| Literature DB >> 35628943 |
Murat Dogru1, Cem Simsek1,2, Takashi Kojima1, Naohiko Aketa1, Kazuo Tsubota1,3, Jun Shimazaki4.
Abstract
The purpose of this study was to investigate the possible effects of the noncontact air puff tonometry (NCT) and Icare rebound tonometry (ICT) on the tear film stability by using the tear stability analysis system (TSAS) and dry eye parameters. Fifteen eyes from fifteen normal healthy subjects were investigated in this study. All subjects underwent TSAS surface regularity index (SRI) examinations, TBUT, and IOP measurements. The mean IOP results measured with NCT were 13.3 ± 1.86 mm Hg, and the mean IOP results measured with ICT were 15.88 ± 3.09 mm Hg (p > 0.05). The mean values of baseline, 5 min, and 10 min of the NCT-SRI and ICR-SRI were tested. There were statistically significant differences between NCT-Baseline SRI, NCT-5 min SRI, and NCT-10 min SRI values (p < 0.05). SRI values significantly increased after NCT. The mean values of the baseline, 5 min, and 10 min of the ICT-SRI were also assessed. There were no statistically significant differences between ICT-Baseline SRI, ICT-5 min SRI, and ICT-10 min SRI values (p > 0.05). The mean TBUT values exhibited a significant decrease at 1 min, 5 min, and 10 min compared with baseline values for the NCT and ICT (p < 0.01). NCT-TBUT and ICT-TBUT values were also compared with each other in the same time period. There were no statistically significant differences between NCT-Baseline and ICT-Baseline TBUT values (p > 0.05). In conclusion, intraocular pressure measurements in routine ophthalmology clinical practices by either NCT or ICT cause deterioration in the tear film stability which might affect tear stability testing when performed soon after IOP measurements. It is best to wait at least for 20-30 min after the IOP measurement before evaluating the tear film and the corneal surface or perform tonometry after the tear film-ocular surface evaluation tests.Entities:
Keywords: IOP measurement; dry eye; tear stability; tonometer
Year: 2022 PMID: 35628943 PMCID: PMC9147104 DOI: 10.3390/jcm11102819
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.964
SRI of subjects before and after IOP measurements. Data are presented as the mean ± SD. SRI, surface regularity index. The time-wise variation in SRI from baseline to 5 and 10 min was observed to be significantly higher for NCT (noncontact tonometry). However, there was no significant difference in terms of SRI for the ICT (Icare tonometry). * p < 0.05.
| The Mean SRI Values | Baseline | 5 min | 10 min |
|---|---|---|---|
|
| 0.96 ± 0.62 | 1.29 ± 0.72 * | 1.25 ± 0.62 * |
|
| 1.04 ± 0.49 | 1.07 ± 0.56 | 1.03 ± 0.77 |
Figure 1The mean TBUTs showed a significant decrease at 1 min, 5 min and 10 min compared with baseline values for the ICT (p < 0.01). NCT-TBUT and ICT-TBUT values are also comparable with each other in the same time period. There were no statistically significant differences between NCT-Baseline and ICT-Baseline TBUT values (p > 0.05). Statistically significant differences were found at 1 min, 5 min and 10 min TBUT values between NCT and ICT (p < 0.05). * represents p < 0.05.
Figure 2The mean change in TBUT from baseline to 10 min after tonometry (ΔTBUT) was significantly lesser for ICT measurements compared with NCT measurements (p < 0.05). * represents p < 0.05.