Shaun Pratt1, Tamsin S Barnes1,2, Nicholas Cowling1, Karla de Klerk1, Joanne Rainger1, Albert Sole-Guitart1, Solomon Woldeyohannes1, Wendy Goodwin1.
Abstract
To compare arterial blood pressure (ABP) measured invasively (IBP) to ABP measured non-invasively (NIBP) via oscillometry in healthy anaesthetised and standing horses using the Bionet BM7Vet. Fourteen horses were anaesthetised for elective procedures (anaesthetised group) and 10 horses were enrolled for standing blood pressure manipulation (standing group). In both groups, IBP and NIBP-corrected to heart level were measured every 3 min using the Bionet BM7Vet. The overall mean difference (bias), standard deviation and limits of agreement (LOA) were calculated for paired IBP and NIBP systolic (SAP), mean (MAP) and diastolic (DAP) blood pressure measurements. In anaesthetised horses, the NIBP cuff was placed at either the proximal tail base or the metacarpus. Invasive MAP was used to retrospectively characterise measurements into hypotensive (≤70 mm Hg), normotensive (71-110 mm Hg) or hypertensive (≥111 mm Hg) subgroups. In standing horses, the NIBP cuff was placed at the tail base only and invasive MAP was manipulated to achieve hypertension (≥126 mm Hg) and hypotension (≤90 mm Hg) using phenylephrine and acepromazine, respectively. When measuring NIBP at the tail in anaesthetised horses, the Bionet BM7Vet failed on 8/185 occasions and overestimated SAP, MAP and DAP during hypotension and normotension. The biases (lower, upper LOA) for MAP were -11.4 (-33.3, 10.5) and -6.0 (-25.8, 13.8) mm Hg, respectively. Hypertension could not be evaluated. When measuring NIBP at the metacarpus in anaesthetised horses, the Bionet BM7Vet failed on 24/65 occasions and underestimated SAP, MAP and DAP when all ABP subgroups were combined. The bias (lower, upper LOA) for pooled MAP was 3.6 (-44.3, 51.6) mm Hg. When measuring NIBP at the tail in standing horses, the Bionet BM7Vet failed on 64/268 occasions and underestimated SAP, MAP and DAP during hypotension, normotension and hypertension. The biases (lower, upper LOA) for MAP were 16.3 (-10.5, 43.1), 16.6 (-19.5, 52.7) and 30.0 (-8.1, 68.0) mm Hg, respectively. Monitoring NIBP on the Bionet BM7Vet in anaesthetised horses overestimated ABP at the tail and underestimated ABP at the metacarpus. The device inaccurately detected hypotension and should be used cautiously. In standing horses, the Bionet BM7Vet underestimated ABP at the tail, especially during pharmacologically induced hypertension.Entities:
Keywords: ABP; Bionet BM7Vet; IBP; NIBP; anaesthesia; horse
Year: 2022 PMID: 35202305 PMCID: PMC8878245 DOI: 10.3390/vetsci9020052
Source DB: PubMed Journal: Vet Sci ISSN: 2306-7381
Arterial blood pressure state classification.
IBP arterial blood pressure measured invasively, MAP mean arterial blood pressure. * Induced with acepromazine 0.05–0.1 mg/kg intravenous. # Induced with phenylephrine hydrochloride 1–5 mcg/kg/m intravenous constant rate infusion.
Figure 1Flow diaphragm for comparing arterial blood pressure measured non-invasively to arterial blood pressure measured invasively using the Bionet BM7Vet in 14 healthy anaesthetised and 9 standing horses. Failures occurred when the Bionet BM7Vet failed to generate a non-invasive arterial blood pressure measurement by the completion of the inflation cycle. Paired measurements were excluded from the analysis when the pulse rate measured by the non-invasive method differed by ≥20% to the true pulse rate. Anaesthetised group: hypotension ≤ 70 mm Hg, normotension 71–110 mm Hg, hypertension ≥ 111 mm Hg. Standing group: hypotension ≤ 90 mm Hg, normotension 91–125 mm Hg, hypertension ≥ 126 mm Hg.
Bland–Altmann analyses for 218 paired arterial blood pressure measurements made non-invasively compared to those made invasively using the Bionet BM7Vet in 14 healthy, anaesthetised horses.
| No. Paired Observations | No. Horses | Bias | SD | LOA-L | LOA-U | ||
|---|---|---|---|---|---|---|---|
|
| |||||||
| Systolic | Pooled | 177 | 14 | −12.9 | 14.1 | −41.0 | 15.2 |
| Hypotension | 42 | 9 | −15.4 | 13.9 | −43.1 | 12.4 | |
| Normotension | 133 | 14 | −12.8 | 14.4 | −41.5 | 16.0 | |
| Hypertension | 2 | 1 | |||||
| Mean | Pooled | 177 | 14 | −6.9 | 10.7 | −28.3 | 14.5 |
| Hypotension | 42 | 9 | −11.4 | 11.0 | −33.3 | 10.5 | |
| Normotension | 133 | 14 | −6.0 | 9.9 | −25.8 | 13.8 | |
| Hypertension | 2 | 1 | |||||
| Diastolic | Pooled | 177 | 14 | −5.4 | 10.3 | −25.9 | 15.2 |
| Hypotension | 42 | 9 | −10.2 | 10.4 | −30.9 | 10.6 | |
| Normotension | 133 | 14 | −4.0 | 9.7 | −23.4 | 15.4 | |
| Hypertension | 2 | 1 | |||||
|
| |||||||
| Systolic | 41 | 9 | 9.0 | 22.1 | −35.1 | 53.2 | |
| Mean | 41 | 9 | 3.6 | 24.0 | −44.3 | 51.6 | |
| Diastolic | 41 | 9 | 0.8 | 26.0 | −51.1 | 52.7 | |
No. number, SD standard deviation, LOA limits of agreement, U upper and L lower, MAP mean arterial pressure. Bias defines accuracy and SD defines precision. The non-invasive arterial blood pressure cuff was placed either at the proximal tail base or the metacarpus. Combined recordings on the proximal tail base (pooled) were separated into hypotension (MAP ≤ 70 mm Hg), normotension (71 ≤ MAP ≤ 110 mm Hg) and hypertension (MAP ≥ 111 mmHg).
Figure 2Bland–Altman plot of agreement for 177 paired mean arterial blood pressure measurements (mm Hg) made non-invasively compared to those made invasively using the Bionet BM7Vet in 14 healthy, anaesthetised horses. The non-invasive arterial blood pressure cuff was placed at the proximal tail base. Repeated observations from the same horse are plotted using the same colour-shape combination. The solid black line indicates mean bias, and the dashed lines indicate lower and upper limits of agreement. The solid red line represents zero bias. IBP invasive blood pressure, NIBP non-invasive blood pressure. Bias defines accuracy.
Figure 3Bland–Altman plot of agreement for 41 paired mean arterial blood pressure measurements (mm Hg) made non-invasively compared to those made invasively using the Bionet BM7Vet in 9 healthy, anaesthetised horses. The non-invasive arterial blood pressure cuff was placed at the metacarpus. Repeated observations from the same horse are plotted using the same colour-shape combination. The solid black line indicates mean bias, and the dashed lines indicate lower and upper limits of agreement. The solid red line represents zero bias. IBP invasive blood pressure, NIBP non-invasive blood pressure. Bias defines accuracy.
Bland–Altmann analyses for 179 paired arterial blood pressure measurements made non-invasively compared to those made invasively using the Bionet BM7Vet in 9 healthy, standing horses.
| Parameter | Tension State | No. Paired Observations | No. Horses | Bias | SD | LOA-L | LOA-U |
|---|---|---|---|---|---|---|---|
| Systolic | Pooled | 179 | 9 | 15.7 | 23.0 | −30.4 | 61.7 |
| Hypotension | 56 | 8 | 10.7 | 17.9 | −25.1 | 46.6 | |
| Normotension | 82 | 9 | 14.4 | 24.4 | −34.5 | 63.3 | |
| Hypertension | 41 | 9 | 23.7 | 24.1 | −24.5 | 71.8 | |
| Mean | Pooled | 179 | 9 | 19.9 | 17.4 | −14.9 | 54.7 |
| Hypotension | 56 | 8 | 16.3 | 13.4 | −10.5 | 43.1 | |
| Normotension | 82 | 9 | 16.6 | 18.0 | −19.5 | 52.7 | |
| Hypertension | 41 | 9 | 30.0 | 19.0 | −8.1 | 68.0 | |
| Diastolic | Pooled | 179 | 9 | 18.4 | 18.2 | −18.0 | 54.7 |
| Hypotension | 56 | 8 | 17.3 | 14.3 | −11.3 | 46.0 | |
| Normotension | 82 | 9 | 14.3 | 18.1 | −21.8 | 50.4 | |
| Hypertension | 41 | 9 | 26.2 | 22.9 | −19.6 | 71.9 |
No. number, SD standard deviation, LOA limits of agreement, U upper and L lower, MAP mean arterial pressure. Bias defines accuracy and SD defines precision. The non-invasive arterial blood pressure cuff was placed at the proximal tail base. Combined recordings (pooled) were separated into hypotension (MAP ≤ 90 mm Hg), normotension (91 ≤ MAP ≤ 125 mm Hg) and hypertension (MAP ≥ 126 mmHg). Hypotension was induced with acepromazine 0.05–0.1 mg/kg intravenous and hypertension was induced with phenylephrine hydrochloride 1–5 mcg/kg/m intravenous constant rate infusion.
Figure 4Bland–Altman plot of agreement for 179 paired mean arterial blood pressure measurements (mm Hg) made non-invasively compared to those made invasively using the Bionet BM7Vet in 9 healthy, standing horses. The non-invasive arterial blood pressure cuff was placed at the proximal tail base. Repeated observations from the same horse are plotted using the same colour-shape combination. The solid black line indicates mean bias, and the dashed lines indicate lower and upper limits of agreement. The solid red line represents zero bias. IBP invasive blood pressure, NIBP non-invasive blood pressure. Bias defines accuracy.