| Literature DB >> 35493850 |
James L Bedford1, Ian M Hanson1.
Abstract
Background and purpose: Real-time portal dosimetry compares measured images with predicted images to detect delivery errors as the radiotherapy treatment proceeds. This work aimed to investigate the performance of a recurrent neural network for processing image metrics so as to detect delivery errors as early as possible in the treatment. Materials and methods: Volumetric modulated arc therapy (VMAT) plans of six prostate patients were used to generate sequences of predicted portal images. Errors were introduced into the treatment plans and the modified plans were delivered to a water-equivalent phantom. Four different metrics were used to detect errors. These metrics were applied to a threshold-based method to detect the errors as soon as possible during the delivery, and also to a recurrent neural network consisting of four layers. A leave-two-out approach was used to set thresholds and train the neural network then test the resulting systems.Entities:
Keywords: Artificial neural network; Electronic portal imaging device; In vivo dosimetry; Volumetric modulated arc therapy
Year: 2022 PMID: 35493850 PMCID: PMC9048084 DOI: 10.1016/j.phro.2022.03.004
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1An analysis of a volumetric modulated arc therapy treatment plan for a patient delivery, seen in AutoDose v1.1. The main panel shows the mean image difference as a percentage of local image intensity for sections of arc consisting of 10 segments. The inset (lower right) shows the expected and actual images for a single section of arc, together with horizontal and vertical profiles through the central axis (Data 1 – expected image, Data 2 – actual image).
Fig. 2Training the recurrent neural network. (a) Network topology, (b) abstraction of one layer of the network, (c) training progress for the nine data sets, (d)-(g) Median index of the first segment at which each error is detected, as a function of error type and magnitude. White cross-hatching indicates that the error is not detected. C: central image signal, M: mean image value, G: root-mean-square error as a percentage of global maximum, L: root-mean-square error as a percentage of local signal, E: error, MSM: multiple separate metrics, RNN: recurrent neural network.
Fig. 3Median index of the first segment at which each error is detected, as a function of error type and magnitude, during testing. White cross-hatching indicates that the error is not detected. MSM: multiple separate metrics; RNN: recurrent neural network.
Fig. 4Index of the first segment at which each error is detected, in the six patients separately, for a fixed level of error, during testing. White cross-hatching indicates that the error is not detected. MSM: multiple separate metrics; RNN: recurrent neural network.
Mean segment index at which errors are detected for multiple separate metrics with threshold and for a recurrent neural network, during testing.
| Patient | Patient | Error Size | MSM | RNN | Relative benefit |
|---|---|---|---|---|---|
| 1 | 4 | Small | 159 | 181 | 1.14 |
| Medium | 129 | 38 | 0.29 | ||
| Large | 78 | 23 | 0.29 | ||
| 1 | 5 | Small | 159 | 105 | 0.66 |
| Medium | 120 | 51 | 0.43 | ||
| Large | 78 | 23 | 0.29 | ||
| 1 | 6 | Small | 159 | 142 | 0.89 |
| Medium | 130 | 60 | 0.46 | ||
| Large | 78 | 23 | 0.29 | ||
| 2 | 4 | Small | 114 | 181 | 1.59 |
| Medium | 84 | 84 | 1.00 | ||
| Large | 40 | 33 | 0.83 | ||
| 2 | 5 | Small | 114 | 151 | 1.32 |
| Medium | 92 | 61 | 0.66 | ||
| Large | 38 | 32 | 0.84 | ||
| 2 | 6 | Small | 115 | 103 | 0.90 |
| Medium | 78 | 77 | 0.99 | ||
| Large | 42 | 24 | 0.57 | ||
| 3 | 4 | Small | 129 | 181 | 1.40 |
| Medium | 131 | 72 | 0.55 | ||
| Large | 59 | 74 | 1.25 | ||
| 3 | 5 | Small | 129 | 181 | 1.40 |
| Medium | 122 | 66 | 0.54 | ||
| Large | 58 | 24 | 0.41 | ||
| 3 | 6 | Small | 129 | 181 | 1.40 |
| Medium | 131 | 80 | 0.61 | ||
| Large | 59 | 24 | 0.41 | ||
MSM: multiple separate metrics; RNN: recurrent neural network.
Small: 2% monitor unit increase, 2 mm aperture opening, 2 mm aperture shift, 10 mm air gap; medium: 4–6% monitor unit increase, 4–6 mm aperture opening, 4–6 mm aperture shift, 20–30 mm air gap; large: 8–10% monitor unit increase, 8–10 mm aperture opening, 8–10 mm aperture shift, 40–50 mm air gap.
Relative benefit defined as quotient of RNN and MSM.
Fig. 5Network output for patient 1 for several error cases. Results less than or equal to zero indicate absence of an error and results greater than zero indicate an error. The output in the grey region at the left is disregarded due to instability of the raw signals.