| Literature DB >> 35645506 |
Suleman Abbas Khan1, Rinee Khanna1, Saumya Navit1, Seema Jabeen1, Nishi Grover1, Saurabh Pramanik2.
Abstract
Aim and objective: To compare the accuracy of radiovisiography (RVG), Root ZX mini Apex locator, and Endo radar's inbuilt Apex Locator during working length determination determination in primary teeth with the actual/direct canal length. Materials and methods: 58 primary teeth indicated for extraction in children of age group 4-12 years were selected. RVG, Root ZX mini Apex Locator, and Endo radar's Apex Locator were used to determine the working length intraorally. Following extraction of the teeth, the actual working length was determined with magnifying loupes (2.5×) using a K-file and an endodontic ruler. The data was tabulated and statistically analyzed.Entities:
Keywords: Apex locators; Digital radiography; Integrated endomotor; Pediatric endodontics; Primary teeth; Root ZX mini; Root canal working length; Working Length Determination
Year: 2022 PMID: 35645506 PMCID: PMC9108817 DOI: 10.5005/jp-journals-10005-2123
Source DB: PubMed Journal: Int J Clin Pediatr Dent ISSN: 0974-7052
Fig. 1Difference in canal measurement using different techniques as compared to actual length
Significance of difference in canal measurement using different techniques as compared to actual length (paired “t”-test)
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|
|
|
|
| ||||
| 1. | Digital radiography | 0.88 | 0.79 | 6.7 | 10.27 | < 0.001 |
| 2. | Root ZX mini EAL | −0.02 | 0.12 | −0.1 | 1.35 | 0.18 |
| 3. | Endo radar EAL | −0.36 | 0.46 | −2.7 | 7.07 | < 0.001 |
Range of difference from actual canal length using different techniques
|
|
|
|
|
| |||
|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
| ||
| 1. | 0-No difference | 6 | 7.1 | 79 | 94 | 46 | 54.8 |
| 2. | ±0.5 mm | 28 | 33.3 | 5 | 6 | 17 | 20.2 |
| 3. | ±1.0 mm | 32 | 38.2 | 0 | 0 | 20 | 23.8 |
| 4. | ±1.5 mm | 10 | 11.9 | 0 | 0 | 0 | 0 |
| 5. | ±2 mm or more | 8 | 9.5 | 0 | 0 | 1 | 1.2 |
| Significance of difference among different techniques (Friedman's test) (Non-parametric ANOVA) | χ2= 118.5; p < 0.001 | ||||||
(EAL, electronic apex locator)