Malfunction of cardiac devices after radiotherapy without direct exposure to ionizing radiation: mechanisms and experimental data.

Massimo Zecchin1, Gaetano Morea2, Mara Severgnini3, Elisabetta Sergi4, Anna Baratto Roldan4, Elisabetta Bianco2, Silvia Magnani2, Antonio De Luca2, Anna Zorzin Fantasia2, Luca Salvatore2, Vittorino Milan5, Gianrossano Giannini4, Gianfranco Sinagra2.   

Abstract

AIMS: Malfunctions of cardiac implantable electronical devices (CIED) have been described after high-energy radiation therapy even in the absence of direct exposure to ionizing radiation, due to diffusion of neutrons (n) causing soft errors in inner circuits. The purpose of the study was to analyse the effect of scattered radiation on different types and models of CIED and the possible sources of malfunctions. METHODS AND
RESULTS: Fifty-nine explanted CIED were placed on an anthropomorphous phantom of tissue-equivalent material, and a high-energy photon (15 MV) radiotherapy course (total dose = 70 Gy) for prostate treatment was performed. All devices were interrogated before and after radiation. Radiation dose, the electromagnetic field, and neutron fluence at the CIED site were measured. Thirty-four pacemakers (PM) and 25 implantable cardioverter-defibrillators (ICD) were analysed. No malfunctions were detected before radiation. After radiation a software malfunction was evident in 13 (52%) ICD and 6 (18%) PM; no significant electromagnetic field or photon radiations were detected in the thoracic region. Neutron capture was demonstrated by the presence of the (198)Au((197)Au + n) or (192)Ir((191)Ir + n) isotope activation; it was significantly greater in ICD than in PM and non-significantly greater in damaged devices. A greater effect in St Jude PM (2/2 damaged), Boston (9/11), and St Jude ICD (3/6) and in older ICD models was observed; the year of production was not relevant in PM.
CONCLUSION: High-energy radiation can cause different malfunctions on CIED, particularly ICD, even without direct exposure to ionizing radiation due to scattered radiation of neutrons produced by the linear accelerator. Published on behalf of the European Society of Cardiology. All rights reserved.
© The Author 2015. For permissions please email: journals.permissions@oup.com.

Entities:  

Keywords:  CIED malfunction; Implantable cardioverter-defibrillator; Pacemaker; Radiotherapy; Scattered radiation; Soft errors

Mesh:

Year:  2015        PMID: 26333380     DOI: 10.1093/europace/euv250

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  12 in total

1.  A randomized in vitro evaluation of transient and permanent cardiac implantable electronic device malfunctions following direct exposure up to 10 Gy.

Authors:  Maria Daniela Falco; Domenico Genovesi; Luciana Caravatta; Clelia Di Carlo; Ekaterina Bliakharskaia; Marianna Appignani; Massimiliano Faustino; Nanda Furia; Enrico Di Girolamo
Journal:  Strahlenther Onkol       Date:  2020-06-19       Impact factor: 3.621

2.  Radiotherapy in patients with cardiac implantable electronic devices: clinical and dosimetric aspects.

Authors:  Giulia Riva; Ombretta Alessandro; Ruggero Spoto; Annamaria Ferrari; Cristina Garibaldi; Federica Cattani; Rosa Luraschi; Elena Rondi; Nicola Colombo; Fulvio Lorenzo Francesco Giovenzana; Carlo Maria Cipolla; Mikolaj Winnicki; Martina Persiani; Fabiana Castelluccia; Massimo Sarra Fiore; Roberto Orecchia; Barbara Alicja Jereczek-Fossa
Journal:  Med Oncol       Date:  2018-04-17       Impact factor: 3.064

3.  Radiotherapy of patients with cardiac implantable electronic devices according to the DEGRO/DGK guideline-is the risk of relevant errors overestimated?

Authors:  Felix Steger; Matthias G Hautmann; Christoph Süß; Ute Hubauer; Ekrem Ücer; Lars Maier; Oliver Kölbl; Carsten Jungbauer
Journal:  Strahlenther Onkol       Date:  2019-08-09       Impact factor: 3.621

4.  Local dose rate effects in implantable cardioverter-defibrillators with flattening filter free and flattened photon radiation.

Authors:  Benjamin Gauter-Fleckenstein; Erol Tülümen; Boris Rudic; Martin Borggrefe; Martin Polednik; Jens Fleckenstein
Journal:  Strahlenther Onkol       Date:  2022-03-10       Impact factor: 4.033

5.  Abdominal implantable cardioverter-defibrillator placement in a patient requiring bilateral chest radiation therapy.

Authors:  Ankur A Karnik; Robert H Helm; Sudarshan B Chatterjee; Kevin M Monahan
Journal:  HeartRhythm Case Rep       Date:  2016-06-16

6.  The effect of stereotactic body radiotherapy (SBRT) using flattening filter-free beams on cardiac implantable electronic devices (CIEDs) in clinical situations.

Authors:  Hossein Aslian; Tomas Kron; Troy Watts; Cagla Akalanli; Nicholas Hardcastle; Peta Lonski; Atousa Montaseri; Barry Hay; James Korte; Kemal Berk; Francesco Longo; Mara Severgnini
Journal:  J Appl Clin Med Phys       Date:  2020-04-11       Impact factor: 2.102

7.  Cause of the "power-on reset" phenomenon other than electric magnetic interference in a patient with a pacemaker.

Authors:  Kyoichiro Yazaki; Masahiro Watarai; Mitsuru Kahata; Asako Kumagai; Koji Inoue; Hiroshi Koganei; Kenji Enta; Masato Otsuka; Yasuhiro Ishii
Journal:  Indian Pacing Electrophysiol J       Date:  2018-02-22

8.  Assessment of Radiation-Induced Malfunction in Cardiac Implantable Electronic Devices.

Authors:  Amin Zagzoog; Matt Wronski; David H Birnie; Cynthia Yeung; Adrian Baranchuk; Jeffrey S Healey; Mehrdad Golian; Usama Boles; Aldo G Carrizo; Suzette Turner; Ahmed Hassan; Elsayed Ali; Sharath K Kumar; Steve Russell; Mohammed Shurrab; Eugene Crystal
Journal:  CJC Open       Date:  2021-07-11

9.  Safety verification of carbon-ion radiotherapy for patients with cardiac implantable electronic devices (CIEDs).

Authors:  Naoko Okano; Makoto Sakai; Kei Shibuya; Kazuhisa Tsuda; Takao Kanzaki; Masato Sano; Yoshiaki Kaneko; Tatsuya Ohno
Journal:  J Radiat Res       Date:  2022-01-20       Impact factor: 2.724

Review 10.  Anaesthetic consideration in patients with cardiac implantable electronic devices scheduled for surgery.

Authors:  Murali Chakravarthy; Dattatreya Prabhakumar; Antony George
Journal:  Indian J Anaesth       Date:  2017-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.