Performance evaluation of new Abbott Alinity hq hematology analyzer.

Kwang-Sook Woo1, In-Hwa Jeong1, Gyu-Dae An1, Hyeon-Ho Lim1, Jin-Yeong Han1.   

Abstract

INTRODUCTION: Abbott Alinity hq is a next-generation automated hematology analyzer providing complete blood count (CBC) with 6-part white blood cells (WBC) differential counts. The purpose of this study was to evaluate the performance of the analyzer to verify the diagnostic and clinical utility of the Abbott Alinity hq automated system.
METHODS: We evaluated specimen stability, precision, linearity, carry-over, and method comparison to assess the performance of Alinity hq. For comparison of the Alinity hq with Sysmex XN-9000, totally 314 samples from adult and pediatric patients including both normal and abnormal hematology profiles were analyzed in parallel. The Alinity hq was also compared with the manual differential counts for the same 314 samples.
RESULTS: At 4°C, the Alinity hq analyzer showed no significant changes in CBC and WBC differential count up to 48 hours. When stored at room temperature (18-25°C), all parameters except the mean platelet volume (MPV) were stable up to 36 hours. The Abbott Alinity hq analyzer demonstrated excellent reproducibility and between-batch precision for all CBC and WBC differential parameters. WBC, red blood cells (RBC), hemoglobin (HGB), and platelets showed good linearity and acceptable carry-over. Comparison with a Sysmex XN-9000 analyzer and manual 400-cell differential showed excellent correlation for CBC and WBC differential count parameters (correlation coefficient = 0.815-0.999) except for mean corpuscular hemoglobin concentration (MCHC) and basophils.
CONCLUSION: We performed initial validation studies and confirmed performance specifications on specimen stability, precision, linearity, carry-over, and method comparison. The Abbott Alinity hq analyzer showed good analytical performance for all standard CBC parameters.
© 2019 John Wiley & Sons Ltd.

Entities:  

Keywords:  Abbott Alinity hq; complete blood count; differential count; hematology analyzer; performance

Mesh:

Substances:

Year:  2019        PMID: 31271528     DOI: 10.1111/ijlh.13082

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Performance Evaluation of the Abbott Alinity hs Blood Slide Maker/Stainer.

Authors:  Min-Sun Kwak; In-Hwa Jeong; Sung-Sook Cho; Kwang-Sook Woo; Jin-Yeong Han
Journal:  Ann Lab Med       Date:  2022-07-01       Impact factor: 4.941

2.  Effects of specimen haemolysis on complete blood count results by Abbott Alinity hq System.

Authors:  Müjgan Ercan; Emiş Deniz Akbulut; Nihayet Bayraktar; Şerif Ercan
Journal:  Biochem Med (Zagreb)       Date:  2021-10-15       Impact factor: 2.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.