Xueting Cai1,2, Yi Qin1, Sixiu Liu1, Zhewei Cheng1, Fan Lu1, Jia Qu1, Ding Chen1.
Abstract
Purpose: To evaluate the influence of intraocular pressure (IOP) measurement while wearing bandage contact lens (BCL) and the effect of Pentacam and Corvis ST's correction systems.Entities:
Year: 2022 PMID: 35586593 PMCID: PMC9110209 DOI: 10.1155/2022/4933555
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.974
IOP values before and after BCL wearing (n = 40).
| IOP (mmHg) | Before BCL wearing | After BCL wearing |
|
|
|---|---|---|---|---|
| IOPt | 14.8 ± 3.2 | 15.7 ± 3.4 | −0.9 ± 2.9 | 0.04 |
| IOPc | 13.7 ± 2.8 | 15.0 ± 4.0 | −1.3 ± 2.4 | 0.002 |
| IOPe | 15.0 ± 3.1 | 15.0 ± 3.3 | 0.1 ± 2.7 | 0.87 |
| IOPs | 15.2 ± 2.8 | 15.5 ± 2.9 | −0.3 ± 2.7 | 0.464 |
| IOPd | 15.1 ± 2.8 | 15.5 ± 2.9 | −0.4 ± 2.7 | 0.318 |
| IOPk | 15.4 ± 3.0 | 16.2 ± 3.1 | −0.8 ± 2.8 | 0.072 |
| IOPo | 15.1 ± 3.0 | 15.3 ± 2.9 | −0.2 ± 3.0 | 0.624 |
| IOPce | 13.5 ± 2.6 | 13.0 ± 3.1 | 0.5 ± 2.5 | 0.25 |
| BIOP | 11.6 ± 2.0 | 12.0 ± 2.7 | −0.4 ± 2.1 | 0.286 |
BCL: bandage contact lens; NCT: noncontact tonometer; IOPc: Corvis ST IOP values; IOPt: NCT IOP values; IOPe: NCT IOP values corrected by Ehlers formula; IOPs: NCT IOP values corrected by Shah formula; IOPd: NCT IOP values corrected by Dreaden formula; IOPk: NCT IOP values corrected by Kohlhaas formula; IOPo: NCT IOP values corrected by Orssengo/Pye formula; IOPce: Corvis ST IOP values corrected by Ehlers formula; BIOP: biomechanical corrected IOP values by Corvis ST.
Figure 1The Bland–Altman plots for the comparison between the IOP values obtained by noncontact tonometer and Corvis ST before BCL wearing. IOP values from noncontact tonometer was higher than corvis ST before BCL wearing (+1.1 mmHg).
Figure 2The Bland–Altman plots for the comparison between the IOP values obtained by noncontact tonometer and Corvis ST after BCL wearing. IOP values from noncontact tonometer was higher than Corvis ST after BCL wearing (+0.7 mmHg).
Comparison between IOP values obtained before BCL wearing and corrected IOP values after BCL wearing (n = 40).
|
|
| |
|---|---|---|
| IOPt1–IOPe2 | −0.2 ± 4.1 | 0.736 |
| IOPt1–IOPs2 | −0.7 ± 3.5 | 0.212 |
| IOPt1–IOPd2 | −0.7 ± 3.2 | 0.166 |
| IOPt1–IOPk2 | −1.4 ± 2.9 | 0.004 |
| IOPt1–IOPo2 | −0.5 ± 3.3 | 0.342 |
| IOPc1–IOPce2 | +0.7 ± 3.2 | 0.179 |
| IOPc1–BIOP2 | +1.7 ± 2.1 | ≤0.001 |
BCL: bandage contact lens; NCT: noncontact tonometer; IOPt1: NCT IOP values before BCL wearing; IOPt2: NCT IOP values after BCL wearing; IOPe2: NCT IOP values after BCL wearing corrected by Ehlers formula; IOPs2: NCT IOP values after BCL wearing corrected by Shah formula; IOPd2: NCT IOP values after BCL wearing corrected by Dresden formula; IOPk2: NCT IOP values after BCL wearing corrected by Kohlhaas formula; IOPo2: NCT IOP values after BCL wearing corrected by Orssengo/Pye formula; IOPc1: Corvis ST IOP values before BCL wearing; IOPc2: Corvis ST IOP values after BCL wearing; IOPce2: Corvis ST IOP values after BCL wearing corrected by Ehlers formula; and BIOP2:biomechanical corrected Corvis ST IOP values after BCL wearing.