Planning Risk-Based SQC Schedules for Bracketed Operation of Continuous Production Analyzers.

James O Westgard1,2, Hassan Bayat3, Sten A Westgard2.   

Abstract

BACKGROUND: To minimize patient risk, "bracketed" statistical quality control (SQC) is recommended in the new CLSI guidelines for SQC (C24-Ed4). Bracketed SQC requires that a QC event both precedes and follows (brackets) a group of patient samples. In optimizing a QC schedule, the frequency of QC or run size becomes an important planning consideration to maintain quality and also facilitate responsive reporting of results from continuous operation of high production analytic systems.
METHODS: Different plans for optimizing a bracketed SQC schedule were investigated on the basis of Parvin's model for patient risk and CLSI C24-Ed4's recommendations for establishing QC schedules. A Sigma-metric run size nomogram was used to evaluate different QC schedules for processes of different sigma performance.
RESULTS: For high Sigma performance, an effective SQC approach is to employ a multistage QC procedure utilizing a "startup" design at the beginning of production and a "monitor" design periodically throughout production. Example QC schedules are illustrated for applications with measurement procedures having 6-σ, 5-σ, and 4-σ performance.
CONCLUSIONS: Continuous production analyzers that demonstrate high σ performance can be effectively controlled with multistage SQC designs that employ a startup QC event followed by periodic monitoring or bracketing QC events. Such designs can be optimized to minimize the risk of harm to patients.
© 2017 American Association for Clinical Chemistry.

Entities:  

Mesh:

Year:  2017        PMID: 29097516     DOI: 10.1373/clinchem.2017.278291

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  14 in total

1.  Sigma metrics for assessing the analytical quality of clinical chemistry assays: a comparison of two approaches: Electronic supplementary material available online for this article.

Authors:  Xiuzhi Guo; Tianjiao Zhang; Xuehui Gao; Pengchang Li; Tingting You; Qiong Wu; Jie Wu; Fang Zhao; Liangyu Xia; Ermu Xu; Ling Qiu; Xinqi Cheng
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

Review 2.  Analytical Sigma metrics: A review of Six Sigma implementation tools for medical laboratories.

Authors:  Sten Westgard; Hassan Bayat; James O Westgard
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

3.  Moving Rate of Positive Patient Results as a Quality Control Tool for High-Sensitivity Cardiac Troponin T Assays.

Authors:  Tingting Li; Shunwang Cao; Yi Wang; Yujuan Xiong; Yuting He; Peifeng Ke; Xianzhang Huang
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

4.  Optimization and validation of patient-based real-time quality control procedure using moving average and average of normals with multi-rules for TT3, TT4, FT3, FT3, and TSH on three analyzers.

Authors:  Chao Song; Jun Zhou; Jun Xia; Deli Ye; Qian Chen; Weixing Li
Journal:  J Clin Lab Anal       Date:  2020-05-03       Impact factor: 2.352

Review 5.  Biological variation: Understanding why it is so important?

Authors:  Tony Badrick
Journal:  Pract Lab Med       Date:  2021-01-04

6.  Risk analysis and assessment based on Sigma metrics and intended use.

Authors:  Yong Xia; Hao Xue; Cunliang Yan; Bowen Li; ShuQiong Zhang; Mingyang Li; Ling Ji
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

7.  Optimizing moving average control procedures for small-volume laboratories: can it be done?

Authors:  Vera Lukić; Svetlana Ignjatović
Journal:  Biochem Med (Zagreb)       Date:  2019-10-15       Impact factor: 2.313

8.  Comparative analysis of calculating sigma metrics by a trueness verification proficiency testing-based approach and an internal quality control data inter-laboratory comparison-based approach.

Authors:  Runqing Li; Tengjiao Wang; Lijun Gong; Peng Peng; Song Yang; Haibin Zhao; Pan Xiong
Journal:  J Clin Lab Anal       Date:  2019-08-06       Impact factor: 2.352

9.  Evaluation of the clinical chemistry tests analytical performance with Sigma Metric by using different quality specifications - Comparison of analyser actual performance with manufacturer data.

Authors:  Murat Keleş
Journal:  Biochem Med (Zagreb)       Date:  2021-12-15       Impact factor: 2.313

10.  Comparison and optimization of various moving patient-based real-time quality control procedures for serum sodium.

Authors:  Yuanyuan Li; Qian Yu; Xiaoyan Zhang; Xiaoling Chen
Journal:  J Clin Lab Anal       Date:  2021-09-14       Impact factor: 2.352

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.