| Literature DB >> 35260588 |
G Papadakis1,2, A K Pantazis3,4, N Fikas3,4, S Chatziioannidou3,4,5, V Tsiakalou3, K Michaelidou6, V Pogka7, M Megariti3, M Vardaki3,5, K Giarentis5, J Heaney8,9, E Nastouli8,9, T Karamitros7, A Mentis7, A Zafiropoulos10, G Sourvinos10, S Agelaki5,11, E Gizeli12,13.
Abstract
Loop-mediated isothermal amplification is known for its high sensitivity, specificity and tolerance to inhibiting-substances. In this work, we developed a device for performing real-time colorimetric LAMP combining the accuracy of lab-based quantitative analysis with the simplicity of point-of-care testing. The device innovation lies on the use of a plastic tube anchored vertically on a hot surface while the side walls are exposed to a mini camera able to take snapshots of the colour change in real time during LAMP amplification. Competitive features are the rapid analysis (< 30 min), quantification over 9 log-units, crude sample-compatibility (saliva, tissue, swabs), low detection limit (< 5 copies/reaction), smartphone-operation, fast prototyping (3D-printing) and ability to select the dye of interest (Phenol red, HNB). The device's clinical utility is demonstrated in cancer mutations-analysis during the detection of 0.01% of BRAF-V600E-to-wild-type molecules from tissue samples and COVID-19 testing with 97% (Ct < 36.8) and 98% (Ct < 30) sensitivity when using extracted RNA and nasopharyngeal-swabs, respectively. The device high technology-readiness-level makes it a suitable platform for performing any colorimetric LAMP assay; moreover, its simple and inexpensive fabrication holds promise for fast deployment and application in global diagnostics.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35260588 PMCID: PMC8904468 DOI: 10.1038/s41598-022-06632-7
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1(a) Overview of the real-time quantitative colorimetric LAMP (qcLAMP) concept. Bottom: (b) Schematic representation of the qcLAMP device and components. (c) Photograph of the device and tubes-holder. (d) Captured image of two Eppendorf-tubes inside the chamber and during operation of the device. (e) Image of the handheld device for performing qcLAMP. (f) Three prototype devices of various colors; parallel production of two boxes requires ~ 12 h.
Figure 2(a) Real-time colorimetric LAMP reactions performed inside the device using the phenol red indicator. The positive reaction (blue line) contained 10 lysed bacteria as template while the negative control contained no bacteria (orange line). (b) Similar reactions to (a), this time performed with the HNB indicator. In both cases, the heating process started at t = 0 min and the heater reached the required temperature (63 °C) after approximately 2 min. All reactions were stopped after 30 min. (c) Left: Real time colorimetric curves monitored during qcLAMP amplification performed with 2 different bacteria concentrations (400 and 1000 CFU/ml) spiked in saliva samples. The two concentrations were chosen based on previously reported detection limits of bacteria using end-point colorimetric LAMP or biosensors (13, 54). Both were successfully detected in less than 17 min while zero background signal was monitored for the negative controls. Right, top: Picture of end-point reactions with 0 and 1000 CFU/ml after 40 min of incubation. Right bottom: Picture of end-point reactions with 0 and 400 CFU/ml after 40 min of incubation.
Figure 3(a) Real-time colorimetric curves corresponding to mut:wt starting template ratios spanning more than four orders of magnitude. (b) Calibration curve produced from real-time data of triplicate colorimetric LAMP measurements with error bars representing standard deviation; y = − 1.165ln(x) + 23.72 (R2 = 0.99). (c) Real-time data corresponding to tissue samples of various mut:wt ratios. The 1% was not consistently produced (yellow vs green line). (d) Comparison of real-time data for 10, 20 and 50% mut:wt tissue samples measured directly in the qcLAMPdevice without DNA purification. (e) Real time colorimetric LAMP results of 12 patient samples analyzed with the qcLAMP 3D device (blind tests). (f) Comparison of ddPCR, Sanger and qcLAMP test results of 12 clinical samples. The ddPCR method provided additional information regarding the concentration of mutant and wild type copies per analyzed sample.
Figure 4(a) Calibration curve using Influenza A DNA template ranging from 101 to 109 copies per reaction; error bars represent standard deviation of at least triplicates (b) Serial dilutions of a SARS-COV-2 positive sample with a reported Ct value of 19 were measured with the qcLAMP method. (c) Correlation (R2 = 0.99) between the qcLAMP time-to-positive results and the viral RNA concentration based on a tenfold serial dilutions. (d) Scatter plot of the Ct values (ranging from 16 to ~ 37) for 38 positive samples versus the qcLAMP time-to-positive (ranging from 12.5 to 23 min). qcLAMP measurements were performed in different days within 3 weeks using stored (− 80 °C) RNA samples. Only one point was missed by qCLAMP in the first run corresponding to a Ct of 35. (e) Diagnostic results of SARS-CoV-2 qcLAMP assay-evaluation with 89 clinical samples including 38 COVID-19 positive and 51 negative. An overall sensitivity of 97% (95% CI: 93–100), specificity of 100% and negative likelihood ratio of 0.026 (95% CI: 0.004–0.182) were calculated.
Figure 5(a) Schematic representation of the workflow of the SARS-CoV-2 swab-to-qcLAMP detection, i.e. without RNA purification. (b) Scattered plot of the Ct values of PCR-confirmed positive samples by using the HNB colorimetric dye inside the LAMP-mix. (c) Table presenting overall diagnostic sensitivity and specificity of the qcLAMP for fresh and frozen samples; the red and blue colors are used to guide the eye and depict fresh and frozen samples, respectively. (d) when considering various Ct cut off values. For Ct < 30 sensitivity becomes 98% (95% CI: 95–101) and negative likelihood ratio 0.016 (95% CI: 0.002–0.109); for Ct < 25, sensitivity is calculated to be 100%.
Comparison of the qcLAMP, Sanger sequencing, ddPCR and qRT-PCR methods for the two applications.
| SARS-CoV-2 ASSAY | |||||||
|---|---|---|---|---|---|---|---|
| qcLAMP | Sanger | ddPCR | qcLAMPcrude | qRT-PCR1 | qcLAMP | qRT-PCR2 | |
| Steps | 1 | 2 | 2 | 1 | 2 | 1 | 2 |
| Temperature | 65 °C | multiple | multiple | 65 °C | 53 °C (RT) (60 °C–95 °C) × 40 cycles | 65 °C | 50 °C (RT) (55 °C–95 °C) × 45 cycles |
| Assay time | N/A | N/A | N/A | 30 min | 90 min | 30 min | 80 min |
| Sample-to-result time | < 2 h | > 24 h | > 2 h | < 40 min POC | > 3 h lab-based | 2 h lab-based | > 3 h lab-based |
| Target/control | V600E/WT | BRAF V600E/WT | |||||
| Preparatory steps | 1 (DNA extraction or tissue heating) | 2 (DNA extraction & thermal cycling) | 2 (DNA extraction & droplets formation) | Crude N/O-P sample | NA extraction | NA extraction | NA extraction |
| Analytical performance | 0.01% (mut:wt ratio) | 5% (mut:wt ratio) | 0.001% (mut:wt ratio) | Not available | 1 copy/reaction | 5 copies/ reaction | Not available |
PCR1 and PCR2 refer to the different protocols used in the two validation sites (Lab. of Clinical Virology and Hellenic Pasteur Institute, respectively).