Mechanical Characterization and Validation of the Dynamic Collimation System Prototype for Proton Radiotherapy.

Theodore Geoghegan1, Kaustubh Patwardhan1, Nicholas Nelson2, Patrick Hill3, Ryan Flynn1, Blake Smith1, Daniel Hyer1.   

Abstract

Radiation therapy is integral to cancer treatments for more than half of patients. Pencil beam scanning (PBS) proton therapy is the latest radiation therapy technology that uses a beam of protons that are magnetically steered and delivered to the tumor. One of the limiting factors of PBS accuracy is the beam cross-sectional size, similar to how a painter is only as accurate as the size of their brush allows. To address this, collimators can be used to shape the beam along the tumor edge to minimize the dose spread outside of the tumor. Under development is a dynamic collimation system (DCS) that uses two pairs of nickel trimmers that collimate the beam at the tumor periphery, limiting dose from spilling into healthy tissue. Herein, we establish the dosimetric and mechanical acceptance criteria for the DCS based on a functioning prototype and Monte Carlo methods, characterize the mechanical accuracy of the prototype, and validate that the acceptance criteria are met. From Monte Carlo simulations, we found that the trimmers must be positioned within ±0.5 mm and ±1.0 deg for the dose distributions to pass our gamma analysis. We characterized the trimmer positioners at jerk values up to 400 m/s3 and validated their accuracy to 50 μm. We measured and validated the rotational trimmer accuracy to ±0.5 deg with a FARO® ScanArm. Lastly, we calculated time penalties associated with the DCS and found that the additional time required to treat one field using the DCS varied from 25-52 s.
Copyright © 2022 by ASME.

Entities:  

Year:  2022        PMID: 35284033      PMCID: PMC8905094          DOI: 10.1115/1.4053722

Source DB:  PubMed          Journal:  J Med Device        ISSN: 1932-6181            Impact factor:   0.743


  20 in total

1.  Intensity modulation methods for proton radiotherapy.

Authors:  A Lomax
Journal:  Phys Med Biol       Date:  1999-01       Impact factor: 3.609

2.  Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking.

Authors:  J B Farr; A E Mascia; W C Hsi; C E Allgower; F Jesseph; A N Schreuder; M Wolanski; D F Nichiporov; V Anferov
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  Characterization of a mini-multileaf collimator in a proton beamline.

Authors:  J Daartz; M Bangert; M R Bussière; M Engelsman; H M Kooy
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

Review 4.  An overview of the comprehensive proton therapy machine quality assurance procedures implemented at The University of Texas M. D. Anderson Cancer Center Proton Therapy Center-Houston.

Authors:  Bijan Arjomandy; Narayan Sahoo; X Ronald Zhu; John R Zullo; Richard Y Wu; Mingping Zhu; Xiaoning Ding; Craig Martin; George Ciangaru; Michael T Gillin
Journal:  Med Phys       Date:  2009-06       Impact factor: 4.071

5.  TOPAS: an innovative proton Monte Carlo platform for research and clinical applications.

Authors:  J Perl; J Shin; J Schumann; B Faddegon; H Paganetti
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

6.  Trimmer sequencing time minimization during dynamically collimated proton therapy using a colony of cooperating agents.

Authors:  Blake R Smith; Daniel E Hyer; Ryan T Flynn; Patrick M Hill; Wesley S Culberson
Journal:  Phys Med Biol       Date:  2019-10-21       Impact factor: 3.609

7.  A method for modeling laterally asymmetric proton beamlets resulting from collimation.

Authors:  Edgar Gelover; Dongxu Wang; Patrick M Hill; Ryan T Flynn; Mingcheng Gao; Steve Laub; Mark Pankuch; Daniel E Hyer
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

Review 8.  Charged particles in radiation oncology.

Authors:  Marco Durante; Jay S Loeffler
Journal:  Nat Rev Clin Oncol       Date:  2009-12-01       Impact factor: 66.675

Review 9.  Secondary neutrons in clinical proton radiotherapy: a charged issue.

Authors:  David J Brenner; Eric J Hall
Journal:  Radiother Oncol       Date:  2008-01-14       Impact factor: 6.280

10.  Improving Head and Neck Cancer Treatments Using Dynamic Collimation in Spot Scanning Proton Therapy.

Authors:  Alexandra Moignier; Edgar Gelover; Dongxu Wang; Blake Smith; Ryan Flynn; Maura Kirk; Liyong Lin; Timothy Solberg; Alexander Lin; Daniel Hyer
Journal:  Int J Part Ther       Date:  2016-03-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.