Gianluca Verona Rinati1, Giuseppe Felici2, Federica Galante2, Alessia Gasparini3,4, Rafael Kranzer5,6, Giulia Mariani2, Matteo Pacitti2, Giuseppe Prestopino1, Andreas Schüller7, Verdi Vanreusel3,4, Dirk Verellen3,4, Claudio Verona1, Marco Marinelli1.
Abstract
PURPOSE: A diamond detector prototype was recently proposed by Marinelli et al. (Medical Physics 2022, https://doi.org/10.1002/mp.15473) for applications in ultrahigh-dose-per-pulse (UH-DPP) and ultrahigh-dose-rate (UH-DR) beams, as used in FLASH radiotherapy (FLASH-RT). In the present study, such so-called flashDiamond (fD) was investigated from the dosimetric point of view, under pulsed electron beam irradiation. It was then used for the commissioning of an ElectronFlash linac (SIT S.p.A., Italy) both in conventional and UH-DPP modalities.Entities:
Keywords: FLASH radiotherapy; diamond detector; dosimetry
Mesh:
Substances:
Year: 2022 PMID: 35652248 PMCID: PMC9543846 DOI: 10.1002/mp.15782
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.506
Main features of the ElectronFlash linac used in this study
| Parameter | Range | |||
|---|---|---|---|---|
| Beam energy | 7 and 9 MeV | |||
| Pulse duration | from 0.5 to 4.0 μs | |||
| Pulse repetition frequency | from single pulse to 500 Hz | |||
| Applicators (mm) | Ext. | Int. | SSDa (cm) | DPPmax (Gy) |
| 30 | 40 | 30 | 71.8 | 11.4 |
| 35 | 40 | 36 | 71.8 | 11.9 |
| 40 | 50 | 40 | 78.6 | 10.6 |
| 50 | 60 | 50 | 81.8 | 9.2 |
| 100 | 110 | 100 | 109.8 | 4.1 |
| 120 | 134 | 124 | 126.8 | 2.7 |
Note: The DPPmax values were obtained by EBT‐XD films, positioned at d max, under 9 MeV irradiation, with the phantom in close contact with the applicator.
aSource‐to‐surface distance.
FIGURE 1Charge per pulse (QPP) of the flashDiamond (fD) detector as a function of the dose‐per‐pulse (DPP) measured by EBT‐XD films. The DPP was varied by changing the PMMA applicator (a) and the pulse duration (b). The red dashed lines are the linear best fits to the experimental data. Deviations from linearity plots are also reported in both cases.
FIGURE 2Percentage depth doses (PDDs) measured by the flashDiamond (fD) prototype and reference detectors in the case of the 100 mm PMMA applicator and 9 MeV beam energy, both in conventional (a) and ultrahigh‐dose‐per‐pulse (UH‐DPP) (b) modalities
R 90 and R 50 of the 9 MeV beams in conventional and ultrahigh‐dose‐per‐pulse (UH‐DPP) modalities, calculated from the percentage depth dose (PDDs) measured with the 100m applicator by the flashDiamond (fD), microDiamond (mD), Advanced Markus, EBT‐XD and silicon diode (SiD) detectors
| Conventional | UH‐DPP | |||
|---|---|---|---|---|
| Detector |
|
|
|
|
| fD | 21.0 | 31.3 | 24.5 | 33.8 |
| EBT‐XD | 20.9 | 31.2 | 25.0 | 34.3 |
| mD | 20.8 | 31.1 | ||
| SiD | 21.2 | 31.2 | ||
| Adv. Markus | 20.8 | 30.9 | ||
Note: All data are expressed in mm.
FIGURE 3Beam profiles measured by the flashDiamond (fD) prototype and reference detectors in the case of the 100 and 40 mm PMMA applicators, 9 MeV beam energy, both in conventional (a) and ultrahigh‐dose‐per‐pulse (UH‐DPP) (b) modalities
Average 80%–20% penumbra, field size (FS) and flatness of the 9 MeV conventional and ultrahigh‐dose‐per‐pulse (UH‐DPP) beams calculated from the profiles with 100 and 40 mm applicators measured by the flashDiamond (fD), silicon diode (SiD) and EBT‐XD
| Conventional | ||||||
|---|---|---|---|---|---|---|
| 100 mm | 40 mm | |||||
| Detector | Penumbra (mm) | FS (mm) | Flatness (%) | Penumbra (mm) | FS (mm) | Flatness (%) |
| fD | 4.7 | 101.5 | 3.0 | 5.1 | 41.1 | 5.2 |
| SiD | 4.7 | 101.4 | 2.8 | 5.8 | 40.7 | 6.9 |
FIGURE 4Output factors (OFs) measured by the flashDiamond (fD) prototype and reference detectors by using all the applicators and the beam energies, both in conventional (a) and ultrahigh‐dose‐per‐pulse (UH‐DPP) (b) modalities
Output factors (OFs) measured by the flashDiamond (fD) prototype and reference detectors by using all the applicators and the beam energies, both in conventional and ultrahigh‐dose‐per‐pulse (UH‐DPP) modalities
| Conventional | UH‐DPP | |||||||
|---|---|---|---|---|---|---|---|---|
| 7 MeV | 9 MeV | 7 MeV | 9 MeV | |||||
|
| fD | SiD | fD | SiD | fD | EBT‐XD | fD | EBT‐XD |
| 30 | 1.89 | 1.91 | 2.03 | 2.04 | 2.56 | 2.48 | 2.90 | 2.81 |
| 35 | 2.02 | 2.02 | 2.11 | 2.12 | 2.66 | 2.74 | 2.91 | 2.94 |
| 40 | 1.81 | 1.81 | 1.89 | 1.93 | 2.35 | 2.40 | 2.59 | 2.62 |
| 50 | 1.68 | 1.70 | 1.74 | 1.74 | 2.10 | 2.14 | 2.23 | 2.26 |
| 100 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 |
| 120 | 0.77 | 0.76 | 0.76 | 0.74 | 0.68 | 0.70 | 0.67 | 0.66 |
R 90 and R 50 of the 7 and 9 MeV conventional and ultrahigh‐dose‐per‐pulse (UH‐DPP) beams extracted from the percentage depth dose (PDDs) measured by the flashDiamond (fD) with different applicators
| 7 MeV | 9 MeV | |||||||
|---|---|---|---|---|---|---|---|---|
| Conventional | UH‐DPP | Conventional | UH‐DPP | |||||
|
|
|
|
|
|
|
|
|
|
| 30 | 15.9 | 23.7 | 18.7 | 26.0 | 20.2 | 29.9 | 23.1 | 32.9 |
| 35 | 16.9 | 24.7 | 18.2 | 25.4 | 21.4 | 31.2 | 24.1 | 33.3 |
| 40 | 17.1 | 24.7 | 19.2 | 26.2 | 21.9 | 31.5 | 25.1 | 34.3 |
| 50 | 17.3 | 25.0 | 18.3 | 25.5 | 22.5 | 31.9 | 25.1 | 33.9 |
| 100 | 15.7 | 23.8 | 17.8 | 24.9 | 21.0 | 31.3 | 24.5 | 33.8 |
| 120 | 14.8 | 23.0 | 17.3 | 24.7 | 19.5 | 30.0 | 24.2 | 33.4 |
Note: All data are expressed in mm.
Average values of the 80%–20% penumbras, field sizes (FSs), and flatness of the 7 and 9 MeV conventional and ultrahigh‐dose‐per‐pulse (UH‐DPP) beams calculated from the profiles measured by the flashDiamond (fD) using applicators of different diameters
| 7 MeV | 9 MeV | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Conventional | UH‐DPP | Conventional | UH‐DPP | |||||||||
|
| Penumbra (mm) | FS (mm) | Flatness (%) | Penumbra (mm) | FS (mm) | Flatness (%) | Penumbra (mm) | FS (mm) | Flatness (%) | Penumbra (mm) | FS (mm) | Flatness (%) |
| 30 | 5.4 | 30.9 | 9.5 | 6.0 | 30.8 | 11.6 | 4.8 | 31.2 | 7.9 | 6.1 | 31.2 | 10.7 |
| 35 | 5.2 | 37.2 | 6.2 | 5.8 | 37.0 | 8.9 | 5.2 | 37.5 | 5.8 | 6.5 | 37.3 | 9.1 |
| 40 | 5.2 | 40.8 | 5.8 | 5.9 | 40.5 | 7.6 | 5.1 | 41.1 | 5.2 | 6.7 | 40.9 | 8.6 |
| 50 | 5.0 | 51.2 | 2.8 | 4.6 | 51.2 | 2.8 | 5.7 | 51.5 | 3.6 | 7.3 | 51.4 | 7.4 |
| 100 | 4.3 | 101.2 | 2.3 | 5.3 | 101.0 | 1.1 | 4.7 | 101.5 | 3.0 | 7.2 | 101.0 | 3.4 |
| 120 | 4.0 | 126.8 | 2.2 | 5.3 | 126.6 | 1.7 | 4.5 | 127.2 | 2.9 | 7.6 | 126.4 | 3.8 |
FIGURE 5Percentage depth dose (PDDs) measured by the flashDiamond (fD) prototype irradiated in ultrahigh‐dose‐per‐pulse (UH‐DPP) modality, by using all the applicators and beam energies, in the case of 9 MeV (a) and 7 MeV (b)
FIGURE 6Beam profiles measured by the flashDiamond (fD) prototype irradiated in ultrahigh‐dose‐per‐pulse (UH‐DPP) modality, by using all the applicators and beam energies, in the case of 9 MeV (a) and 7 MeV (b)