| Literature DB >> 35459140 |
Chun Yue1, Na Ding1, Lu-Lu Xu1, Ya-Qian Fu1, Yuan-Wei Guo1, Yan-Yi Yang1, Xian-Mei Zhao1, Zhi-Feng Sheng2,3.
Abstract
BACKGROUND: Early detection and timely prophylaxis can retard the progression of osteoporosis. The purpose of this study was to determine the validity of peripheral Dual Energy X-ray Absorptiometry (DXA) test for osteoporosis screening. We examined peripheral bone mineral density (BMD) using AKDX-09 W-I DXA densitometer. Firstly, we acquired BMD data from manufacturer-supplied density-gradient phantoms and 30 volunteers to investigate its accuracy and precision, then we measured BMD for 150 volunteers using both AKDX (left forearm) and Hologic Discovery Wi (left forearm, left hip and L1 - L4 vertebrae) simultaneously. Correlation relationship of BMD results acquired from two instruments was assessed by simple linear regression analysis, the Receiver Operating Characteristic (ROC) curves and Areas Under the Curves (AUCs) were evaluated for the diagnostic value of left forearm BMD measured by AKDX in detecting osteoporosis.Entities:
Keywords: Bone mineral density; Forearm bone densitometry; Osteoporosis; Prescreening
Mesh:
Year: 2022 PMID: 35459140 PMCID: PMC9027342 DOI: 10.1186/s12891-022-05325-6
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Basic Characteristics of the Participants(n = 150, 38 males and 112 females)
| Variables | Mean ± SD ( | Range ( | Mean ± SD ( | Range ( |
|---|---|---|---|---|
| Age | 48.9 ± 15.9 | 20 ~ 84 | 62.0 ± 6.9 | 47 ~ 84 |
| Height (cm) | 158.8 ± 7.2 | 145.0 ~ 184.5 | 157.4 ± 6.8 | 145.0 ~ 173.5 |
| Weight (kg) | 58.5 ± 9.7 | 43.8 ~ 101.5 | 58.5 ± 9.2 | 45.5 ~ 81.9 |
| BMI (kg/cm2) | 23.2 ± 3.0 | 17.7 ~ 30.1 | 23.6 ± 2.8 | 18.5 ~ 30.1 |
| AKDX BMD(g/cm2) | 0.406 ± 0.047 | 0.284 ~ 0.507 | 0.401 ± 0.053 | 0.284 ~ 0.531 |
| Hologic distal 1/3 of radius BMD(g/cm2) | 0.640 ± 0.087 | 0.422 ~ 0.842 | 0.618 ± 0.112 | 0.074 ~ 0.891 |
| Hologic spine L1–L4 BMD(g/cm2) | 0.921 ± 0.145 | 0.560 ~ 1.350 | 0.858 ± 0.156 | 0.556 ~ 1.351 |
| Hologic femoral neck BMD(g/cm2) | 0.735 ± 0.128 | − 3.3 ~ 4.1 | 0.678 ± 0.107 | 0.463 ~ 1.053 |
| Hologic total hip BMD(g/cm2) | 0.881 ± 0.132 | 0.56 ~ 1.35 | 0.834 ± 0.125 | 0.559 ~ 1.156 |
| AKDX BMD T score | − 1.401 ± 0.672 | −3.24 ~ 0.5 | − 1.531 ± 0.720 | −3.24 ~ 0.02 |
| Hologic distal 1/3 of radius BMD T score | −1.429 ± 1.195 | −4.5 ~ 1.6 | −1.854 ± 1.347 | −4.5 ~ 1.4 |
| Hologic spine L1–L4 BMD T score | −0.584 ± 1.257 | − 3.7 ~ 3.2 | − 1.138 ± 1.356 | − 3.7 ~ 3.2 |
| Hologic femoral neck BMD T score | −0.863 ± 1.183 | −3.3 ~ 4.1 | −1.420 ± 0.973 | −3.3 ~ 2.0 |
| Hologic total hip BMD T score | −0.155 ± 1.084 | −2.8 ~ 3.7 | −0.579 ± 0.996 | −2.8 ~ 1.9 |
BMI body mass index, SD standard deviation, BMD bone mineral density
Precision of Measurement--AKDX and Hologic (n = 30)
| BMD Mean Value±SD | RMS-SD (g/cm2) | Precision (CV%) | |||
|---|---|---|---|---|---|
| Low | AKDX | 0.300 ± 0.001 | < 0.01 | 0.033 | 0.40 |
| Hologic | 0.432 ± 0.003 | 0.002 | 0.50 | ||
| Medium | AKDX | 0.492 ± 0.002 | < 0.01 | 0.003 | 0.20 |
| Hologic | 0.671 ± 0.006 | 0.006 | 0.85 | ||
| High | AKDX | 0.582 ± 0.002 | < 0.01 | 0.003 | 0.19 |
| Hologic | 0.886 ± 0.007 | 0.007 | 0.82 | ||
| Healthy volunteers | AKDX | 0.402 ± 0.044 | < 0.01 | 0.019 | 1.65 |
| Hologic | 0.633 ± 0.072 | 0.009 | 1.48 | ||
Low, Medium, and High densities gradient phantom, SD standard deviation, RMS-SD root mean square standard deviation, % CV percent coefficient of variation
Linear correlation coefficients(r) of the dual-energy X-ray absorptiometry (DXA) measurements with AKDX to Hologic
| Region (BMD, g/cm2) | R | Slope | Intercept | S.E.E. (g/cm2) |
|---|---|---|---|---|
| ultradistal radius | 0.777 | 0.536 | 0.187 | 0.031 |
| distal 1/3 of radius | 0.670 | 0.400 | 0.150 | 0.041 |
| Spine L1–L4 | 0.430 | 0.170 | 0.250 | 0.012 |
| Femoral neck | 0.449 | 0.180 | 0.280 | 0.011 |
| Total hip | 0.559 | 0.220 | 0.210 | 0.011 |
R Correlation coefficient, S.E.E Standard errors of the estimate
Fig. 1The scatter plots of AKDX T-scores and DXA T-scores
Fig. 2Receiver operating characteristic curves of AKDX in identifying the risk of osteoporosis with reference to Hologic DXA. A, B Receiver operator characteristic curves on the performance of AKDX (AKDX T score < − 1) in identifying subjects at risk of suboptimal bone health (Hologic DXA T score < − 1), (A) AKDX forearm distal VS Hologic distal 1/3 of radius, (B) AKDX forearm distal VS Hologic any site. C, D Receiver operator characteristic curves on the performance of AKDX (AKDX T score ≤ − 2.5) in identifying subjects at risk of suboptimal bone health (Hologic DXA T score ≤ − 2.5), (C) AKDX forearm distal VS Hologic distal 1/3 of radius, (D) AKDX forearm distal VS Hologic any site. Any site: distal one-third radius, L1–L4 vertebrae, femoral neck and total hip
The performance of AKDX in identifying subjects with suboptimal bone health (Hologic DXA T-score < −1)
| Sen.(%) | Spe.(%) | AUC | 95% CI | ||
|---|---|---|---|---|---|
| distal 1/3 of radius ( | 0.630 | 0.739 | 0.708 | 0.573 ~ 0.843 | 0.004 |
| Any site ( | 0.761 | 0.900 | 0.879 | 0.792 ~ 0.967 | <0.001 |
| AKDX T-scores of −1 (distal 1/3 of radius) | 0.885 | 0.435 | |||
| AKDX T-scores of −1 (Any site) | 0.877 | 0.667 | |||
Significant p-values are italicized. AUC area under curve, CI confidence interval, Sen. sensitivity, Spe. specificity, Any site, any site of the distal 1/3 of radius of left forearm, spine L1 - L4, femoral neck, total hip
The performance of AKDX in identifying subjects with osteoporosis (Hologic DXA T-score ≤ −2.5)
| Sen.(%) | Spe.(%) | AUC | 95% CI | ||
|---|---|---|---|---|---|
| distal 1/3 of radius ( | 0.769 | 0.804 | 0.823 | 0.732 ~ 0.915 | < 0.001 |
| Any site ( | 0.704 | 0.780 | 0.778 | 0.677 ~ 0.880 | < 0.001 |
| AKDX T-scores of −2.5 (distal 1/3 of radius) | 0.348 | 1 | |||
| AKDX T-scores of −2.5 (Any site) | 0.292 | 0.981 | |||
| AKDX T-scores of −1.6 (distal 1/3 of radius) | 0.900 | 0.654 | |||
| AKDX T-scores of −1.2 (Any site) | 0.900 | 0.377 | |||
| AKDX T-scores of −2.0 (distal 1/3 of radius) | 0.327 | 0.900 | |||
| AKDX T-scores of −2.1 (Any site) | 0.375 | 0.900 | |||
Significant p-values are italicized. AUC area under curve, CI confidence interval, Sen. sensitivity, Spe. specificity. Any site, any site of the distal 1/3 of radius of left forearm, spine L1 - L4, femoral neck, total hip