A mobile laminar airflow unit to reduce air bacterial contamination at surgical area in a conventionally ventilated operating theatre.

C Pasquarella1, G E Sansebastiano, S Ferretti, E Saccani, M Fanti, U Moscato, G Giannetti, S Fornia, P Cortellini, P Vitali, C Signorelli.   

Abstract

The aim of this study was to evaluate the efficacy of a mobile laminar airflow (LAF) unit in reducing bacterial contamination at the surgical area in an operating theatre supplied with turbulent air ventilation. Bacterial sedimentation was evaluated during 76 clean urological laparotomies; in 34 of these, a mobile LAF unit was added. During each operation, settle plates were placed at four points in the operating theatre (one at the patient area and three at the perimeter), a nitrocellulose membrane was placed on the instrument table and an additional membrane near the wound. During four operations, particle counting was performed to detect particles > or =0.5 microm. Mean bacterial sedimentation on the nitrocellulose membrane on the instrument table was 2730 cfu/m(2)/h under standard ventilation conditions, whereas it decreased significantly to a mean of 305 cfu/m(2)/h when the LAF unit was used, i.e. within the suggested limit for ultraclean operating theatres (P=0.0001). The membrane near the wound showed a bacterial sedimentation of 4031 cfu/m(2)/h without the LAF unit and 1608 cfu/m(2)/h with the unit (P=0.0001). Particle counts also showed a reduction when the LAF unit was used. No significant difference was found at the four points in the operating theatre between samplings performed with, and without, the LAF unit. Use of a mobile LAF unit with turbulent air ventilation can reduce bacterial contamination at the surgical area in high-risk operations (e.g. prosthesis implant).

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17669550     DOI: 10.1016/j.jhin.2007.05.022

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  11 in total

1.  Indoor air quality in university environments.

Authors:  Mara Di Giulio; Rossella Grande; Emanuela Di Campli; Soraya Di Bartolomeo; Luigina Cellini
Journal:  Environ Monit Assess       Date:  2009-11-26       Impact factor: 2.513

2.  Sterility of the personal protection system in total joint arthroplasty.

Authors:  Kenneth A Kearns; Dan Witmer; Junaid Makda; Javad Parvizi; Donald Jungkind
Journal:  Clin Orthop Relat Res       Date:  2011-11       Impact factor: 4.176

3.  Laminar air flow reduces particle load in TKA-even outside the LAF panel: a prospective, randomized cohort study.

Authors:  Stephanie Kirschbaum; Hagen Hommel; Peggy Strache; Roland Horn; Roman Falk; Carsten Perka
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-11-09       Impact factor: 4.342

4.  Five-year microbiological monitoring of wards and operating theatres in southern Italy.

Authors:  V La Fauci; C Genovese; A Facciolà; M A R Palamara; R Squeri
Journal:  J Prev Med Hyg       Date:  2017-06

5.  Mobile laminar air flow screen for additional operating room ventilation: reduction of intraoperative bacterial contamination during total knee arthroplasty.

Authors:  D Sossai; G Dagnino; F Sanguineti; F Franchin
Journal:  J Orthop Traumatol       Date:  2011-11-10

6.  Surgical Smoke and Airborne Microbial Contamination in Operating Theatres: Influence of Ventilation and Surgical Phases.

Authors:  Francesco Romano; Samanta Milani; Jan Gustén; Cesare Maria Joppolo
Journal:  Int J Environ Res Public Health       Date:  2020-07-27       Impact factor: 3.390

7.  Air sampling procedures to evaluate microbial contamination: a comparison between active and passive methods in operating theatres.

Authors:  Christian Napoli; Vincenzo Marcotrigiano; Maria Teresa Montagna
Journal:  BMC Public Health       Date:  2012-08-02       Impact factor: 3.295

Review 8.  Human factors and ergonomics as a patient safety practice.

Authors:  Pascale Carayon; Anping Xie; Sarah Kianfar
Journal:  BMJ Qual Saf       Date:  2013-06-28       Impact factor: 7.035

9.  Does a mobile laminar airflow screen reduce bacterial contamination in the operating room? A numerical study using computational fluid dynamics technique.

Authors:  Sasan Sadrizadeh; Ann Tammelin; Peter V Nielsen; Sture Holmberg
Journal:  Patient Saf Surg       Date:  2014-06-26

10.  Comparison of three distinct clean air suits to decrease the bacterial load in the operating room: an observational study.

Authors:  Piotr Kasina; Ann Tammelin; Anne-Marie Blomfeldt; Bengt Ljungqvist; Berit Reinmüller; Carin Ottosson
Journal:  Patient Saf Surg       Date:  2016-01-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.