Dosimetric accuracy of three dose calculation algorithms for radiation therapy of in situ non-small cell lung carcinoma.

Manda Švabić Kolacio1, David Rajlić1, Milan Radojčić2, Đeni Smilović Radojčić1,3, Nevena Obajdin1, Dea Dundara Debeljuh1,3,4, Slaven Jurković1,3.   

Abstract

Background: Study determines differences in calculated dose distributions for non-small cell lung carcinoma (NSC LC) patients. NSC LC cases were investigated, being the most common lung cancer treated by radiotherapy in our clinical practice. Materials and methods: A retrospective study of 15 NSCLC patient dose distributions originally calculated using standard superposition (SS) and recalculated using collapsed cone (CC ) and Monte Carlo (MC) based algorithm expressed as dose to medium in medium (MCDm) and dose to water in medium (MCDw,) was performed so that prescribed dose covers at least 99% of the gross target volume (GTV). Statistical analysis was performed for differences of conformity index (CI), heterogeneity index (HI), gradient index (GI), dose delivered to 2% of the volume (D2%), mean dose (Dmean) and percentage of volumes covered by prescribed dose (V70Gy). For organs at risk (OARs), Dmean and percentage of volume receiving 20 Gy and 5Gy (V20Gy, V5Gy) were analysed.
Results: Statistically significant difference for GTVs was observed between MCDw and SS algorithm in mean dose only. For planning target volumes (PTVs), statistically significant differences were observed in prescribed dose coverage for CC, MCDm and MCDw. The differences in mean CI value for the CC algorithm and mean HI value for MCDm and MCDw were statistically significant. There is a statistically significant difference in the number of MUs for MCDm and MCDw compared to SS.
Conclusion: All investigated algorithms succeed in managing the restrictive conditions of the clinical goals. This study shows the drawbacks of the CC algorithm compared to other algorithms used.
© 2022 Greater Poland Cancer Centre.

Entities:  

Keywords:  Monaco; TPS; XiO standard superposition; non-small cell lung carcinoma; radiation therapy

Year:  2022        PMID: 35402037      PMCID: PMC8989458          DOI: 10.5603/RPOR.a2022.0013

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


Introduction

Accuracy of dose calculation algorithms is one of the prerequisites for successful radiation therapy outcome. Thorough investigation of possible limitations of calculation algorithms is important because the total error in delivered absorbed dose of 5% can lead up to at least 20% of change in Tumour Control Probability (TCP) and Normal Tissue Complication Probability (NTCP) [1, 2]. Therefore, the quantification and characterization of possible sources of errors in absorbed dose calculation is required. The introduction of a more advanced Elekta Monaco treatment planning system (TPS) at the Radiation Oncology Department of the University Hospital Rijeka opened the possibility of optimisation and calculation of absorbed dose distributions for the planning of certain tumour sites using different dose calculation algorithms. Until recently, the radiation therapy planning technique of choice for Non-Small Cell Lung Carcinoma (NSCLC) was forward planned intensity modulated radiation therapy (IMRT) using the standard superposition (SS) analytical dose calculation algorithm built in the Elekta XiO TPS. After commissioning and clinical implementation of the Elekta Monaco TPS, the intention was to transfer the workflow for NSCLC cases to the advanced TPS. Elekta Monaco TPS uses the collapsed cone (CC) analytical dose calculation algorithm for forward planning IMRT and a Monte Carlo (MC) based algorithm for inverse planning IMRT. The clinical impact of using different dose calculation algorithms, especially comparisons taking into account MC based algorithms, have been in focus of research since their implementation in radiation oncology and are still an interesting research topic [3-9]. Considering the known differences in dose calculation engines of the abovementioned algorithms, especially in low-density media, it was decided to investigate lung cancer patients, specifically NSCLC cases, because they are the most common in our clinical practice. Prior to implementation of SS algorithm into clinical practice, a thorough validation of absorbed dose calculation accuracy of the algorithm compared to measurements in the CIRS Thorax semi-anthropomorphic phantom was performed [10]. In order to investigate the dosimetric accuracy of absorbed dose calculation of the algorithms built in the Elekta Monaco TPS, the same set of calculations and measurements was repeated [11, 12]. Discrepancies between measured and calculated dose in lung density equivalent (LDE) and bone density equivalent (BDE) regions of the phantom were observed for the algorithms built in the Elekta Monaco TPS. For the MC based Monaco algorithm, similar behaviour in BDE was reported in earlier studies [13, 14]. Magnitude of reported discrepancies could be related to the dose calculation algorithm used. Considering that, observed discrepancies in the LDE regions raise the question of possible clinical implications for lung cancer patients. Thus, the dosimetric evaluation results in semi-anthropomorphic phantom incited the main focus of this paper to be an investigation of possible clinical impact. Consequently, the evaluation of performance and possible limitations of algorithms built in Monaco TPSs for patients with in-situ NSCLC, considering the large regions of low-density tissue encompassed in irradiated volumes, was examined.

Materials and methods

Treatment planning

This work was performed using devices which are in clinical use at the Radiation Oncology Department of University Hospital Rijeka: linear accelerator Siemens Oncor Expression, Siemens Somatom Open CT simulator (Siemens Healthineers, Erlangen, Germany) and TPSs XiO v.5.10.02 and Monaco v.5.11.02 (Elekta, Stockholm, Sweden). The intention of research was not a thorough examination of algorithm performances but the verification of acceptability of calculated absorbed dose for NSCLC patients. Nevertheless, it is necessary to emphasize the basic differences within dose calculation engines built in algorithms used in this study. Namely, the Elekta XiO SS analytical algorithm utilizes the patient CT data as water with variable electron densities for absorbed dose calculation and, consequently, expresses dose as dose to water in water [15]. In the Elekta Monaco CC analytical algorithm the dose concept is to calculate the dose to actual medium and, therefore, report dose to medium [16, 17]. The absorbed dose calculation algorithm built in the Elekta Monaco TPS used for inverse IMRT planning is based on MC simulation (X-ray Voxel Monte Carlo, XVMC) in a patient model [18]. It provides two dose reporting modes: dose to medium in medium (MCDm) and dose to water in medium (MCDw). For the MC based algorithm, there is neither general agreement [13, 19–21] nor recommendation which reporting mode should be used for expressing dose in clinical cases. Thus, absorbed dose distributions created for this retrospective study were calculated and optimized using both approaches. Statistical uncertainty (SU) used for both dose calculations was 0.5% per control point. It is worth noting that the results of the investigation of our group using ab initio MC simulation for the absorbed dose calculation in low-density media and simple geometries show that the MCDm and MCDw reporting modes can be considered equivalent [22]. Further work will be dedicated to verify the aforementioned results in clinical cases.

Phantom measurements

The dosimetric accuracy of investigated algorithms for 6MV photon beam was initially verified through absorbed dose measurements in a water phantom. TPS dose point values were used for the analytical algorithms used in this study. For both calculation options of the MC algorithm, the dose calculation was averaged over a 0.081 cm3 spherical volume with 81 dose evaluation points. Discrepancies for examined algorithms between measured and calculated data in water were determined according to international recommendations [11, 12, 23]. Evaluation of the results was performed using the expression: where, D is calculated dose at a particular point in the phantom, D is the measured dose at the point of calculation in the phantom, D is the measured dose at the reference point in the phantom. The algorithms were further validated using the semi-anthropomorphic CIRS Thorax phantom, made from three different density regions (water density equivalent — WDE, LDE and BDE) (Fig. 1). The phantom was scanned using the standard CT protocol for lung cancer patients (120 kVp tube voltage, 95 mAs, 3 mm slice thickness, extended field of view), contoured and used for dose calculations. Measurements were performed at 10 points in the phantom using the PTW30013 ionization chamber.
Figure 1

Cross section of CIRS Thorax phantom with three regions of different densities: water density equivalent (WDE) region (plastic water, ρ = 1.003 g/cm3) with measuring points 1–5, lung density equivalent (LDE) region (ρ = 0.207 g/cm3) with measuring points 6–9 and bone density equivalent (BDE) region (ρ = 1.506 g/cm3) with measuring point 10

Absorbed dose calculations provided by above mentioned algorithms were verified in a semi-anthropomorphic phantom for different beam arrangements following the recommendations [11, 12]. Discrepancies were calculated using Eq. 1. Tolerances were determined considering the complexity of the measurement geometry [24]. Furthermore, the results with large measurement uncertainty (areas of large dose gradient and radiation fields smaller than 4 × 4 cm2) were excluded from the analyses. For points located in the WDE region of the phantom, the tolerances were up to ± 3%, while in the non-water equivalent regions, the tolerance limits were up to ± 4%. One has to bear in mind that dosimetric validation has its own intrinsic limitations depending on the uncertainties of the dosimetrical system used, especially for measurements in non-water density equivalent media [25].

Clinical study design

To investigate possible implications of used calculation algorithms on radiation therapy treatment, a retrospective study of 15 NSCLC patients was performed. Dose distributions for these patients were originally calculated using the SS algorithm. For the purpose of this study, the absorbed dose distributions were created again using the Elekta Monaco inverse and forward planning IMRT modes by applying the MC based and CC algorithms, respectively. Beam angles and target margins were maintained in order to avoid variability in the results due to different beam arrangements. The patients underwent CT simulation in a supine position with free breathing conditions using standard lung cancer patient CT acquisition protocol. Conventional fractionation of 2 Gy per fraction was prescribed while the total dose was patient dependent. It was determined by the radiation oncologist considering patient age, general condition, comorbidities and lungs functional status (FEV1-forced expiratory volume in the first second and DLCO-diffusing capacity of the lungs for carbon monoxide). GTV was defined as a visible tumour along with enlarged lymph nodes with short axis diameter larger than 10 mm. The PTV margin for non-central lesions was 8–10 mm, except in the craniocaudal direction for which it was 10–15 mm depending on the tumour location [26-28]. GTV volumes ranged from 28.37 to 130.05 cm3 and PTV volumes varied between 95.3 and 366.3 cm3. All patients had solid tissue equivalent density GTVs. These patients were selected for their relatively small GTV volumes compared to their corresponding PTV volumes and large VPTV/VGTV ratios with PTVs including a high amount of lung tissue. In Figure 2 typical transversal CT slice of a patient included in the study is shown. Large difference in PTV and GTV volumes is noticeable.
Figure 2

Example of typical transversal CT slice of a patient included in the study showing large difference in the planning target volume (PTV) (grey line) and the gross target volume (GTV) (white line) volume as well as large lung volume included in the PTV

Planning objectives regarding the target volumes (TVs) and relevant organs at risk (OARs) were achieved in accordance with international guidelines [27, 29]. All dose distributions were calculated and optimized in a way that at least 99% of the GTV should be covered by the prescribed dose. The dose calculation voxel size was 3 × 3 × 3 mm3 for all calculations. For inverse IMRT the minimum open field size was 4 cm2, the statistical uncertainty was kept at 0.5% per control point and the minimum number of monitor units (MUs) per segment was 2. The IMRT delivery mode was step and shoot. To facilitate intercomparation regardless of the original patient dose prescription, all absorbed dose distributions were set to 70 Gy (2 Gy per fraction, 35 fractions).

Statistical analysis

Calculated dose volume histograms (DVH) were analysed to quantify the dose to TVs and relevant OARs. Maximum dose defined as dose delivered to 2% of the volume (D2%), mean dose (Dmean) and percentage of volumes covered by the prescribed dose were also used to validate dose distribution differences for planning target volumes (PTVs) and gross tumour volumes (GTVs). For the OARs (ipsilateral, contralateral and whole lungs), Dmean and percentage of volume receiving 20 Gy and 5 Gy (V20Gy, V5Gy), were validated [30-32]. Additional analysed parameters were: heterogeneity index (HI) (Eq. 2), conformity index (CI) (Eq. 3) and gradient index (GI) (Eq. 4) defined as: where, D is dose delivered to 2% of the volume, D is dose delivered to 98% of the target volume, TV1 is target volume covered by the prescribed dose, TV is target volume and VR is total volume of prescribed dose. V% represents the volume irradiated by 50% of prescribed dose. The statistical analysis method used was two-tailed Student’s t-test, calculated using TIBCO Statistica version 13.5.0.17 (TIBCO Software Inc). A value of p < 0.05 was considered statistically significant.

Results

Phantom measurements

Calculated discrepancies for examined algorithms in water (Eq. 1) were found to be within tolerances proposed for different beam arrangements and levels of complexity [24]. Comparison of calculated and measured absorbed dose at respective points in the semi-anthropomorphic CIRS Thorax phantom for evaluated algorithms is shown in Figures 3 and 4. In Figure 3 differences between measured and calculated absorbed dose for different beam arrangements in the WDE region of the phantom are shown for five measuring points (points 1–5, Fig. 1) and dose calculation algorithms used with respective tolerances [24].
Figure 3

Box and Whisker plot for differences between measured and calculated absorbed dose in WDE region of the semi-anthropomorphic phantom at five measuring points (points 1–5, Fig. 1) for all evaluated algorithms. SS — standard superposition; CC — collapsed cone; MCDm — Monte Carlo dose to medium; MCDw — Monte Carlo recalculation to dose to water from dose to medium

Figure 4

Box and Whisker plot for differences between measured and calculated absorbed dose in lung density equivalent (LDE) regions of the semi-anthropomorphic phantom at four measuring points (6–9, Fig. 1) for all evaluated algorithms. SS — standard superposition; CC — collapsed cone; MCDm — Monte Carlo dose to medium; MCDw — Monte Carlo recalculation to dose to water from dose to medium

In Figure 4 differences between measured and calculated absorbed dose in the LDE regions of the phantom are shown for four measuring points (points 6–9, Fig. 1) and dose calculation algorithms used with respective tolerances [24]. In the WDE region of the semi-anthropomorphic phantom (Fig. 3), the mean value of differences for the SS algorithm was 0.05% (from −3.2% to +3.2% with standard deviation SD = 1.23%) and in the LDE region (Fig. 4), it was 0.42% (−1.9% to +3.8%, SD = 1.27%). Considering the good agreement of the measured and calculated data, SS was taken as the algorithm of choice for absorbed dose distribution calculations for radiation therapy of in-situ NSCLC at that time. Also, absorbed dose calculated using algorithms built in Monaco TPS was compared to measured absorbed dose values. The results in the WDE region (Fig. 3.) are in accordance with the proposed tolerances [24]. For MCDm calculation, mean difference between the measured and calculated absorbed dose, was −0.59% (from −3.21% up to 2.64%, SD = 1.24%) and for MCDw mean difference was −0.58% (from −3.24% to 2.58%, SD = 1.25%). The results obtained for the CC algorithm show slightly larger deviations compared to the results of remaining algorithms. The mean difference for CC is −0.62% (from −3.19% to 4.33%, SD = 1.49%) in the WDE region. In the LDE region (Fig. 4.) of the phantom, the mean differences between the measured and calculated absorbed dose values for all measuring points were: −1.76% (−6.1% to +3.9%, SD = 2.29%) for CC, and 0.12% (−1.5% to +3.6%, SD = 1.21%) for MCDm, and 0.01% (−2.0% to +3.5%, SD = 1.24%) for MCDw. These results indicate that for SS, MCDm and MCDw differences between calculated values and measurements in the LDE region were in tolerance. Additionally, it should be emphasized that the differences between MCDm and MCDw compared with measured absorbed doses for all investigated points in the LDE region are almost the same. On the other hand, more than 20% of results for the CC algorithm were out of tolerance (Fig. 4).

Clinical study

A comparison of absorbed dose distributions calculated using CC, MCDm and MCDw versus SS has been carried out for the 15 NSCLC patient data sets. Parameters based on dose calculated using aforementioned algorithms in comparison with SS and corresponding statistical analysis results for NSCLC patients in this study are shown in Tables 1, 2.
Table 1

Colapsed cone (CC ), Monte Carlo (MC) based algorithm expressed as dose to medium in medium (MCDm) and dose to water in medium (MCDw) absorbed dose calculation comparison with standard superposition (SS) for the gross target volume (GTVs) and the planning target volume (PTVs): mean values of relevant parameters with related standard deviations and p-values

StructureParameterSSCCCC vs. SSMCDmMCDm vs. SSMCDwMCDw vs. SS
Mean (SD)Mean (SD)p-ValueMean (SD)p-valueMean (SD)p-value
GTVV70Gy (%)99.84 (0.16)99.67 (0.16)0.05721799.86 (0.18)0.79492099.72 (0.33)0.377364
Dmean (Gy)72.39 (0.35)72.21 (0.70)0.53588372.15 (0.13)0.08880472.01 (0.23)0.020618
D2% (Gy)73.41 (0.13)73.36 (0.23)0.62127773.26 (0.26)0.16813073.42 (0.21)0.891330
CI0.27 (0.05)0.28 (0.06)0,7779530.25 (0.07)0,5195270.25 (0.07)0,470062
HI1.03 (0.01)1.04 (0.01)0.0572761.03 (0.00)0.4354281.03 (0.01)0.17643
PTVV70Gy (%)92.52 (2.17)86.16 (3.07)0.00020396.96(1.55)0.00016196.91 (1.78)0.00425
Dmean (Gy)71.85 (0.26)71.51 (0.65)0.19172971.75 (0.21)0.43135371.68 (0.27)0.212450
D2% [Gy]73.37 (0.1)73.38 (0.20)0.97771973.34 (0.34)0.81252773.39 (0.17)0.810185
CI0.76 (0.05)0.7 (0.04)0.0245550.77 (0.07)0.8559470.75 (0.07)0.751894
GI6.19 (1.03)6.32 (1.09)0.8094515.32 (0.67)0.0640045.38 (0.61)0.076091
HI1.05 (0.01)1.07 (0.01)0.0012491.04 (0.01)0.0010321.04 (0.01)0.005861

CI — conformity index; GI — gradient index; HI — homogeneity index; p-values in bold are statistically significant

Table 2

Mean values of relevant parameters with related standard deviations for ipsilateral, contralateral and whole lungs for standard superposition (SS), Colapsed cone (CC ), Monte Carlo (MC) based algorithm expressed as dose to medium in medium (MCDm) and dose to water in medium (MCDw), with p-values determined by comparison of CC, MCDm, MCDw with SS are shown

StructureParameterCC vs.SSMCDm vs. SSMCDw vs. SS
Mean diff. (SD)p-valueMean diff. (SD)p-valueMean diff. (SD)p-value
Ispilateral lungV5Gy (%)0.57 (2.31)0.94−0.12 (3.53)0.99−0.17 (3.57)0.98

V20Gy (%)−1.39 (1.32)0.83−1.99 (2.13)0.76−1.68 (2.26)0.79

Dmean [Gy]−0.41 (0.64)0.91−0.16 (1.38)0.960.01 (1.44)1.00

Contralateral lungV5Gy (%)1.53 (3.03)0.802.41 (4.68)0.681.82 (5.01)0.76

V20Gy (%)−0.16 80.42)0.85−0.22 (0.64)0.78−0.18 (0.67)0.83

Dmean [Gy]0.25 (0.34)0.740.41 (0.63)0.590.4 (0.71)0.60

Whole lungsV5Gy (%)0.53 (2.65)0.930.89 (3.79)0.880.8 (3.72)0.89

V20Gy (%)−0.71 (0.48)0.83−0.87 (1.19)0.80−0.86 (1.14)0.80

Dmean [Gy]−0.04 (0.29)0.980.22 (0.82)0.910.22 (0.8)0.91

SD — standard deviation

TVs and additional structure

Results in Table 1 represent the comparison of CC, MCDm and MCDw mean values of examined relevant TV parameters with standard deviations (SD) in relation to SS algorithm values taken as reference. Statistical analysis was performed for GTVs as well as PTVs. For the GTVs, there is no statistically significant difference neither in prescribed dose coverage (V70Gy) nor in CI, HI and D2% when CC and MC data are compared with SS. On the other hand, there was a statistically significant difference in Dmean between MCDw and SS (p = 0.020618). Results of the PTV analysis show different trends. Namely, for the PTVs, there is a statistically significant difference for CC (V70Gy) with the mean value of 86.16% and SD = 3.07% (p = 0.000203), and a statistically significant difference for MCDm (V70Gy) with the mean value of 96.96% and SD = 1.55% (p = 0.000161), and a statistically significant difference for MCDw (V70Gy) with the mean value of 96.91% and SD = 1.78% (p = 0.004251) in comparison with the SS mean value of 92.52% and SD = 2.17%. Additionally, for the PTVs, statistically significant differences in CI and HI were observed. The mean CI has a lower value for the CC algorithm of 0.70, SD = 0.04 compared to the SS algorithm mean value of 0.76 and SD = 0.05 (p = 0.024555). The mean HI has a higher value for CC of 1.07, SD=0.10 compared to the SS mean value of 1.05 and SD = 0.01 (p = 0.001249). The mean HI has a lower value for MCDm (1.04, SD = 0.01, p = 0.001032), and MCDw (1.04, SD = 0.01, p = 0.005861) compared to the SS mean value. To further investigate the location of dose differences, an additional structure PTV minus GTV (PTV-GTV) was created and V70Gy was analysed for CC, MCDm and MCDw compared to SS calculated absorbed dose distributions. Analysis of the additional structure (PTV-GTV) revealed that the statistically significant difference was found for average V70Gy coverage for CC (79.34%, SD = 4.42%, p = 0.000363) and for MCDm and MCDw (95.42%, SD = 2.24%, p = 0.000075 and 95.06%, SD = 2.98%, p = 0.000372, respectively) compared to SS (88.34%, SD = 3.24%).

OARs

A comparison of absorbed dose distributions was also performed for OARs (ipsilateral, contralateral and whole lungs) where the differences could be expected. The comparison of results for respective OAR is shown in Table 2. For the ipsilateral, contralateral and total lungs no statistically significant differences in Dmean, V20Gy and V5Gy were found.

Number of MUs

In addition, the average number of MUs required for the optimization of absorbed dose distributions calculated by different algorithms used in this study has also been analysed. The results show that there is a statistically significant difference in the number of MUs for MCDm (average MU = 403, SD = 49 and p = 0.000037) and MCDw (average MU = 400, SD = 51 and p = 0.000088) compared to SS (average MU = 285, SD = 19). On the other hand, there was no statistically significant difference in the number of MUs for CC (average MU = 298, SD = 10) compared to SS.

Discussion

Comparison of calculated and measured absorbed dose in the LDE region of semi-anthropomorphic phantom showed that discrepancies for the CC algorithm were the most pronounced with about 20% out of tolerance values (Fig. 2). On the other hand, for other algorithms all discrepancies were in tolerance. Findings in this study were in accordance with published data [33, 34] which show discrepancies in the LDE region between absorbed dose values measured and calculated using CC algorithm. Additionally, deviations between measured and MCDm/MCDw calculated absorbed doses in the LDE region are in agreement with the results from a previously published study where the deviations between MCDm/MCDw calculated absorbed doses were compared to ab initio MC simulation [22]. Further analysis of clinical data related to GTVs has shown statistically significant difference only in Dmean when calculated data using MCDw and SS algorithms are compared (p = 0.020618). Despite the statistical significance, Dmean averages are above the prescribed dose and the D2% criteria are fulfilled for both algorithms. This difference, while being statistically significant, is in practice clinically irrelevant because all GTV objectives are fulfilled in both cases. From this perspective, all calculated absorbed dose distributions in this study should be appropriate for clinical use. There is quite a different situation when PTVs were analysed. While mean values of the GI, Dmean and D2% show no statistically significant difference, for V70Gy, CI and HI values statistically significant differences were observed. For mean CI, statistically significant difference for CC compared to the SS algorithm (p = 0.024555) implies inferior conformity of absorbed dose distributions calculated using CC compared to SS. The mean HI shows higher value for CC compared to SS (p = 0.001249) which implies inferior homogeneity of absorbed dose distributions calculated using CC compared to SS. Additionally, the mean HI has a lower value for MCDm (p = 0.001032) and MCDw (p = 0.005861) compared to SS which implies superior homogeneity of absorbed dose distributions calculated using MCDm and MCDw compared to SS. Observed lower PTV coverage in lung tissue for analytical algorithms compared to the MC based algorithm is in accordance with previously reported findings [35]. An example of difference in V70Gy in DVH comparison of PTVs and GTVs for one patient from the study is shown in Figure 5. Considering the V70Gy for the PTV, lower coverage for CC and higher coverage for the MC algorithm when compared to SS is apparent. For GTV the differences in V70Gy are negligible.
Figure 5

Example of dose volume histograms (DVH) comparison of standard superposition (SS) with collapsed cone (CC ) (A), Monte Carlo dose to medium (MCDm) (B) and Monte Carlo recalculation to dose to water from dose to medium (MCDw) (C) for planning target volume (PTV) (black lines) and gross target volume (GTV) (grey lines) coverage for a randomly chosen patient from this study

The analysis of additional structure (PTV-GTV) revealed that the average V70Gy coverage decreases more for CC than for SS, when compared with PTV. On the other hand, the results for the MC algorithm imply that its use could be a step forward when compared to SS. Namely, results of the comparisons for the (PTV-GTV) volume confirm that both MCDm (mean difference = 7.36%, SD = 2.13%) and MCDw (mean difference = 6.99%, SD = 2.59%) could provide better coverage than the SS algorithm. A result of a recent study [36] confirms superiority of MC based algorithms in comparison to SS and CC algorithms, especially in GTV coverage. On the other hand, for the CC algorithm, the (PTV-GTV) V70Gy coverage decreases in comparison with absorbed dose distribution calculated by the SS algorithm (mean difference = −8.83%, SD = 3.81%) and also decreases even more than the results for the PTV (Tab. 1). This implies that most of the differences in calculated dose are located in the low-density part of the PTV considering the high mean coverage of GTV regardless of algorithm used. This result confirms original findings concerning the differences in absorbed dose over the LDE region of the semi-anthropomorphic phantom when measured values and values calculated using the CC algorithm are compared. Statistical analysis was performed for relevant OARs as well (Tab. 2). Considering the large volumes of OARs included in the analysis compared to the volumes of PTVs and GTVs in the chosen group of patients, as well as small differences in CI and GI across the compared absorbed dose distributions, the differences in mean values of analysed parameters in the OARs proved to be negligible and statistically non-significant. The results related to the average number of MUs required for the optimization of absorbed dose distributions show that there is a statistically significant difference when the number of MUs for MCDm and MCDw is compared to SS, which represents a 40% and 41% increase, respectively. This can lead to a substantial increase of treatment duration. On the other hand, there was no statistically significant difference in the number of MUs between investigated analytical algorithms. Patients with lung cancer have a large probability of inter- and intra-therapy organ and target volume movement due to the respiratory cycle. For that reason, it is necessary to ensure high coverage of target volumes with the prescribed dose and duration of the therapy as short as possible [37]. Additionally, the complexity should be minimized while maintaining all the dose goals for target volumes and organs at risk, especially when intra-therapy image guided radiotherapy (IGRT) and respiratory gating are not available. It is worth noting that in this retrospective study, target volumes and OARs in the thorax are contoured using free breathing CT and empirically determined margins, which introduces a possible substantial geometric uncertainty because the PTV margins are not individualized. For that reason, margins must be large enough to account for such geometric uncertainties. This often results in a larger irradiated volume with higher probability of damage of surrounding healthy tissue, primarily pulmonary. Consequently, lower total prescribed absorbed dose must be applied since lung toxicity is a limiting factor. Considering all the aforementioned, it can be concluded that the use of the CC algorithm for radiotherapy planning of NSCLC cases is inferior to the SS algorithm, which might result in lower quality of overall radiation therapy treatment. Regarding the results of this study for the MC algorithm in the LDE region, both reporting modes (MCDm and MCDw) show equivalent dose volume statistics for relevant TVs and OARs. Additionally, they provide higher prescribed dose coverage for the PTVs than SS or CC. Additionally, equivalent or superior dose conformity, heterogeneity for TVs and dose parameters for the OARs can be achieved, but at a cost of increased complexity of the treatment and longer radiation therapy time frames.

Conclusion

This work reports the results of a study on the clinical aspects of using different dose calculation algorithms for in-situ NSCLC radiation therapy. The three algorithms built in the Elekta XiO and Monaco treatment planning systems can technically succeed in managing the very restrictive conditions of the clinical goals according to the international standards and are appropriate for treatment planning of NSCLC tumour patients. Nevertheless, this retrospective study shows that the CC algorithm exhibits drawbacks. For this reason, CC users should bear in mind the requirement for higher prescribed dose coverage and higher CI values than the minimal recommendations to achieve adequate treatment quality for NSCLC patients. Namely, lower dose coverage of the target volume and, thus, lower absorbed dose than prescribed, can influence the inferior chances of survival in groups of patients with NSCLC [38, 39]. Additionally, the MC based algorithm used for step and shoot IMRT can fulfil the planning objectives at the expense of higher treatment complexity as well as longer treatment duration, which can also have an impact on therapy quality for NSCLC patients.
  31 in total

1.  Monte Carlo- versus pencil-beam-/collapsed-cone-dose calculation in a heterogeneous multi-layer phantom.

Authors:  Thomas Krieger; Otto A Sauer
Journal:  Phys Med Biol       Date:  2005-02-17       Impact factor: 3.609

2.  The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: implications for tumour coverage.

Authors:  Lasse Rye Aarup; Alan E Nahum; Christina Zacharatou; Trine Juhler-Nøttrup; Tommy Knöös; Håkan Nyström; Lena Specht; Elinore Wieslander; Stine S Korreman
Journal:  Radiother Oncol       Date:  2009-03-16       Impact factor: 6.280

3.  Advanced kernel methods vs. Monte Carlo-based dose calculation for high energy photon beams.

Authors:  Irina Fotina; Peter Winkler; Thomas Künzler; Jochen Reiterer; Isabell Simmat; Dietmar Georg
Journal:  Radiother Oncol       Date:  2009-11-18       Impact factor: 6.280

4.  A conformation number to quantify the degree of conformality in brachytherapy and external beam irradiation: application to the prostate.

Authors:  A van't Riet; A C Mak; M A Moerland; L H Elders; W van der Zee
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-02-01       Impact factor: 7.038

5.  Accuracy of dose calculation algorithms for static and rotational IMRT of lung cancer: A phantom study.

Authors:  Carla Sini; Sara Broggi; Claudio Fiorino; Giovanni Mauro Cattaneo; Riccardo Calandrino
Journal:  Phys Med       Date:  2015-03-20       Impact factor: 2.685

6.  Reinforcing of QA/QC programs in radiotherapy departments in Croatia: results of treatment planning system verification.

Authors:  Slaven Jurković; Manda Svabić; Ana Diklić; Deni Smilović Radojčić; Dea Dundara; Mladen Kasabašić; Ana Ivković; Dario Faj
Journal:  Med Dosim       Date:  2012-12-13       Impact factor: 1.482

7.  About the non-consistency of PTV-based prescription in lung.

Authors:  S Lebredonchel; T Lacornerie; E Rault; A Wagner; N Reynaert; F Crop
Journal:  Phys Med       Date:  2017-03-30       Impact factor: 2.685

8.  Evaluation of heterogeneity dose distributions for Stereotactic Radiotherapy (SRT): comparison of commercially available Monte Carlo dose calculation with other algorithms.

Authors:  Wataru Takahashi; Hideomi Yamashita; Naoya Saotome; Yoshio Iwai; Akira Sakumi; Akihiro Haga; Keiichi Nakagawa
Journal:  Radiat Oncol       Date:  2012-02-09       Impact factor: 3.481

9.  Dose Calculation Comparisons between Three Modern Treatment Planning Systems.

Authors:  Courtney Bosse; Ganesh Narayanasamy; Daniel Saenz; Pamela Myers; Neil Kirby; Karl Rasmussen; Panayiotis Mavroidis; Niko Papanikolaou; Sotirios Stathakis
Journal:  J Med Phys       Date:  2020-10-13

10.  Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media.

Authors:  Djeni Smilovic Radojcic; Bozidar Casar; David Rajlic; Manda Svabic Kolacio; Ignasi Mendez; Nevena Obajdin; Dea Dundara Debeljuh; Slaven Jurkovic
Journal:  Radiol Oncol       Date:  2020-09-16       Impact factor: 2.991

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.