Site-specific alert criteria to detect patient-related errors with 3D EPID transit dosimetry.

Igor Olaciregui-Ruiz1, Roel Rozendaal1, Ben Mijnheer1, Anton Mans1.   

Abstract

PURPOSE: To assess the sensitivity of various EPID dosimetry alert indicators to patient-related variations and to determine alert threshold values that ensure excellent error detectability.
METHODS: Our virtual dose reconstruction method uses in air EPID measurements to calculate virtual 3D dose distributions within a CT data set. Patient errors are introduced by transforming the plan-CT into an error-CT data set. Virtual patient dose distributions reconstructed using the plan-CT and the error-CT data set are compared to the planned dose distributions by γ(3%/3 mm) and DVH analysis using seven indicators: ΔDISOC , γ-mean, near γ-max, γ-pass rate, ΔPTVD 2 , ΔPTVD 50, and ΔPTVD 98 . Translation and rotation patient setup errors and uniform contour changes are studied for 104 VMAT plans of 4 treatment sites. Lung expansions and contractions to simulate changes in lung density are considered for 26 IMRT lung plans. A ROC curve is generated for each combination of error and indicator. For each ROC curve, the AUC value and the optimal alert threshold value of the indicator are determined.
RESULTS: AUC values for γ-indicators and ΔPTVD 2 are consistently higher than for ΔDISOC and ΔPTVD 98 . For VMAT plans, error detectability to patient position shifts is worse for pelvic treatments and best for head-and-neck and brain plans. Excellent detectability is observed for 5 mm translations in head-and-neck plans (AUC = 0.94) and for 4° rotations in brain plans (AUC = 0.89). All sites but prostate show good-to-excellent detectability (AUC > 0.8) for 10 mm translations and 8° rotations and excellent detectability (AUC > 0.9) for ±6 mm patient contour changes. For head-and-neck, excellent detectability is obtained with γ-mean and γ-pass rate threshold values of around 0.63 and 83%, respectively. For brain and rectum, these threshold values are 0.53 and 90%, respectively. In IMRT lung plans, expansions of 3 mm and contractions of 6 mm are detected (AUC > 0.8).
CONCLUSIONS: By combining virtual dose reconstructions with synthetic patient data, we developed a framework to assess the sensitivity of our 3D EPID transit dosimetry method to patient-related variations. The detectability of each introduced error is specific to the treatment site and indicator used. Optimal alert criteria can be determined to ensure excellent detectability for each combination of error type and indicator. The alert threshold values and the magnitude of the error that can be detected are site-specific. In situations where the minimum error that can be detected is larger than the clinically desirable action level, EPID transit dosimetry must be used in combination with IGRT procedures to ensure correct patient positioning and early detection of anatomy variations.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  ROC analysis; anatomical changes; in vivo EPID dosimetry; patient setup; sensitivity

Mesh:

Year:  2018        PMID: 30372521     DOI: 10.1002/mp.13265

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Augmented reality-guided positioning system for radiotherapy patients.

Authors:  Chunying Li; Zhengda Lu; Mu He; Jianfeng Sui; Tao Lin; Kai Xie; Jiawei Sun; Xinye Ni
Journal:  J Appl Clin Med Phys       Date:  2022-01-05       Impact factor: 2.102

Review 2.  In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Igor Olaciregui-Ruiz; Sam Beddar; Peter Greer; Nuria Jornet; Boyd McCurdy; Gabriel Paiva-Fonseca; Ben Mijnheer; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2020-08-29

3.  Optimisation of a composite difference metric for prompt error detection in real-time portal dosimetry of simulated volumetric modulated arc therapy.

Authors:  James L Bedford; Ian M Hanson
Journal:  Br J Radiol       Date:  2021-03-18       Impact factor: 3.039

4.  Evaluation of Daily CT for EPID-Based Transit In Vivo Dosimetry.

Authors:  Bin Feng; Lei Yu; Enwei Mo; Liyuan Chen; Jun Zhao; Jiazhou Wang; Weigang Hu
Journal:  Front Oncol       Date:  2021-11-02       Impact factor: 6.244

5.  A recurrent neural network for rapid detection of delivery errors during real-time portal dosimetry.

Authors:  James L Bedford; Ian M Hanson
Journal:  Phys Imaging Radiat Oncol       Date:  2022-04-20

6.  Extending in aqua portal dosimetry with dose inhomogeneity conversion maps for accurate patient dose reconstruction in external beam radiotherapy.

Authors:  Igor Olaciregui-Ruiz; Julia-Maria Osinga-Blaettermann; Karen Ortega-Marin; Ben Mijnheer; Anton Mans
Journal:  Phys Imaging Radiat Oncol       Date:  2022-04-14

7.  Reduction of systematic dosimetric uncertainties in volumetric modulated arc therapy triggered by patient-specific quality assurance.

Authors:  Anton Mans; Roel Rozendaal; Tomas Janssen; Eugène Damen; Jochem Kaas; Anke van Mourik; Ben Mijnheer
Journal:  Phys Imaging Radiat Oncol       Date:  2022-01-20

8.  Characterization of automatic treatment planning approaches in radiotherapy.

Authors:  Geert Wortel; Dave Eekhout; Emmy Lamers; René van der Bel; Karen Kiers; Terry Wiersma; Tomas Janssen; Eugène Damen
Journal:  Phys Imaging Radiat Oncol       Date:  2021-07-13
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.