The effect of the choice of patient model on the performance of in vivo 3D EPID dosimetry to detect variations in patient position and anatomy.

Igor Olaciregui-Ruiz1, Roel Rozendaal1, Simon van Kranen1, Ben Mijnheer1, Anton Mans1.   

Abstract

PURPOSE: In vivo EPID dosimetry is meant to trigger on relevant differences between delivered and planned dose distributions and should therefore be sensitive to changes in patient position and patient anatomy. Three-dimensional (3D) EPID back-projection algorithms can use either the planning computed tomography (CT) or the daily patient anatomy as patient model for dose reconstruction. The purpose of this study is to quantify the effect of the choice of patient model on the performance of in vivo 3D EPID dosimetry to detect patient-related variations.
METHODS: Variations in patient position and patient anatomy were simulated by transforming the reference planning CT images (pCT) into synthetic daily CT images (dCT) representing a variation of a given magnitude in patient position or in patient anatomy. For each variation, synthetic in vivo EPID data were also generated to simulate the reconstruction of in vivo EPID dose distributions. Both the planning CT images and the synthetic daily CT images could be used as patient model in the reconstructions yielding e D pCT and e D dCT EPID reconstructed dose distributions respectively. The accuracy of e D pCT and e D dCT reconstructions was evaluated against absolute dose measurements made in different phantom setups, and against dose distributions calculated by the treatment planning system (TPS). The comparison was performed by γ-analysis (3% local dose/2 mm). The difference in sensitivity between e D pCT and e D dCT reconstructions to detect variations in patient position and in patient anatomy was investigated using receiver operating characteristic analysis and the number of triggered alerts for 100 volumetric modulated arc therapy plans and 12 variations.
RESULTS: e D dCT showed good agreement with both absolute point dose measurements (<0.5%) and TPS data (γ-mean = 0.52 ± 0.11). The agreement degraded with e D pCT , with the magnitude of the deviation varying with each specific case. e D dCT readily detected combined 3 mm translation setup errors in all directions (AUC = 1.0) and combined 3° rotation setup errors around all axes (AUC = 0.86) whereas e D pCT showed good detectability only for 12 mm translations (AUC = 0.85) and 9° rotations (AUC = 0.80). Conversely, e D pCT manifested a higher sensitivity to patient anatomical changes resulting in AUC values of 0.92/0.95 for a 6 mm patient contour expansion/contraction compared to 0.70/0.64 with e D dCT . Using |ΔPTVD50 | > 3% as clinical tolerance level, the percentage of alerts for 6 mm changes in patient contour were 85%/27% with e D pCT / e D dCT .
CONCLUSIONS: With planning CT images as patient model, EPID dose reconstructions underestimate the dosimetric effects caused by errors in patient positioning and overestimate the dosimetric effects caused by changes in patient anatomy. The use of the daily patient position and anatomy as patient model for in vivo 3D EPID transit dosimetry improves the ability of the system to detect uncorrected errors in patient position and it reduces the likelihood of false positives due to patient anatomical changes.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  ROC analysis; in vivo EPID dosimetry; patient position errors; sensitivity

Mesh:

Year:  2019        PMID: 31674038     DOI: 10.1002/mp.13893

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  A new registration algorithm of electronic portal imaging devices images based on the automatic detection of bone edges during radiotherapy.

Authors:  Cheng Chen; Chaoyan Wu; Yahua Zhong; Conghua Xie; Yunfeng Zhou; Hui Liu; Jun Zhang; Jiuling Sheng; Dazheng Jiang; Hongli Zhao; Haijun Yu
Journal:  Sci Rep       Date:  2020-06-24       Impact factor: 4.379

Review 2.  In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Igor Olaciregui-Ruiz; Sam Beddar; Peter Greer; Nuria Jornet; Boyd McCurdy; Gabriel Paiva-Fonseca; Ben Mijnheer; Frank Verhaegen
Journal:  Phys Imaging Radiat Oncol       Date:  2020-08-29

3.  Portal dosimetry in radiotherapy repeatability evaluation.

Authors:  Krzysztof Ślosarek; Dominika Plaza; Aleksandra Nas; Marta Reudelsdorf; Jacek Wendykier; Barbara Bekman; Aleksandra Grządziel
Journal:  J Appl Clin Med Phys       Date:  2020-12-12       Impact factor: 2.102

4.  Evaluation of Daily CT for EPID-Based Transit In Vivo Dosimetry.

Authors:  Bin Feng; Lei Yu; Enwei Mo; Liyuan Chen; Jun Zhao; Jiazhou Wang; Weigang Hu
Journal:  Front Oncol       Date:  2021-11-02       Impact factor: 6.244

5.  A recurrent neural network for rapid detection of delivery errors during real-time portal dosimetry.

Authors:  James L Bedford; Ian M Hanson
Journal:  Phys Imaging Radiat Oncol       Date:  2022-04-20
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.