Eberson Damião Dos Santos de Bitencourt1, Carlos Franco Voegeli2, Gabriela Dos Santos Onzi3, Sara Cardoso Boscato3, Carine Ghem4, Terezinha Munhoz5.
Abstract
BACKGROUND: The speed and quality of information have become essential items in the release of laboratory reports. The Sysmex(®)SP1000-I device has been developed to prepare and stain smear slides. However, for a device to be cleared for use in the laboratory routine it must pass through a validation process.Entities:
Keywords: Automation, laboratory; Blood cell count/methods; Diagnostic equipment; Flow cytometry; Validation studies
Year: 2013 PMID: 24478606 PMCID: PMC3905822 DOI: 10.5581/1516-8484.20130121
Source DB: PubMed Journal: Rev Bras Hematol Hemoter ISSN: 1516-8484
Settings for blood smear performed by the Sysmex® SP-1000i equipment calculated from the hematocrit of the patient
|
|
|
| |
| Level 1 | < 20 | 31 | 180 |
| Level 2 | 20-25 | 30 | 175 |
| Level 3 | 25-30 | 29 | 165 |
| Level 4 | 30-35 | 29 | 160 |
| Level 5 | 35-40 | 29 | 155 |
| Level 6 | 40-45 | 29 | 150 |
| Level 7 | 45-50 | 29 | 120 |
| Level 8 | > 50 | 29 | 100 |
Evaluation of the quality of slides produced by the Sysmex® CellaVision® DM96 for each hematocrit level
|
|
|
|
|
| |
| Level 1 | 10 | 16.68 ± 1.87 | 9.9 ± 1.3 | 99.85 ± 0.24 | 15.24 ± 5.53 |
| Level 2 | 10 | 22.96 ± 1.13 | 10.1 ± 2.1 | 100 | 15.72 ± 11.42 |
| Level 3 | 10 | 26.96 ± 1.39 | 9.4 ± 1.5 | 99.85 ± 0.24 | 6 ± 3.68 |
| Level 4 | 10 | 32.49 ± 1.38 | 9.6 ± 1.8 | 99.81 ± 0.32 | 13.25 ± 11.21 |
| Level 5 | 10 | 37.2 ± 1.47 | 8.7 ± 1.5 | 99.9 ± 0.21 | 8.35 ± 4.81 |
| Level 6 | 10 | 41.16 ± 0.78 | 9.6 ± 1.1 | 100 | 10.67 ± 6.15 |
| Level 7 | 10 | 46.44 ± 1.12 | 9.1 ± 1.4 | 100 | 15.64 ± 13.62 |
| Level 8 | 10 | 52.27 ± 1.95 | 9.5 ± 1.5 | 100 | 14.2 ± 7.75 |
Number of slides and leukocytes improperly classified by the CellaVision® DM96 using slides prepared by the Sysmex® SP-1000i
|
|
|
|
| ||
|
|
|
| |||
| Level 1 | 10 | 0 | 0 | 0 | 0 |
| Level 2 | 10 | 1 | 1 | 0 | 0 |
| Level 3 | 10 | 1 | 1 | 0 | 1 |
| Level 4 | 10 | 2 | 2 | 1 | 0 |
| Level 5 | 10 | 1 | 1 | 0 | 0 |
| Level 6 | 10 | 1 | 1 | 0 | 0 |
| Level 7 | 10 | 0 | 0 | 0 | 0 |
| Level 8 | 10 | 2 | 2 | 1 | 1 |
| Total | 80 | 8 (10.0%) | 8 (10.0%) | 2 (2.5%) | 2 (2.5%) |
Correlation for the pre-classification of cells performed by the CellaVison® DM96 and the classification by the professional.
|
|
|
|
| Band neutrophils | 0.802 | < 0.001 |
| Segmented neutrophils | 0.963 | < 0.001 |
| Eosinophils | 0.958 | < 0.001 |
| Basophils | 0.568 | < 0.001 |
| Lymphocytes | 0.985 | < 0.001 |
| Monocytes | 0.791 | < 0.001 |
| Myelocytes | 0.465 | < 0.001 |
| Metamyelocytes | 0.456 | < 0.001 |
| Plasmocytes | 0.373 | < 0.001 |
| Atypical Lymphocytes | 0.866 | < 0.001 |
| Erythroblasts | 0.708 | < 0.001 |
Figure 1Linear regression and scatter plots for band neutrophils, segmented neutrophils, eosinophils, monocytes, lymphocytes, atypical lymphocytes, immature granulocytes and erythroblasts