| Literature DB >> 35441716 |
Marco Cavallone1,2, Patrik Gonçalves Jorge3, Raphaël Moeckli3, Claude Bailat3, Alessandro Flacco2, Yolanda Prezado4,5, Rachel Delorme6,7.
Abstract
BACKGROUND: Ultra-high dose-rate (UHDR) irradiations (>40 Gy/s) have recently garnered interest in radiotherapy (RT) as they can trigger the so-called "FLASH" effect, namely a higher tolerance of normal tissues in comparison with conventional dose rates when a sufficiently high dose is delivered to the tissue. To transfer this to clinical RT treatments, adapted methods and practical tools for online dosimetry need to be developed. Ionization chambers remain the gold standards in RT but the charge recombination effects may be very significant at such high dose rates, limiting the use of some of these dosimeters. The reduction of the sensitive volume size can be an interesting characteristic to reduce such effects.Entities:
Keywords: FLASH radiotherapy; dosimetry; ion recombination; ionization chamber; ultra-high dose-rate
Mesh:
Year: 2022 PMID: 35441716 PMCID: PMC9539950 DOI: 10.1002/mp.15675
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.506
FIGURE 1Experimental irradiation setup. Right: picture of the eRT6 Oriatron. Left: picture of the RW3 water equivalent phantom in which the Razor Nano Chamber was inserted. The picture also shows the EBT3 film placed at the phantom surface for simultaneous reference dose measurement. Series of measurements were performed with SSDs of 2, 0.5, 0.3, and 0.2 m
Uncertainties on various parameters finally affecting the ICE uncertainty
| Measurements affected by uncertainties | Contributions | Description | Uncertainty | Total uncertainty (method) |
|---|---|---|---|---|
| EBT3 dose |
| SD between two films during calibration | 0.5% | 0–15 Gy: 3.4% ; 15–30 Gy: 6% (Quadratic sum) |
|
| Uncertainty on IC dose during calibration | 1.5% | ||
| σFilm, SD | Inhomogeneity of films during UHDR irradiations (SD in 2 mm diameter) | 3% | ||
|
| Deviation between channels above 15 Gy | 5% | ||
| Dose at RNC position |
| 3.4% (0–15 Gy) ; 6% (15–30 Gy) | 0–15 Gy: 4.2% ; 15–30 Gy: 6.5% (Quadratic sum) | |
| σkf | Depth dose corrections—SD between two measurements | 2.5% | ||
|
| σQ | Repeatability of RNC response | 0.4% | 2.5% (Monte Carlo approach) |
|
| 4.2% (0–15 Gy) ; 6.5% (15–30 Gy) | |||
| ICE | σQ | Repeatability of RNC response | 0.4% | 5.8% ( |
| σQsat | 2.5% | |||
| ICE fitting curves | σICE | 5.8% | 1–8% (Monte Carlo approach) | |
|
| 4.2% (0–15 Gy) ; 6.5% (15–30 Gy) |
FIGURE 2Linearity measurement to calibrate the RNC response under reference electron beam of 8 MeV delivered at conventional dose rate (Elekta Synergy, CHUV). The uncertainty on the dose was of 1.5% (horizontal error bars), and that on the RNC charge response was of 0.6% (vertical error bars)
FIGURE 3Linearity: Charge RNC response as a function of the total dose delivered at high dose rate with the eRT6 accelerator. The beam parameter used were a frequency of 10 Hz, a Grid Tension of 200 V and an SSD of 0.5 m. Two series were performed at a pulse duration of 1 and 3 μs, providing DPP of 1.9 and 5.5 Gy, respectively. The uncertainty on the RNC charge response was of 0.43% (vertical error bars)
FIGURE 4RNC charge response as a function of the dose‐per‐pulse (DPP). Two series were performed at pulse lengths of 1 μs (red dots) and 3 μs (blue dots), with DPP ranging from 0.01 to 30 Gy. The black linear dashed curve represents the saturation charge , that is, the charge that would be collected to the RNC electrodes if no recombination occurred. The uncertainty on DPP was of 5% (horizontal error bars) and that on the RNC charge response was of 0.43% (vertical error bars)
Parameters of the fit obtained for 1 and 3 μs with the logistic recombination model proposed by Petersson et al. and described in Equation 5
| Parameters |
|
|
|
|
|---|---|---|---|---|
| Fit 1 μs | 0.76 | 0.61 | 5.12 | 0.97 |
| Fit 3 μs | 0.97 | 0.35 | 3.28 | 0.94 |
FIGURE 5Ion collection efficiency of the RNC as a function of the dose per pulse, for pulse lengths of 1 μs (red curve) and 3 μs (blue curve), fitted with the logistic recombination model. The uncertainties on ICE (see section 2.5) are represented by the shaded region around the main curves. The RNC ICE results are compared with those of the Markus Advanced ion chamber of PTW for pulse lengths of 0.5–1.8 μs, as published by Petersson et al.