Commissioning and performance testing of the first prototype of AlignRT InBore™ a Halcyon™ and Ethos™-dedicated surface guided radiation therapy platform.

Daniel Nguyen1, Jad Farah2, Nicolas Barbet3, Mustapha Khodri3.   

Abstract

PURPOSE: To commission and assess the performance of AlignRT InBore™, a Halcyon™ and Ethos™-dedicated Surface Guided Radiation Therapy (SGRT) platform which combines ceiling-mounted cameras for patient setup and bore-mounted cameras for in-bore tracking.
METHODS: To check the potential impact of InBore™ cameras on dose delivery, 16 SRS, H&N, breast and pelvis patients' quality assurance (QA) treatment plans were measured with/without AlignRT InBore™ and using ArcCHECK® and SRS MapCHECK®. Impact on image quality was determined using Catphan® 540 phantom and considering all available MV and CBCT protocols (head, breast, chest and pelvis). The stability, accuracy and overall performance of AlignRT InBore™ was assessed using an MV Cube and anthropomorphic phantoms.
RESULTS: Comparison of 2D dose distributions with/without AlignRT InBore™ showed no impact on treatment delivery for all 16 QA checks (p-value > 0.25). 2D and CBCT images showed no artefacts or change in the contrast-to-noise ratio, resolution and noise values measured with Catphan® 540. Anti-collision sensors were unaffected by the bore-mounted cameras. Additionally, AlignRT InBore™ cameras allowed for motion detection with sub-0.5 mm accuracy and sub-0.4 mm stability with surface coverage of >50 × 60 × 35 cc. Accurate transition (sub-0.3 mm) from virtual to treatment isocentres was achieved. Finally, Halcyon™ rotations during CBCT and beam delivery resulted in limited camera vibrations with translation uncertainty <0.5 mm in left-right and anterior-posterior directions and <0.1 mm in head-feet direction.
CONCLUSION: AlignRT InBore™ provides SGRT setup and intrafraction monitoring capabilities with a performance comparable to standard SGRT solutions while having no adverse effect on Halcyon™.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Halcyon™; Motion management; Setup verification; Surface guidance

Mesh:

Year:  2020        PMID: 33189046     DOI: 10.1016/j.ejmp.2020.10.024

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  5 in total

1.  Reproducibility of Deep-Inspiration Breath Hold treatments on Halcyon™ performed using the first clinical version of AlignRT InBore™: Results of CYBORE study.

Authors:  F Lorchel; D Nguyen; A Mamou; N Barbet; J Camoesas; Y Degluaire; I Pouchard; Y Queffelec; F Renoult; M Khodri; J Farah
Journal:  Clin Transl Radiat Oncol       Date:  2022-05-18

2.  Treatment planning comparison of volumetric modulated arc therapy with the trilogy and the Halcyon for bilateral breast cancer.

Authors:  Tao Sun; Xiutong Lin; Guifang Zhang; Qingtao Qiu; Chengqiang Li; Yong Yin
Journal:  Radiat Oncol       Date:  2021-02-18       Impact factor: 3.481

3.  Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery.

Authors:  Lei Zhang; Sarath Vijayan; Sheng Huang; Yulin Song; Tianfang Li; Xiang Li; Elizabeth Hipp; Maria F Chan; Hsiang-Chi Kuo; Xiaoli Tang; Grace Tang; Seng Boh Lim; Dale Michael Lovelock; Ase Ballangrud; Guang Li
Journal:  J Appl Clin Med Phys       Date:  2021-03-29       Impact factor: 2.102

4.  Intra-fraction motion monitoring during fast modulated radiotherapy delivery in a closed-bore gantry linac.

Authors:  Laurence Delombaerde; Saskia Petillion; Caroline Weltens; Tom Depuydt
Journal:  Phys Imaging Radiat Oncol       Date:  2021-10-30

5.  Analysis of a surface imaging system using a six degree-of-freedom couch.

Authors:  Xiaodong Zhao; Elizabeth L Covington; Richard A Popple
Journal:  J Appl Clin Med Phys       Date:  2022-07-12       Impact factor: 2.243

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.