Literature DB >> 20089323

3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry.

Anton Mans1, Peter Remeijer, Igor Olaciregui-Ruiz, Markus Wendling, Jan-Jakob Sonke, Ben Mijnheer, Marcel van Herk, Joep C Stroom.   

Abstract

BACKGROUND AND
PURPOSE: To demonstrate the feasibility of back-projection portal dosimetry for accurate 3D dosimetric verification of volumetric-modulated arc therapy (VMAT), pre-treatment as well as in vivo.
MATERIALS AND METHODS: Several modifications to our existing approach were implemented to make the method applicable to VMAT: (i) gantry angle-resolved data acquisition, (ii) calculation of the patient transmission, (iii) compensation for detector 'flex' and (iv) 3D dose reconstruction and evaluation.
RESULTS: Planned and EPID-(Electronic Portal Image Detector)-reconstructed dose distributions show good agreement for pre-treatment verification of two prostate, a stereotactic lung and a head-and-neck VMAT plan and for in vivo verification of VMAT treatments of prostate and lung cancer. Averaged over pre-treatment verifications, planned and measured isocentre dose ratios were -1.2% (range [-4.7%,1.8%]). 3D gamma analysis (3% maximum dose, 3mm) revealed mean gamma gamma(mean)=0.37 [0.34,0.39], maximum 1% gamma gamma(1%)=0.72 [0.66,0.81] and percentage of points with gamma1 P(gamma)(1)=99% [97%,100%]. For in vivo verification, the average isocentre dose ratio was -1.2% [-0.8%,-1.7%], gamma(mean)=0.52 [0.40,0.64], gamma(1%)=0.92 [0.76,1.08] and P(gamma)(1)=96% [93%,100%].
CONCLUSIONS: Our portal dosimetry method was successfully adapted for verification of VMAT treatments, pre-treatment as well as in vivo. Copyright 2010 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20089323     DOI: 10.1016/j.radonc.2009.12.020

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  37 in total

1.  Comparative analysis of SmartArc-based dual arc volumetric-modulated arc radiotherapy (VMAT) versus intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma.

Authors:  Tsair-Fwu Lee; Pei-Ju Chao; Hui-Min Ting; Su-Hua Lo; Yu-Wen Wang; Chiu-Ching Tuan; Fu-Min Fang; Te-Jen Su
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

2.  Dose verification for respiratory-gated volumetric modulated arc therapy.

Authors:  Jianguo Qian; Lei Xing; Wu Liu; Gary Luxton
Journal:  Phys Med Biol       Date:  2011-07-13       Impact factor: 3.609

3.  A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

Authors:  P Rowshanfarzad; H L Riis; S J Zimmermann; M A Ebert
Journal:  Br J Radiol       Date:  2015-04-23       Impact factor: 3.039

4.  A dual two dimensional electronic portal imaging device transit dosimetry model based on an empirical quadratic formalism.

Authors:  Y I Tan; M Metwaly; M Glegg; S P Baggarley; A Elliott
Journal:  Br J Radiol       Date:  2015-05-13       Impact factor: 3.039

5.  Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy.

Authors:  Aiping Ding; Lei Xing; Bin Han
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

6.  Comparison of 2D and 3D gamma analyses.

Authors:  Kiley B Pulliam; Jessie Y Huang; Rebecca M Howell; David Followill; Ryan Bosca; Jennifer O'Daniel; Stephen F Kry
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

7.  Video-rate optical dosimetry and dynamic visualization of IMRT and VMAT treatment plans in water using Cherenkov radiation.

Authors:  Adam K Glaser; Jacqueline M Andreozzi; Scott C Davis; Rongxiao Zhang; Brian W Pogue; Colleen J Fox; David J Gladstone
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

8.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

9.  Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel.

Authors:  Naoki Hayashi; Ryan L Malmin; Yoichi Watanabe
Journal:  J Radiat Res       Date:  2014-01-20       Impact factor: 2.724

10.  Dose verification of volumetric modulated arc therapy (VMAT) by use of in-treatment linac parameters.

Authors:  Akihiro Haga; Akira Sakumi; Yukari Okano; Saori Itoh; Naoya Saotome; Satoshi Kida; Hiroshi Igaki; Kenshiro Shiraishi; Hideomi Yamashita; Kuni Ohtomo; Keiichi Nakagawa
Journal:  Radiol Phys Technol       Date:  2013-03-12
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.