| Literature DB >> 35722048 |
Josefin Seth Caous1,2, Karin Svensson Malchau2,3, Max Petzold4, Ylva Fridell3, Henrik Malchau2,3,5, Linda Ahlstrom3,6, Peter Grant2,7, Annette Erichsen Andersson3,6.
Abstract
Background: Airborne bacteria present in the operating room may be a cause of surgical site infection, either contaminating the surgical wound directly, or indirectly via e.g. surgical instruments. The aim of this study was to evaluate if instrument and assistant tables equipped with local unidirectional airflow reduce bacterial contamination of the instrument area to ultra clean levels, during orthopedic implant surgery in an operating room with displacement ventilation.Entities:
Keywords: Bacterial count; Colony forming unit; Instrument table; Orthopedic surgery; Surgical instruments; Unidirectional airflow
Year: 2022 PMID: 35722048 PMCID: PMC9198428 DOI: 10.1016/j.infpip.2022.100222
Source DB: PubMed Journal: Infect Prev Pract ISSN: 2590-0889
Figure 1Instrument (A) and assistant (B) tables with UDAF units at the end of the tables and the air sampler centered on the tables. The operating room (C) with conventional ventilation and instrumentation tables placed at the end of the operating table. Instrumentation tables equipped with units creating a HEPA filtered horizontal unidirectional airflow (UDAF) over the surgical instruments were evaluated. Measurements were performed with the UDAF units either active or inactive. A holder for the air sampler was used, to enable reproduceable sampling at nearly the same location during every surgery. During the study only one active UDAF table, either instrument (A) or assistant (B) table, was used at a time. Example of the positioning of the tables relative the conventional ventilation of the operating room is shown in C.
Overview of surgery types during which measurements were sampled. Measurements were performed during 55 surgeries, however measurements from 48 surgeries were included in the analysis. Surgery types where one surgery has been excluded, due to too short procedure time, ongoing infection, or growth on reference plates, are indicated with∗
| Ass. table UDAF on | Ass. table UDAF off | Inst. table UDAF on | Inst. table UDAF off | Inst. table conv. | Sum | |
|---|---|---|---|---|---|---|
| Intramedullary nail | 3 | 4 | 5∗ | 5∗ | 2∗ | 19 |
| Plate osteosynthesis | 4 | 2∗ | 4∗ | 4∗ | 3 | 17 |
| Hemiarthroplasty | 2 | 3 | 3 | 3 | 5∗ | 16 |
| Other implants | 1 | 2 | 3 | |||
| Sum | 10 | 11 | 12 | 12 | 10 | 55 |
Local UDAF units reduce cfu in air and surface samples taken from instrumentation tables during ongoing orthopaedic surgery. The number of colony-forming units (cfu) was analysed in air and surface samples gathered from instrument and assistant tables equipped with unidirectional airflow (UDAF) units (either active or inactive during measurements) and conventional instrument tables. The mean and median of cfu in samples depending on condition is presented, as well as % of air samples fulfilling the recommendation of ≥10 cfu/m3 air. For instrument samples the limit was set to 0 cfu
| Sample | Active UDAF unit | Inactive/no UDAF unit | |
|---|---|---|---|
| No. of surgeries (air samples) | 10 (40) | 10 (38) | |
| Mean cfu/m3/surgery ± SD | 0.3 ± 0.78 | 13.4 ± 13,25 | |
| Median cfu/m3/surgery (min-max) | 0 (0–2,5) | 9,4 (2,8–45,5) | <0.001 |
| Air samples with cfu ≤ 10 | 100 % | 68 % | |
| No. of surgeries (instrument samples) | 10 (20) | 10 (20) | |
| Mean cfu/cm2/h/surgery ± SD | 0.004 ± 0.127 | 0.052 ± 0.0476 | |
| Median cfu/cm2/h/surgery (min-max) | 0 (0–0.4) | 0.035 (0-0,13) | 0.001 |
| Instrument samples with cfu = 0 | 95 % | 45 % | |
| No. of surgeries (air samples) | 10 (39) | 10 (39) | |
| Mean cfu/m3/surgery ± SD | 0.2 ± 0.26 | 5.0 ± 2.77 | |
| Median cfu/m3/surgery (min-max) | 0.1 (0–0.8) | 4.9 (1.3–10.8) | 0.002 |
| Air samples with cfu ≤ 10 | 100 % | 92 % | |
| No. of surgeries (instrument samples) | 10 (30) | 10 (30) | |
| Mean cfu/cm2/h/surgery ± SD | 0.004 ± 0.0027 | 0.17 ± 0.022 | |
| Median cfu/cm2/h/surgery (min-max) | 0 (0–0.02) | 0.015 (0–0.07) | 0.1 |
| Instrument samples with cfu = 0 | 93 % | 73 % | |
| Not Applicable | |||
| No. of surgeries (air samples) | 8 (32) | ||
| Mean cfu/m3/surgery ± SD | 8.0 ± 1.52 | ||
| Median cfu/m3/surgery (min-max) | 7.8 (2.8–14.0) | 0.5 | |
| Air samples with cfu ≤ 10 | 72 % | ||
| No. of surgeries (instrument samples) | 8 (24) | ||
| Mean cfu/cm2/h/surgery ± SD | 0.02 ± 0.022 | ||
| Median cfu/cm2/h/surgery (min-max) | 0.015 (0–0.06) | 0.9 | |
| Instrument samples with cfu = 0 | 67 % | ||
UDAF, unidirectional air flow; cfu, colony forming unit; SD, standard deviation.
Kruskal-Wallis nonparametric test, comparison between group.
Comparison between instrument table with active or inactive UDAF.
Comparison between instrument table with active UDAF and Conventional table.
Compared to instrument table with UDAF turned off.
Figure 2Local UDAF units significantly reduced air and surface contamination around the surgical instruments during orthopedic implant surgery. Results from active air sampling with a Sartorius MD-8 air scanner (left) and passive surface sampling with instrument dummies (stainless-steel coupons with a size of 4.3 x 4.3 cm) (right) at assistant tables (top row) and instrument tables (bottom row).
Regression analysis of mean cfu in samples from instrument dummies and air, and association with local UDAF, number of door openings, and persons present in the OR. Multiple linear regression was used to assess how air quality above the instrumentation tables as well as instrument dummy contamination was affected by number of door openings and number of persons present during surgery, as well as if local UDAF units were used or not
| Compared factors | Air contamination | Instrument contamination | ||||
|---|---|---|---|---|---|---|
| R-squared (adjusted) | Coefficient (95% CI) | R-squared (adjusted) | Coefficient (95% CI) | |||
| 0.45 (0.34) | 0.47 (0.37) | |||||
UDAF | -10.6 (-19.8–1.3) | 0.03∗ | -0.05 (-0.8–-0.01) | <0.01∗ | ||
Door openings | -0.4 (-3.2–2.5) | 0.8 | <-0.01 (-0.02 – 0.00) | 0.07 | ||
Persons | 4.2 (-1.2–9.6) | 0.1 | <0.01 (-0.02 – 0.00) | 0.9 | ||
| 0.67 (0.60) | 0.27 (0.13) | |||||
UDAF | -4.8 (-6.6–-2.9) | <0.001∗ | -0.01 (-0.3–0.00) | 0.1 | ||
Door openings | 0.4 (-0.3–1.1) | 0.2 | <0.01 (-0.00 – 0.01) | 0.3 | ||
Persons | 0.06 (-0.8–0.9) | 0.9 | <0.01 (-0.00 – 0.01) | 0.3 | ||
| 0.72 (0.66) | 0.25 (0.09) | |||||
UDAF | -7.6 (-10.7–-4.6) | <0.001∗ | -0.01 (-0.03–0.00) | 0.1 | ||
Door openings | -0.5 (-1.7–0.7) | 0.4 | <-0.01 (-0.01 – 0.00) | 0.5 | ||
Persons | 0.7 (-1.2–2.5) | 0.5 | <0.01 (-0.01 – 0.01) | 0.9 | ||
CI, confidence interwall; ∗ Significant at 5% level.