angle dependencies. Differences between predicted and measured dose in the patients were smaller than the measurement uncertainty for the majority of the measurements.
CONCLUSIONS: In vivo dosimetry using microMOSFETs in MUPIT and Utrecht Interstitial Fletcher applicators has proved to be feasible. Reimaging should be performed after detection of differences larger than 10% between predicted and measured dose to verify the applicator configuration. Movement of the applicator relative to the target or organs at risk is undetectable with this method.
Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Brachytherapy; Feasibility study; Gynecological; In vivo dosimetry; microMOSFET

Mesh:

Year:  2017        PMID: 28528720     DOI: 10.1016/j.brachy.2017.04.240

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  4 in total

Review 1.  Review of strategies for MRI based reconstruction of endocavitary and interstitial applicators in brachytherapy of cervical cancer.

Authors:  José Richart; Vicente Carmona-Meseguer; Teresa García-Martínez; Antonio Herreros; Antonio Otal; Santiago Pellejero; Ana Tornero-López; José Pérez-Calatayud
Journal:  Rep Pract Oncol Radiother       Date:  2018-07-23

2.  Two-dimensional real-time quality assurance dosimetry system using μ-Al2O3:C,Mg radioluminescence films.

Authors:  Luana F Nascimento; Dirk Verellen; Jo Goossens; Lara Struelens; Filip Vanhavere; Paul Leblans; Mark Akselrod
Journal:  Phys Imaging Radiat Oncol       Date:  2020-10-05

Review 3.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28

4.  In Vivo Verification of Treatment Source Dwell Times in Brachytherapy of Postoperative Endometrial Carcinoma: A Feasibility Study.

Authors:  Antonio Herreros; José Pérez-Calatayud; Facundo Ballester; Jose Barrera-Gómez; Rosa Abellana; Joana Melo; Luis Moutinho; Luca Tagliaferri; Ángeles Rovirosa
Journal:  J Pers Med       Date:  2022-05-31
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.