Performance Evaluation of the Abbott Alinity hs Blood Slide Maker/Stainer.

Min-Sun Kwak1,2, In-Hwa Jeong1,2, Sung-Sook Cho2, Kwang-Sook Woo1,2, Jin-Yeong Han1,2.   

Abstract

Entities:  

Year:  2022        PMID: 35177570      PMCID: PMC8859557          DOI: 10.3343/alm.2022.42.4.482

Source DB:  PubMed          Journal:  Ann Lab Med        ISSN: 2234-3806            Impact factor:   4.941


× No keyword cloud information.
Dear Editor, Abbott Laboratories (Abbott Diagnostics, Santa Clara, CA, USA) recently released the Alinity h-series for high-throughput, automated hematology analysis [1, 2]. Since the introduction of automated slide makers and stainers in the late 1970s, only a few studies have evaluated the performance of automated slide makers [3-6]. We evaluated the performance of the Abbott Alinity hs automated slide maker and stainer and compared the white blood cell (WBC) differential counts between slides prepared with Alinity hs and manual smears. The Institutional Review Board of Dong-A University Hospital, Busan, Korea, approved this study (DAUHIRB-19-194). Peripheral blood specimens from 50 healthy controls and 115 patients referred to the Department of Laboratory Medicine, Dong-A University Hospital between September 2019 and February 2020 were collected in K3-EDTA VACUETTE blood collection tubes (Greiner Bio-One GmbH, Kremsmunster, Austria). We included at least one third to half the total number of abnormal (patient) specimens, as suggested by the CLSI and International Council for Standardization in Hematology guidelines [7, 8]. The blood specimens were processed within four hours of phlebotomy, and the residuals after laboratory tests were used for the study. All automated complete blood cell count (CBC) and WBC differential counts were obtained using the Abbott Alinity hq analyzer. We analyzed the blood smear quality, repeatability, and cellular carry-over and compared WBC differential counts with those obtained using manual smears [7, 8]. Both Alinity hs and manual smears were stained using Abbott Wright-Giemsa staining reagents and the Alinity hs stainer [1]. Statistical analyses were performed using MedCalc (version 14.8.1; MedCalc Software Ltd., Ostend, Belgium) and EP Evaluator Release 11 (Data Innovation LLC, South Burlington, VT, USA). Agreement and the interpretation criteria were adapted from the manufacturer’s acceptance limits and state-of-the-art criteria [1, 7]. Macroscopic examination revealed that all 165 Alinity hs smears showed an acceptable quality in terms of length, margins, and thickness, with a smooth and gradual transition from thick to thin. When the grading scheme was applied, all 50 normal (100%) and 107 abnormal (93.0%) specimens were graded as “slight precipitation.” Six (5.2%) and two (1.7%) abnormal specimens showed moderate and heavy precipitation, respectively; however, all specimens were judged acceptable for morphological assessment and WBC differential count. The distribution of monocytes between the margins and body of the smears was assessed in 50 normal and 106 abnormal specimens with a total WBC count >3×109/L (which is above the reference range, 3,000–9,300×109/L, used in the laboratory). In all normal (100%) and 104/106 abnormal (98.1%) specimens, the number of monocytes in the margins did not exceed that in the body of the smear by >2–3 times, which is regarded acceptable [4, 5]. Nine abnormal specimens showed a WBC count <3×109/L (below the lower reference intervals) and were excluded from the analysis. We examined repeatability using 10 blood smears prepared from five normal blood specimens with the Alinity hs, using a manual 200-cell WBC differential count. All smears showed acceptable SD values ranging 1.7–2.9 for neutrophils, 1.4–2.6 for lymphocytes, and 0.6–1.9 for monocytes (Table 1) [4].
Table 1

Repeatability based on five abnormal (patient) specimens with normal CBC results

Neutrophils (%) (mean ± SD)Lymphocytes (%) (mean ± SD)Monocytes (%) (mean ± SD)
Specimen 171.4 ± 1.723.7 ± 1.44.3 ± 1.0
Specimen 266.9 ± 2.124.0 ± 1.76.3 ± 0.9
Specimen 361.0 ± 2.929.2 ± 2.47.1 ± 1.9
Specimen 469.4 ± 2.327.7 ± 2.61.6 ± 0.6
Specimen 571.8 ± 1.820.5 ± 1.84.5 ± 1.2

Abbreviation: CBC, complete blood cell count.

Two specimens with numerous smudge cells (100 smudge cells on average based on the 200-cell differential count) and leukemic promyelocytes (98.7% promyelocytes on average based on the 200-cell differential count) showed no specific smudge or promyelocytic cells, respectively, in three high-low runs. The size, shape, Hb stainability, and distribution of red blood cells were comparable between Alinity hs and manual smears in all the 50 normal and 115 abnormal blood specimens evaluated. Morphological assessment revealed that platelet size, granularity, and the presence of clumps were also comparable. A 400-cell WBC differential count by two examiners revealed a remarkable accordance between Alinity hs and manual smears for the major WBC types (Fig. 1). Correlations in the neutrophil, lymphocyte, and monocyte percentages between hs and manual smears were excellent, with a correlation coefficient (r) of 0.961, 0.955, and 0.93, respectively. Eosinophils and basophils showed excellent or moderate correlation between hs and manual smears, with r of 0.934 and 0.684, respectively (data not shown), in normal and abnormal specimens.
Fig. 1

Comparison of 400-cell WBC differential count results between Abbott Alinity hs and manual smears. (A) Neutrophils, (B) Lymphocytes, (C) Monocytes. Blue lines indicate 95% confidence intervals.

Abbreviation: WBC, white blood cell.

Even in the era of digital morphology, slide makers and stainers have not been thoroughly evaluated for performance [7, 9]. To the best of our knowledge, this was the first comprehensive evaluation of the Abbott Alinity hs slide maker and stainer. The results demonstrate that the systems perform well in terms of blood smear quality, repeatability, and cellular carry-over and that comparable results with manual smears in routine laboratory work for major WBC differential counts.
  8 in total

1.  Initial performance evaluation of the UniCel® DxH slide maker/stainer Coulter® cellular analysis system.

Authors:  W Brown; M Keeney; B D Hedley
Journal:  Int J Lab Hematol       Date:  2013-09-13       Impact factor: 2.877

2.  Performance evaluation of new Abbott Alinity hq hematology analyzer.

Authors:  Kwang-Sook Woo; In-Hwa Jeong; Gyu-Dae An; Hyeon-Ho Lim; Jin-Yeong Han
Journal:  Int J Lab Hematol       Date:  2019-07-04       Impact factor: 2.877

3.  Performance evaluation of the prototype Abbott Alinity hq hematology analyzer.

Authors:  Yolien Van der Beken; Annelien Van Dalem; Guy Van Moer; Eddy Segers; Serge Damiaens; Johannes Hoffmann; Gabriella Lakos; Kristin Jochmans
Journal:  Int J Lab Hematol       Date:  2019-05-21       Impact factor: 2.877

4.  Automatic wedge smears preparation may cause traumatic morphological changes in peripheral blood cells.

Authors:  Antonio La Gioia; Maurizio Fumi; Paola Pezzati; Fiamma Balboni; Ylenia Pancione; Lucia Rocco; Silvia Sale; Maria Bombara; Marcello Fiorini; Fabiana Fiorini; Vincenzo Rocco
Journal:  J Clin Pathol       Date:  2017-08-03       Impact factor: 3.411

5.  Comparison of three staining methods in the automated digital cell imaging analyzer Sysmex DI-60.

Authors:  Hyeong Nyeon Kim; Mina Hur; Hanah Kim; Mikyoung Park; Seung Wan Kim; Hee-Won Moon; Yeo-Min Yun
Journal:  Clin Chem Lab Med       Date:  2018-11-27       Impact factor: 3.694

6.  ICSH guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting.

Authors:  C Briggs; N Culp; B Davis; G d'Onofrio; G Zini; S J Machin
Journal:  Int J Lab Hematol       Date:  2014-03-25       Impact factor: 2.877

7.  Performance of automated slidemakers and stainers in a working laboratory environment - routine operation and quality control.

Authors:  E Simson; M G Gascon-Lema; D L Brown
Journal:  Int J Lab Hematol       Date:  2009-02-11       Impact factor: 2.877

8.  Validation of the Sysmex sp-1000i automated slide preparer-stainer in a clinical laboratory.

Authors:  Eberson Damião Dos Santos de Bitencourt; Carlos Franco Voegeli; Gabriela Dos Santos Onzi; Sara Cardoso Boscato; Carine Ghem; Terezinha Munhoz
Journal:  Rev Bras Hematol Hemoter       Date:  2013
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.