Abstract
BACKGROUND: A novel robotic-assisted bipolar radiofrequency (RF) multifunctional vessel seal-and-transection instrument (SynchroSeal, Intuitive Surgical, Inc., Sunnyvale, CA) has been developed. The objective of the current paper is to describe the design of SynchroSeal based on bench studies, assess the safety of SynchroSeal in ex vivo and in vivo porcine studies, and provide early clinical context.Entities:
Keywords: Radiofrequency energy; Robotic assistance; Safety; Transection; Ultrasonic energy; Vessel sealing
Mesh:
Year: 2022 PMID: 35109833 PMCID: PMC8811999 DOI: 10.1186/s12893-022-01483-5
Source DB: PubMed Journal: BMC Surg ISSN: 1471-2482 Impact factor: 2.102
Fig. 1Cross-sectional schematic of tissue clamped in the SynchroSeal jaws
Bench study evaluations of SynchroSeal and Harmonic Ace+7
| Variable | SynchroSeal | Harmonic Ace+7 | Mean difference | P-value |
|---|---|---|---|---|
| Mean slip force, N ± SD | 8.4 ± 1.0 | 3.1 ± 0.4 | 5.3 | 0.0002 |
| Mean grip pressure, kg/cm2 ± SD | 3.0 ± 0.2 | 4.2 ± 0.5 | − 1.2 | 0.0002 |
| Mean seal time, s ± SD | 1.5 ± 0.4 | 11.6 ± 2.5 | − 9.9 s | < 0.0001 |
| Mean peak jaw temperature, °C ± SD | 109.7 ± 7.2 | 247.4 ± 8.6 | − 137.7 | 0.0051 |
| Mean time to cool to 40 °C, s ± SD | 53.6 ± 2.1 | 68.0 ± 3.5 | − 14.4 | 0.0051 |
SD standard deviation of the mean
Fig. 2Thermal images of (left) SynchroSeal and (right) Harmonic Ace+7 jaws during peak activation
Fig. 3Graphical representation of SynchroSeal and Harmonic Ace+7 jaw temperatures throughout activation and cool down
Fig. 4Graphical representation of mean burst pressures under Sync and Seal modes
Fig. 5Histology images of an in vivo porcine artery sealed and transected with a single pedal press in a Sync mode and b Seal mode
Fig. 6Vessel sealing and transection using SynchroSeal’s Sync mode for A ovarian bundle, B lineic vein, and C short gastric bundle during sealing (left images) and after sealing (right images). Yellow circles show the location of the seals
Single center, single user bariatrics analysis comparing Vessel Sealer Extend (Erbe VIO da Vinci generator) and SynchroSeal (E-100 da Vinci generator)
| Procedure and variables | Vessel Sealer Extend | SynchroSeal | p-value |
|---|---|---|---|
| Sleeve gastrectomy | |||
| Procedures, n | 49 | 53 | |
| Mean activations per procedure, n ± SD | 57 ± 25 a | 79 ± 26 | < 0.001 |
| Mean accumulated activation time per procedure, s ± SD | 249 ± 114 a | 119 ± 38 | < 0.001 |
| Mean time tool used per procedure, s ± SD | 1030 ± 560 | 1229 ± 445 | 0.053 |
| Mean tool usage as a proportion of time for all tools used per procedure, % ± SD | 35 ± 13 | 43 ± 17 | 0.007 |
| Gastric bypass (Roux-en-Y) | |||
| Procedures, n | 33 | 52 | |
| Mean activations per procedure, n ± SD | 40 ± 20 | 74 ± 44 | < 0.001 |
| Mean accumulated activation time per procedure, s ± SD | 179 ± 96 | 108 ± 65 | < 0.001 |
| Mean time tool used per procedure, s ± SD | 934 ± 769 | 1403 ± 809 | 0.010 |
| Mean tool usage as a proportion of time for all tools used per procedure, % ± SD | 20 ± 13 | 34 ± 18 | < 0.001 |
SD standard deviation of the mean
a Data missing for 4 cases