Literature DB >> 19134039

Accuracy of an electronic apex locator in primary teeth with root resorption.

O Angwaravong1, P Panitvisai.   

Abstract

AIM: To evaluate whether root resorption of primary molar teeth affects the accuracy of the Root ZX apex locator and to compare the Root ZX at different meter readings with direct canal measurement.
METHODOLOGY: Sixty extracted primary molar teeth with root resorption affecting one sixth to one third of root length were used. The teeth were embedded in an alginate model. A K-type file was used in association with a Root ZX apex locator to measure canal length. Measurements were recorded using the Root ZX meter reading 'Apex' and '0.5 bar'. Actual tooth length was measured with a K-file to the major foramen. All measurements were read under a stereomicroscope at 15x magnification. The deviation of the Root ZX measurement from the actual canal length was determined.
RESULTS: Mean differences between Root ZX length meter reading 'Apex' and actual length were 0.01 +/- 0.23 mm whereas mean differences between Root ZX length meter reading '0.5 bar' and actual length were -0.33 +/- 0.30 mm. The Root ZX was 96.7% accurate to within +/- 0.5 mm of the apical foramen when compared with the actual canal length of primary molars with root resorption.
CONCLUSIONS: Using a criterion of +/- 0.5 mm, the accuracy of the Root ZX was high and not affected by root resorption. When compared with direct canal measurement, the error in locating the apical foramen was smaller with measurement at meter reading 'Apex' than meter reading '0.5 bar'.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19134039     DOI: 10.1111/j.1365-2591.2008.01476.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  16 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study.

Authors:  A Dandempally; R Muppa; M K Duddu; P Bhupatiraju; S Nallanchakrava
Journal:  Eur Arch Paediatr Dent       Date:  2013-06-18

3.  Evaluation of the multifrequency electronic apex locator Joypex 5 in primary teeth.

Authors:  E J N L Silva; D R Herrera; E J Souza-Júnior; T P Rosa
Journal:  Eur Arch Paediatr Dent       Date:  2013-07-24

4.  In vitro comparison between apex locators, direct and radiographic techniques for determining the root canal length in primary teeth.

Authors:  L Z Adriano; J C Barasuol; M Cardoso; M Bolan
Journal:  Eur Arch Paediatr Dent       Date:  2019-01-02

5.  Evaluation of the Working Length Determination Accuracy by Cone-beam Computed Tomography in Primary Teeth.

Authors:  John Shibin; Prathima Gs; Suganya M; Nandhakumar S; Sanguida Adimoulame; Kavitha M
Journal:  Int J Clin Pediatr Dent       Date:  2022

6.  The effect of sonic activation of irrigant on postoperative pain after root canal treatment in primary molar teeth: a randomized, clinical study.

Authors:  Hüsniye Gümüş; Ebru Delikan
Journal:  Clin Oral Investig       Date:  2020-11-12       Impact factor: 3.573

7.  Comparative evaluation of root ZX Mini® apex locator and radiovisiography in determining the working length of primary molars: An In Vivo study.

Authors:  Rucha N Davalbhakta; Niraj S Gokhale; Shivayogi M Hugar; Chandrashekhar M Badakar; A Gowtham; Sanjana P Soneta
Journal:  J Oral Biol Craniofac Res       Date:  2021-02-17

8.  Accuracy of Root ZX in teeth with simulated root perforation in the presence of gel or liquid type endodontic irrigant.

Authors:  Hyeong-Soon Shin; Won-Kyung Yang; Mi-Ri Kim; Hyun-Jung Ko; Kyung-Mo Cho; Se-Hee Park; Jin-Woo Kim
Journal:  Restor Dent Endod       Date:  2012-08-29

9.  Accuracy of an Electronic Apex Locator for Working Length Determination in Primary Anterior Teeth.

Authors:  Zahra Bahrololoomi; Ali Asghar Soleymani; Jalil Modaresi; Mahsa Imanian; Malihe Lotfian
Journal:  J Dent (Tehran)       Date:  2015-04

10.  A smart rotary technique versus conventional pulpectomy for primary teeth: A randomized controlled clinical study.

Authors:  Negar Mokhtari; Alireza-Sarraf Shirazi; Masoumeh Ebrahimi
Journal:  J Clin Exp Dent       Date:  2017-11-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.