F Lorchel1, D Nguyen1, A Mamou2, N Barbet1, J Camoesas3, Y Degluaire1, I Pouchard1, Y Queffelec3, F Renoult1, M Khodri1, J Farah3.
Abstract
Background and purpose: To investigate the reproducibility of deep-inspiration breath hold (DIBH) breast cancer treatments on Halcyon™ performed using the first clinical version of AlignRT InBore™ (Vision RT Ltd., London, UK), a Halcyon's SGRT dedicated solution. Materials and methods: The ease and feasibility of DIBH treatments was retrospectively investigated for the initial 22 left breast cancer patients treated on Halcyon™ using AlignRT InBore™. Setup time, Cone beam computed tomography (CBCT) imaging and analysis time as well as overall treatment time were recorded. Online and offline review of CBCT images was undertaken to verify the compliance of breast, heart, spine, sternum and diaphragmatic domes positions.Entities:
Keywords: AlignRT InBore™; Breast cancer; Deep-inspiration breath hold; Halcyon™
Year: 2022 PMID: 35662884 PMCID: PMC9156859 DOI: 10.1016/j.ctro.2022.05.002
Source DB: PubMed Journal: Clin Transl Radiat Oncol ISSN: 2405-6308
Fig. 1Boxplot representation of session duration (left) and setup time (right) for all 22 patients.
Fig. 2Mean setup time (up) and session duration (down) as a function of treatment fraction (left) and date (right).
Fig. 3Boxplot representation of CBCT shifts for all 22 patients in vertical (left), longitudinal (middle), and lateral directions (right).
Mean ± standard deviation, Median (Q1–Q3) and Magnitude of CBCT shifts per organ following offline review of 115 CBCT images.
| Organ | Translation | Mean ± standard deviation [mm] | Median (Q1–Q3) [mm] | Mean Magnitude ± standard deviation [mm] | Median (Q1–Q3) Magnitude [mm] |
|---|---|---|---|---|---|
| Breast | Vertical | −0.7 ± 1.9 | 0 (−2.0–0.0) | 2.3 ± 2.4 | 2.0 (0.0–3.5) |
| Longitudinal | 0.2 ± 2.3 | 0 (0.0–0.0) | |||
| Lateral | 0.1 ± 1.5 | 0 (0.0–1.0) | |||
| Heart | Vertical | −0.2 ± 2.0 | 0 (−1.0–1.0) | 2.9 ± 2.0 | 2.3 (1.4–4.2) |
| Longitudinal | −0.4 ± 2.3 | 0 (−1.0–0.0) | |||
| Lateral | 0.1 ± 1.8 | 0 (−1.0–1.0) | |||
| Diaphragmatic Domes | Longitudinal | −2.5 ± 5.7 | −2.0 (−6.0–1.0) | 4.5 ± 4.0 | 4.0 (1.0–7.0) |
| Spine | Vertical | 0.5 ± 2.0 | 0 (−1.0–1.0) | 3.3 ± 2.2 | 3.0 (2.0–4.2) |
| Longitudinal | −0.1 ± 2.8 | 0 (−1.0–1.0) | |||
| Lateral | 0.5 ± 1.9 | 0 (0.0–1.0) | |||
| Sternum | Vertical | −0.1 ± 1.9 | 0 (−1.0–1.0) | 3.2 ± 2.4 | 2.4 (1.4–4.6) |
| Longitudinal | −0.4 ± 3.1 | 0 (−2.0–1.0) | |||
| Lateral | 0.0 ± 1.8 | 0 (−1.0–1.0) | |||
| All | Vertical | −0.5 ± 2.4 | −0.5 (−2.0–1.0) | 3.4 ± 2.5 | 3.0 (1.4–4.6) |
| Longitudinal | 0.4 ± 2.9 | 0 (−1.0–1.8) | |||
| Lateral | 0.3 ± 1.7 | 0 (−1.0–1.0) |
Mean ± standard deviation of setup, imaging and treatment times extracted from Aria, manually recorded by the therapists or exported from the SGRT treatment reports.
| Patient | Session duration (min) Aria logs | Setup time (min) | Monitoring duration – ceiling cameras (min) | Setup duration – InBore cameras (min) | CBCT BH duration (s) | CBCT analysis duration (s) | End CBCT to beginning BH treatment (s) | Treatment duration – InBore cameras (min) | Monitoring duration – InBore cameras (min) | Number of treatment BH |
|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 17.4 ± 3.1 | 2.5 ± 1.5 | 1.7 ± 1.0 | 32 ± 4 | 139 ± 45 | 4.0 ± 0.4 | 8.6 ± 1.6 | 4 ± 0 | ||
| 2 | 17.2 ± 3.4 | 2.2 ± 1.3 | 1.5 ± 0.7 | 30 ± 2 | 122 ± 19 | 4.4 ± 0.5 | 8.3 ± 1.2 | 4 ± 0 | ||
| 3 | 14.6 ± 2.8 | 2.0 ± 1.3 | 1.1 ± 0.4 | 27 ± 2 | 101 ± 21 | 4.1 ± 1.3 | 7.3 ± 1.2 | 4 ± 0 | ||
| 4 | 14.0 ± 4.3 | 3.1 ± 1.1 | 1.2 ± 0.4 | 30 ± 3 | 117 ± 31 | 3.6 ± 1.5 | 7.2 ± 1.6 | 4 ± 0 | ||
| 5 | 14.8 ± 4.5 | 2.6 ± 1.5 | 1.5 ± 0.9 | 30 ± 5 | 130 ± 41 | 4.6 ± 2.1 | 8.6 ± 3.0 | 4 ± 0 | ||
| 6 | 15.0 ± 3.1 | 2.2 ± 1.1 | 1.4 ± 0.6 | 31 ± 3 | 118 ± 40 | 3.5 ± 1.2 | 7.4 ± 1.6 | 4 ± 0 | ||
| 7 | 14.9 ± 5.7 | 2.7 ± 2.5 | 1.4 ± 0.6 | 29 ± 3 | 98 ± 35 | 2.4 ± 1.1 | 5.9 ± 1.0 | 3 ± 0 | ||
| 8 | 16.1 ± 3.8 | 2.8 ± 0.9 | 1.2 ± 0.4 | 32 ± 3 | 118 ± 46 | 4.6 ± 1.6 | 8.4 ± 2.1 | 3 ± 1 | ||
| 9 | 12.7 ± 3.5 | 2.6 ± 0.9 | 1.4 ± 0.4 | 29 ± 3 | 106 ± 43 | 2.5 ± 0.8 | 5.9 ± 1.0 | 3 ± 0 | ||
| 10 | 12.9 ± 2.0 | 2.2 ± 1.0 | 1.1 ± 0.4 | 28 ± 5 | 125 ± 45 | 3.4 ± 1.0 | 7.0 ± 1.1 | 4 ± 1 | ||
| 11 | 10.7 ± 1.8 | 1.9 ± 0.7 | 1.1 ± 0.2 | 27 ± 3 | 108 ± 32 | 2.2 ± 0.5 | 5.5 ± 0.8 | 3 ± 0 | ||
| 12 | 14.5 ± 2.7 | 3.4 ± 0.9 | 1.3 ± 0.8 | 29 ± 3 | 113 ± 35 | 3.3 ± 0.6 | 7.0 ± 1.2 | 3 ± 0 | ||
| 13 | 12.6 ± 2.7 | 2.6 ± 0.9 | 1.4 ± 0.5 | 29 ± 2 | 108 ± 44 | 3.5 ± 0.9 | 7.1 ± 1.2 | 4 ± 0 | ||
| 14 | 12.7 ± 3.0 | 2.4 ± 0.9 | 1.2 ± 0.5 | 28 ± 3 | 117 ± 30 | 3.5 ± 1.3 | 7.0 ± 1.7 | 4 ± 0 | ||
| 15 | 12.2 ± 2.1 | 2.1 ± 0.9 | 1.1 ± 0.4 | 29 ± 3 | 113 ± 38 | 3.1 ± 0.8 | 6.4 ± 1.0 | 3 ± 1 | ||
| 16 | 13.6 ± 2.3 | 2.7 ± 1.2 | 6.2 ± 1.2 | 29 ± 2 | 121 ± 23 | 2.6 ± 0.5 | 6.2 ± 1.2 | 4 ± 0 | ||
| 17 | 13.9 ± 2.5 | 3.6 ± 1.5 | 1.3 ± 0.4 | 29 ± 4 | 119 ± 29 | 3.7 ± 0.8 | 7.5 ± 1.1 | 4 ± 0 | ||
| 18 | 11.8 ± 2.4 | 2.9 ± 1.5 | 0.9 ± 0.2 | 29 ± 3 | 123 ± 23 | 2.7 ± 1.1 | 6.0 ± 1.3 | 3 ± 1 | ||
| 19 | 15.9 ± 3.8 | 4.4 ± 2.0 | 1.2 ± 0.3 | 29 ± 3 | 122 ± 48 | 4.4 ± 3.0 | 8.1 ± 2.7 | 3 ± 1 | ||
| 20 | 14.4 ± 5.7 | 2.8 ± 2.9 | 1.7 ± 2.3 | 27 ± 3 | 114 ± 34 | 3.6 ± 1.4 | 7.6 ± 3.1 | 4 ± 1 | ||
| 21 | 17.6 ± 4.3 | 3.8 ± 1.8 | 1.7 ± 1.1 | 28 ± 2 | 143 ± 33 | 4.7 ± 1.3 | 9.2 ± 1.8 | 4 ± 1 | ||
| 22 | 12.4 ± 2.3 | 2.9 ± 1.1 | 1.0 ± 0.2 | 27 ± 2 | 117 ± 40 | 2.7 ± 0.5 | 6.2 ± 0.7 | 4 ± 1 |