Marcus Chun Jin Tan1, Heng Yoong Chloe Choy2, Victor Koh Teck Chang1, Maria Cecilia Aquino1, Chelvin Cheryl Agnes Sng1, Dawn Ka Ann Lim1, Seng Chee Loon1, Paul Chew Tec Kuan1.
Abstract
PURPOSE: To determine 2-year efficacy of the PAUL Glaucoma Implant (PGI), a novel glaucoma tube shunt in patients with advanced glaucoma. PARTICIPANTS: Patients with glaucoma refractory to maximum medical therapy or previous failed glaucoma surgery.Entities:
Mesh:
Year: 2022 PMID: 35180153 PMCID: PMC9148669 DOI: 10.1097/IJG.0000000000001998
Source DB: PubMed Journal: J Glaucoma ISSN: 1057-0829 Impact factor: 2.290
FIGURE 1Picture of the Paul Glaucoma Implant (Advanced Ophthalmic Innovations, Singapore) with relevant dimensions.
Baseline Demographic and Ocular Characteristics of Enrolled Participants (n=45)
| Age (y), mean±SD | 64.3±12.2 |
| Sex, n (%) | |
| Male | 36 (80) |
| Female | 9 (20) |
| Ethnicity, n (%) | |
| Chinese | 34 (75.5) |
| Malay | 7 (15.6) |
| Indian | 2 (4.4) |
| Others | 2 (4.4) |
| Mean visual acuity (logMAR), mean±SD | 0.82±1.0 |
| Mean medicated intraocular pressure (mm Hg), mean±SD | 19.8±6.3 |
| Mean vertical cup to disc ratio, mean±SD | 0.72±0.21 |
| Lens status, n (%) | |
| Phakic | 31 (68.9) |
| Pseudophakic | 14 (31.1) |
| No. classes of intraocular pressure-lowering medications, n (%) | |
| 1 | 1 (2.2) |
| 2 | 8 (17.8) |
| 3 | 15 (33.3) |
| 4 | 21 (46.7) |
| Mean±SD | 3.2±0.8 |
| Diagnosis, n (%) | |
| Primary open-angle glaucoma | 19 (42.2) |
| Primary angle-closure glaucoma | 11 (24.4) |
| Neovascular glaucoma | 4 (8.9) |
| Others | 11 (24.4) |
| Previous glaucoma procedures, n (%) | |
| Trabeculectomy | 9 (20.0) |
| Glaucoma tube shunt | 7 (15.5) |
Others=silicone oil related (n=1), hyphema related (n=1), anterior cleavage syndrome (n=1), uveitis related (n=1), inflammatory iris chafing (n=1), Ice syndrome (n=2), angle recession (n=1), juvenile open angle glaucoma (n=2), subluxed morcher (n=1).
logMAR indicates logarithm of the minimum angle of resolution.
FIGURE 2Kaplan-Meier survival curve for eyes implanted with the PAUL Glaucoma Implant (Advanced Ophthalmic Innovations, Singapore) over 2 years. Time to failure was defined as time from implantation to the first of 2 consecutive study visits after 3 months in which IOP <5 mm Hg, or >18 mm Hg, loss of light perception or to reoperation for glaucoma or implant related complications.
FIGURE 3Line graph showing intraocular (IOP) trend before and after surgery up to 2 years, compared to both the highest preoperative IOP and preoperative medicated IOPs. The error bars indicate 95% confidence limits.
Postoperative Outcomes of Patients (n=45)
| Outcome | Data (Mean±SD) |
|---|---|
| Mean visual acuity (logMAR) | |
| Preoperative | 0.82±0.94 |
| 12 mo | 0.80±0.97 |
| 24 mo | 0.96±1.17 |
| Mean intraocular pressure (mm Hg) | |
| Preoperative | 19.8±6.3 |
| 1 mo | 13.0±5.4 |
| 6 mo | 14.0±3.8 |
| 12 mo | 14.2±3.5 |
| 18 mo | 13.9±4.3 |
| 24 mo | 13.9±3.7 |
| Mean number of classes of intraocular pressure lowering medications | |
| Preoperative | 3.2±0.8 |
| 12 mo | 0.07±0.25 |
| 18 mo | 0.07±0.25 |
| 24 mo | 0.29±0.65 |
P<0.01.
FIGURE 4Line graph showing the number of intraocular pressure (IOP) lowering medications used before and after surgery. Error bars indicate 95% confidence limits.
Comparison Between the PGI, AGV, and BGI
| PGI 2 y | ABC 5 y | AVB 5 y | AVB 2 y | ||||
|---|---|---|---|---|---|---|---|
| PGI (n=45) | AGV (n=143) | BGI (n=133) | AGV (n=124) | BGI (n=114) | AGV (n=129) | BGI (n=121) | |
| Failure | 8 (17.8%) | 64 (44.7%) | 52 (39.4%) | 53% | 40% | 62 (48.4%) | 54 (44.2%) |
| Mean IOP | 13.9±3.7 | 14.7±4.4 | 12.7±4.5 | 16.6±5.9 | 13.6±5.0 | 14.4±6.7 | 13.7±4.9 |
| No. med | 0.29±0.65 | 2.2±1.4 | 1.8±1.5 | 1.8±1.5 | 1.2±1.3 | 1.43±1.49 | 0.77±1.06 |
| Complications requiring intervention | 4 (8.9%) | 11 (7.7%) | 22 (16.5%) | 63 (51%) | 58 (51%) | 22 (17.1%) | 31 (25.6%) |
ABC indicates Ahmed Baerveldt Comparison; AGV, Ahmed Glaucoma Valve; AVB, Ahmed Versus Baerveldt; BGI, Baerveldt Glaucoma Implant; IOP, intraocular pressure; PGI, PAUL Glaucoma Implant.