| Literature DB >> 35453777 |
Aleksandra Królikowska1, Anna Mika2, Bartosz Plaskota1, Maciej Daszkiewicz1, Monika Kentel3, Anna Kołcz1, Maciej Kentel3, Robert Prill4, Dorota Diakowska5, Paweł Reichert6, Artur Stolarczyk7, Łukasz Oleksy7,8.
Abstract
The Athletic Shoulder (ASH) test was introduced as a tool for quantifying the ability to produce and transfer force across the shoulder girdle. Whether using the portable isometric-based strength training device Active5™ is a reliable alternative to a gold standard force plate for ASH testing purposes remains unknown; therefore, the present study determined the reliability and validity of Active5™ usage in the ASH test compared to force plates. Fifty-one healthy participants performed the ASH test using Active5™ and K-Force plates in three separate sessions. The maximal force was measured bilaterally in a prone position at three shoulder abduction angles, precisely at 180°, 135°, and 90°. The first rater carried out the first and third sessions, spaced at a one-week interval. A second rater performed the second session. The reliability was assessed using the intraclass correlation coefficient (ICC). The linear Pearson's correlation coefficient (r) calculation was used to determine the relationship between ASH test results using the two devices. The ICC = 0.77-0.99 result indicated good to excellent reliability for Active5™ usage. A high to a very high correlation between the two devices at 180° and 90° was noted (r = 0.75-0.95). This data supports the isometric-based strength training device Active5™ as a reliable and valid tool for ASH test performance.Entities:
Keywords: Active5; biomedical monitoring; force plates; injury prevention; physiotherapy; rehabilitation; sports medicine; upper limb
Year: 2022 PMID: 35453777 PMCID: PMC9024613 DOI: 10.3390/biology11040577
Source DB: PubMed Journal: Biology (Basel) ISSN: 2079-7737
Figure 1Technical characteristics of portable isometric-based strength training device Active5™: (a) Length and width; (b) width and height. The weight exceeds 131 g, including the battery.
Figure 2Technical characteristics of K-Force Plates separately for right and left limbs: (a) Length and width; (b) length, width, and height. The weight exceeds 1600 g.
Figure 3The preparation of the participant before the consecutive series of tests.
Illustration of consecutive tests being performed.
| Active5™ | Consecutive Tests | K-Force Plate | Consecutive Tests |
|---|---|---|---|
|
| 180°; I test |
| 180°; I test |
|
| 135°; Y test |
| 135°; Y test |
|
| 90°, T test |
| 90°; T test |
Intrarater reliability of the ASH test in the female group using both studied devices.
| Studied Limb | Shoulder Abduction | Active5™ | K-Force Plates |
|---|---|---|---|
| Dominant limb | 180° | 0.932 (0.883, 0.960) | 0.963 (0.933, 0.979) |
| 135° | 0.959 (0.929, 0.976) | 0.979 (0.963, 0.988) | |
| 90° | 0.941 (0.899, 0.966) | 0.979 (0.964, 0.988) | |
| Nondominant limb | 180° | 0.948 (0.908, 0.971) | 0.979 (0.963, 0.988) |
| 135° | 0.866 (0.776, 0.921) | 0.982 (0.969, 0.990) | |
| 90° | 0.942 (0.900, 0.966) | 0.959 (0.929, 0.976) |
Values expressed as intraclass correlation coefficient (ICC) and 95% confidence interval (lower bound, upper bound) of the ICC.
Inter-rater reliability of the ASH test in the female group using both studied devices.
| Studied Limb | Shoulder Abduction | Active5™ | K-Force Plates |
|---|---|---|---|
| Dominant limb | 180° | 0.969 (0.946, 0.982) | 0.976 (0.956, 0.986) |
| 135° | 0.967 (0.942, 0.981) | 0.983 (0.970, 0.990) | |
| 90° | 0.976 (0.955, 0.985) | 0.984 (0.972, 0.991) | |
| Nondominant limb | 180° | 0.947 (0.904, 0.970) | 0.968 (0.937, 0.983) |
| 135° | 0.955 (0.922, 0.974) | 0.983 (0.970, 0.990) | |
| 90° | 0.962 (0.935, 0.978) | 0.981 (0.966, 0.989) |
Values expressed as intraclass correlation coefficient (ICC) and 95% confidence interval (lower bound, upper bound) of the ICC.
The results of force measurements at 180° shoulder abduction using Active5™ and K-Force Plates.
| Force Measured at 180° Shoulder Abduction in Females (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 5.67 | 1.68 | 29.65 | 0.34 | 0.94 |
| K-Force Plates | 5.37 | 1.63 | 30.37 | 0.33 | 0.91 | ||
| Session 2 | Active5™ | 5.56 | 1.61 | 28.91 | 0.33 | 0.91 | |
| K-Force Plates | 5.23 | 1.65 | 31.53 | 0.34 | 0.94 | ||
| Session 3 | Active5™ | 5.37 | 1.51 | 28.16 | 0.31 | 0.85 | |
| K-Force Plates | 5.32 | 1.64 | 30.94 | 0.34 | 0.94 | ||
| Nondominant | Session 1 | Active5™ | 4.95 | 1.60 | 32.29 | 0.32 | 0.88 |
| K-Force Plates | 4.66 | 1.68 | 36.15 | 0.34 | 0.94 | ||
| Session 2 | Active5™ | 4.75 | 1.60 | 33.59 | 0.32 | 0.88 | |
| K-Force Plates | 4.46 | 1.54 | 34.58 | 0.32 | 0.88 | ||
| Session 3 | Active5™ | 4.86 | 1.64 | 33.89 | 0.34 | 0.94 | |
| K-Force Plates | 4.50 | 1.72 | 38.14 | 0.35 | 0.96 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).
Correlation between ASH test results using Active5™ and K-Force plates in the dominant upper limb in the female group.
| Correlation between ASH Test Results Using Active5™ and K-Force Plates in the Dominant Limb in Females | ||||
|---|---|---|---|---|
| Session 1 | 180° | <0.0001 | 0.875 | 0.766 |
| 135° | 0.625 | −0.104 | 0.010 | |
| 90° | <0.0001 | 0.931 | 0.868 | |
| Session 2 | 180° | <0.0001 | 0.879 | 0.773 |
| 135° | 0.591 | −0.115 | 0.013 | |
| 90° | <0.0001 | 0.945 | 0.893 | |
| Session 3 | 180° | <0.0001 | 0.850 | 0.722 |
| 135° | 0.710 | −0.079 | 0.001 | |
| 90° | <0.0001 | 0.919 | 0.845 | |
Values expressed as a p value of statistical significance; r value: Pearson’s correlation coefficient and r2 indicating the proportion of variance.
Correlation between ASH test results using Active5™ and K-Force plates in the nondominant upper limb in the female group.
| Correlation between ASH Test Results Using Active5™ and K-Force Plates in the Nondominant Limb in Females | ||||
|---|---|---|---|---|
| Session 1 | 180° | <0.0001 | 0.855 | 0.731 |
| 135° | 0.580 | −0.118 | 0.014 | |
| 90° | <0.0001 | 0.862 | 0.744 | |
| Session 2 | 180° | <0.0001 | 0.865 | 0.749 |
| 135° | 0.795 | −0.055 | 0.003 | |
| 90° | <0.0001 | 0.905 | 0.819 | |
| Session 3 | 180° | <0.0001 | 0.840 | 0.706 |
| 135° | 0.726 | −0.075 | 0.005 | |
| 90° | <0.0001 | 0.888 | 0.789 | |
Values expressed as a p value of statistical significance; r value: Pearson’s correlation coefficient and r2 indicating the proportion of variance.
The results of force measurements at 135° shoulder abduction using Active5™ and K-Force Plates in the female group.
| Force Measured at 135° Shoulder Abduction in Females (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 5.99 | 2.03 | 34.00 | 0.41 | 1.14 |
| K-Force Plates | 7.12 | 3.26 | 45.79 | 0.66 | 1.82 | ||
| Session 2 | Active5™ | 5.95 | 2.03 | 34.17 | 0.41 | 1.13 | |
| K-Force Plates | 7.15 | 3.16 | 44.29 | 0.64 | 1.77 | ||
| Session 3 | Active5™ | 6.11 | 2.26 | 37.11 | 0.46 | 1.27 | |
| K-Force Plates | 7.29 | 3.28 | 45.01 | 0.66 | 1.82 | ||
| Nondominant | Session 1 | Active5™ | 5.50 | 2.04 | 37.06 | 0.41 | 1.13 |
| K-Force Plates | 6.46 | 3.17 | 49.16 | 0.64 | 1.77 | ||
| Session 2 | Active5™ | 5.42 | 1.98 | 36.61 | 0.40 | 1.11 | |
| K-Force Plates | 6.41 | 3.10 | 48.34 | 0.63 | 1.74 | ||
| Session 3 | Active5™ | 5.48 | 2.14 | 39.11 | 0.43 | 1.19 | |
| K-Force Plates | 6.59 | 3.16 | 47.99 | 0.64 | 1.77 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).
The results of force measurements at 90° shoulder abduction using Active5™ and K-Force Plates in the female group.
| Force Measured at 90° of Shoulder Abduction in Females (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 6.88 | 2.99 | 43.50 | 0.61 | 1.69 |
| K-Force Plates | 6.59 | 3.17 | 48.20 | 0.64 | 1.77 | ||
| Session 2 | Active5™ | 6.69 | 3.00 | 44.80 | 0.61 | 1.68 | |
| K-Force Plates | 6.51 | 3.27 | 50.23 | 0.66 | 1.82 | ||
| Session 3 | Active5™ | 6.96 | 2.91 | 41.94 | 0.59 | 1.63 | |
| K-Force Plates | 6.51 | 3.29 | 50.56 | 0.67 | 1.85 | ||
| Nondominant | Session 1 | Active5™ | 6.09 | 2.39 | 39.25 | 0.48 | 1.32 |
| K-Force Plates | 5.62 | 2.36 | 41.97 | 0.48 | 1.32 | ||
| Session 2 | Active5™ | 5.95 | 2.40 | 40.39 | 0.49 | 1.35 | |
| K-Force Plates | 5.61 | 2.37 | 42.40 | 0.48 | 1.32 | ||
| Session 3 | Active5™ | 6.22 | 2.35 | 37.90 | 0.48 | 1.32 | |
| K-Force Plates | 5.55 | 2.42 | 43.71 | 0.49 | 1.35 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).
Intrarater reliability of the ASH test using both studied devices in the male group.
| Studied Limb | Shoulder Abduction | Active5™ | K-Force Plates |
|---|---|---|---|
| Dominant limb | 180° | 0.900 (0.794, 0.953) | 0.936 (0.855, 0.972) |
| 135° | 0.961 (0.917, 0.982) | 0.968 (0.932, 0.985) | |
| 90° | 0.925 (0.842, 0.965) | 0.982 (0.961, 0.992) | |
| Nondominant limb | 180° | 0.912 (0.806, 0.960) | 0.968 (0.931, 0.985) |
| 135° | 0.891 (0.762, 0.950) | 0.980 (0.958, 0.991) | |
| 90° | 0.945 (0.884, 0.974) | 0.978 (0.952, 0.990) |
Values expressed as intraclass correlation coefficient (ICC) and 95% confidence interval (lower bound, upper bound) of the ICC.
Inter-rater reliability of the ASH test using both studied devices in the male group.
| Studied Limb | Shoulder Abduction | Active5™ | K-Force Plates |
|---|---|---|---|
| Dominant limb | 180° | 0.949 (0.892, 0.976) | 0.959 (0.912, 0.981) |
| 135° | 0.972 (0.940, 0.987) | 0.964 (0.924, 0.984) | |
| 90° | 0.970 (0.935, 0.986) | 0.990 (0.979, 0.989) | |
| Nondominant limb | 180° | 0.903 (0.796, 0.955) | 0.968 (0.931, 0.985) |
| 135° | 0.954 (0.902, 0.979) | 0.954 (0.902, 0.979) | |
| 90° | 0.965 (0.925, 0.984) | 0.989 (0.975, 0.995) |
Values expressed as intraclass correlation coefficient (ICC) and 95% confidence interval (lower bound, upper bound) of the ICC.
The results of force measurements at 180° shoulder abduction using Active5™ and K-Force Plates.
| Force Measured at 180° of Shoulder Abduction in Males (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 9.48 | 3.00 | 31.65 | 0.57 | 1.57 |
| K-Force Plates | 9.18 | 3.39 | 36.95 | 0.65 | 1.80 | ||
| Session 2 | Active5™ | 9.33 | 2.95 | 31.65 | 0.56 | 1.55 | |
| K-Force Plates | 8.90 | 2.99 | 33.65 | 0.57 | 1.57 | ||
| Session 3 | Active5™ | 9.28 | 3.43 | 36.98 | 0.66 | 1.82 | |
| K-Force Plates | 8.74 | 2.90 | 33.25 | 0.55 | 1.52 | ||
| Nondominant | Session 1 | Active5™ | 8.49 | 2.75 | 32.44 | 0.53 | 1.46 |
| K-Force Plates | 7.72 | 2.90 | 37.57 | 0.55 | 1.52 | ||
| Session 2 | Active5™ | 8.12 | 2.50 | 30.82 | 0.48 | 1.32 | |
| K-Force Plates | 7.40 | 2.61 | 35.25 | 0.50 | 1.38 | ||
| Session 3 | Active5™ | 8.05 | 2.65 | 32.92 | 0.51 | 1.41 | |
| K-Force Plates | 7.62 | 2.70 | 35.53 | 0.52 | 1.44 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).
Correlation between ASH test results using Active5™ and K-Force plates in the dominant upper limb in the male group.
| Correlation between ASH Test Results Using Active5™ and K-Force Plates in the Dominant Limb in Males | ||||
|---|---|---|---|---|
| Session 1 | 180° | <0.0001 | 0.743 | 0.552 |
| 135° | 0.170 | 0.272 | 0.074 | |
| 90° | <0.0001 | 0.682 | 0.465 | |
| Session 2 | 180° | <0.0001 | 0.855 | 0.731 |
| 135° | 0.164 | 0.275 | 0.076 | |
| 90° | <0.0001 | 0.666 | 0.443 | |
| Session 3 | 180° | <0.0001 | 0.846 | 0.716 |
| 135° | 0.064 | 0.362 | 0.131 | |
| 90° | <0.0001 | 0.698 | 0.487 | |
Values expressed as a p value of statistical significance; r value: Pearson’s correlation coefficient and r2 indicating the proportion of variance.
Correlation between ASH test results using Active5™ and K-Force plates in the nondominant upper limb in the male group.
| Correlation between ASH Test Results Using Active5™ and K-Force Plates in the Nondominant Limb in Males | ||||
|---|---|---|---|---|
| Session 1 | 180° | <0.0001 | 0.791 | 0.626 |
| 135° | 0.152 | 0.283 | 0.080 | |
| 90° | <0.0001 | 0.642 | 0.412 | |
| Session 2 | 180° | <0.0001 | 0.788 | 0.621 |
| 135° | 0.097 | 0.326 | 0.106 | |
| 90° | 0.001 | 0.618 | 0.382 | |
| Session 3 | 180° | <0.0001 | 0.865 | 0.748 |
| 135° | 0.147 | 0.286 | 0.082 | |
| 90° | <0.0001 | 0.707 | 0.499 | |
Values expressed as a p value of statistical significance; r value: Pearson’s correlation coefficient and r2 indicating the proportion of variance.
The results of force measurements at 135° of shoulder abduction using Active5™ and K-Force Plates.
| Force Measured at 135° of Shoulder Abduction in Males (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 7.65 | 2.95 | 38.62 | 0.56 | 1.55 |
| K-Force Plates | 5.76 | 1.84 | 31.96 | 0.35 | 0.96 | ||
| Session 2 | Active5™ | 7.54 | 3.05 | 40.48 | 0.58 | 1.60 | |
| K-Force Plates | 5.83 | 1.84 | 31.67 | 0.35 | 0.96 | ||
| Session 3 | Active5™ | 7.61 | 3.26 | 42.79 | 0.62 | 1.71 | |
| K-Force Plates | 5.87 | 1.91 | 32.51 | 0.36 | 0.99 | ||
| Nondominant | Session 1 | Active5™ | 6.75 | 2.34 | 34.66 | 0.45 | 1.24 |
| K-Force Plates | 5.12 | 1.75 | 34.25 | 0.33 | 0.91 | ||
| Session 2 | Active5™ | 6.87 | 2.75 | 40.10 | 0.53 | 1.46 | |
| K-Force Plates | 5.11 | 1.65 | 32.35 | 0.31 | 0.85 | ||
| Session 3 | Active5™ | 6.55 | 2.86 | 43.63 | 0.55 | 1.52 | |
| K-Force Plates | 5.19 | 1.74 | 33.49 | 0.33 | 0.91 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).
The results of force measurements at 135° of shoulder abduction using Active5™ and K-Force Plates.
| Force Measured at 90° of Shoulder Abduction in Males (kg) | |||||||
|---|---|---|---|---|---|---|---|
| Studied Limb | Session | Device | x | SD | CV % | SEM | MDC |
| Dominant | Session 1 | Active5™ | 5.95 | 2.04 | 34.39 | 0.39 | 1.08 |
| K-Force Plates | 5.37 | 1.71 | 31.87 | 0.32 | 0.88 | ||
| Session 2 | Active5™ | 5.95 | 2.11 | 35.60 | 0.40 | 1.11 | |
| K-Force Plates | 5.33 | 1.75 | 32.88 | 0.33 | 0.91 | ||
| Session 3 | Active5™ | 6.05 | 2.31 | 38.18 | 0.44 | 1.21 | |
| K-Force Plates | 5.41 | 1.71 | 31.60 | 0.32 | 0.88 | ||
| Nondominant | Session 1 | Active5™ | 5.36 | 1.65 | 30.87 | 0.31 | 0.85 |
| K-Force Plates | 4.76 | 1.34 | 28.30 | 0.25 | 0.69 | ||
| Session 2 | Active5™ | 5.41 | 1.88 | 34.80 | 0.36 | 0.99 | |
| K-Force Plates | 4.75 | 1.41 | 29.67 | 0.27 | 0.74 | ||
| Session 3 | Active5™ | 5.46 | 1.89 | 34.75 | 0.36 | 0.99 | |
| K-Force Plates | 4.77 | 1.34 | 28.11 | 0.25 | 0.69 | ||
Values are expressed as arithmetic mean (x), standard deviation (SD), coefficient of variation (CV), and standard error of measurement (SEM); minimal detectable change (MDC).