Literature DB >> 24028789

Initial performance evaluation of the UniCel® DxH slide maker/stainer Coulter® cellular analysis system.

W Brown1, M Keeney, B D Hedley.   

Abstract

INTRODUCTION: Despite the advancements in instrumentation within hematology laboratories, there is still a need for review of a peripheral blood film (PBF). For a thorough PBF evaluation, it is critical that a well spread and stained film is available.
METHODS: In this study, we evaluated an automatic slide maker/stainer (DxH-SMS, Beckman Coulter) compared with manually prepared blood films on 124 normal and abnormal samples. The primary goal of the study was to determine whether or not the DxH-SMS was able to consistently and reproducibly prepare and stain blood films of exemplary quality, without carryover between specimens. Additionally, repeatability of white blood cell distribution, comparability of morphology to reference methodologies, and grading of acceptance criteria outlined in the CLSI document H20-A2 were assessed.
RESULTS: Carryover was not an issue and repeatability was within expected limits. There was excellent agreement of the 5-part differential between the automated blood films made by the DxH-SMS compared with the manually prepared reference blood film. There was no difference in identification and enumeration of blasts, variant lymphocytes, or nucleated red blood cells (P < 0.05). Red cell morphology showed excellent agreement.
CONCLUSION: Blood films prepared by the DxH-SMS are of excellent quality, reproducible, and compare well with manually prepared slides. Introduction to our laboratory has improved and standardized slide quality.
© 2013 John Wiley & Sons Ltd.

Entities:  

Keywords:  Automated slidemaker/stainer; blood film quality; peripheral blood film; stain quality

Mesh:

Year:  2013        PMID: 24028789     DOI: 10.1111/ijlh.12150

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Performance evaluation of Mindray CAL 8000(BC-6800 and SC-120) hematology analyzer and slidemaker/stainer.

Authors:  Hwan Tae Lee; Pil-Whan Park; Yiel-Hea Seo; Kyung-Hee Kim; Ja Young Seo; Ji-Hun Jeong; Moon Jin Kim; Jeong-Yeal Ahn
Journal:  J Clin Lab Anal       Date:  2016-09-23       Impact factor: 2.352

2.  Performance Evaluation of the Abbott Alinity hs Blood Slide Maker/Stainer.

Authors:  Min-Sun Kwak; In-Hwa Jeong; Sung-Sook Cho; Kwang-Sook Woo; Jin-Yeong Han
Journal:  Ann Lab Med       Date:  2022-07-01       Impact factor: 4.941

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.