| Literature DB >> 30254903 |
Amruta Jaiswal1, Kuan-Gen Huang1.
Abstract
The introduction of newer vessel sealing systems has revolutionized techniques of hemostasis during laparoscopic surgery. These devices allow for rapid sequential tissue and vessel sealing, coagulation, and transection. Despite of widespread use of newer advanced bipolar and ultrasonic devices, monopolar and conventional bipolar electro-surgery still carry weightage due to wider range of tissue effect, dissection capabilities, cost effectiveness, and ease of availability. Here in we discussed different types of commonly available energy sources in terms of mechanism, efficacy and safety as thorough knowledge is utmost important for surgeon to choose appropriate instrument for surgical procedure.Entities:
Keywords: Electrosurgery; Energy devices; Gynecology; Laparoscopy
Year: 2017 PMID: 30254903 PMCID: PMC6135185 DOI: 10.1016/j.gmit.2017.08.002
Source DB: PubMed Journal: Gynecol Minim Invasive Ther ISSN: 2213-3070
Comparative studies of different electrosurgical device.
| Author | Type of study | Device | Sample size (N) | Procedure | Operative time (Min.) (Mean) | Blood loss (mL) | Postoperative pain score | Complication | Hospital stay (days) | Inference |
|---|---|---|---|---|---|---|---|---|---|---|
| Anna fagoti 2014 et al. | Randomized, controlled trial | TB vs. standard electrosurgery (SES) | N = 71 (excluded 21 due to intraoperative criteria). | Laparoscopic radical hysterectomy with bilateral pelvic lymphadenectomy | TB-85 | TB-50 | At 24 h | TB-0 | TB-3 | TB associated with short operative time and less postoperative pain |
| Hakan ayatan et al 2014. | Randomized prospective study | LS vs. Enseal vs. PK | N = 45 | Total laparoscopic hysterectomy | LS-52.4 | LS-138 | – | – | LS-1.1 | No significant difference except more blood loss in Enseal group |
| Ralf Rothmund et al 2013 | Prospective, randomized, controlled trial | Enseal vs. standard bipolar | N = 160, Enseal-80 bipolar – 80 | Laparoscopic Supracervical hysterectomy | Enseal-78.18 Bipolar – 86.3 | Enseal-50 mL ( | No significant difference | No significant difference | Enseal-2.01 Bipolar – 2.17 | EnSeal device is at least as reliable as the conventional electrocoagulation technique in laparoscopic supracervical hysterectomy (LASH). |
| Total resection time was shorter in the experimental group, and the other investigated clinical parameters were not inferior in the experimental group compared with the control group | ||||||||||
| Janssen et al. 2011 | Randomized controlled trial | LS vs. CB | N = 140 | Laparoscopic hysterectomy | LS-148.1 | LS-234.1 mL | – | – | LS-2.9 | No significant differences in operating time and blood loss |
| Hsuan su et al.2011 | Retrospective study | PK vs. CES | N-194 | Laparoscopic myomectomy | PK-117.8 | PK-190.4 | – | – | PK-2.7 | PK has advantage of less blood loss |
| Demirturk et al (2007) | Retrospective study | HS vs. LS | N = 40 | Total laparoscopic hysterectomy with salpingo-oophorectomy | HS-90.95 | HS-152.63 | – | – | HS-3.42 | LS has advantage of less operative time and less blood loss compared to HS |
| Lee et al. 2007 | Retrospective case–control study | PK vs. CB | N = 76 | Laparoscopic radical hysterectomy with pelvic lymphadenectomy | PK-172 | PK-397 mL | – | Less for PK | PK - 6.9 | PK has advantage of less blood loss, shorter operative time and less postoperative complications |
| Wang et al. 2005 | Prospective, non randomized trial | PK vs. CB | N = 62 | LAVH | PK-87.6 | PK-196.8 | – | – | PK-3.2 | Operation time, blood loss, transfusion rate, length of hospital stay: no significant difference |
Conventional bipolar-CB, Conventional Electrosurgery-CES, Harmonic scalpel-HS, Ligasure-LS, Plasma kinetic gyrus-PK, standard electrosurgery-SES, Thunderbeat-TB.
Different types of available energy sources and tissue effect produce by them.2343
| Type | Tissue effect | |
|---|---|---|
| Monopolar | Vaporization, fulguration, desiccation, coaptation | |
| Conventional bipolar | Desiccation, coaptation | |
| Advanced bipolar | Ligasure, pk gyrus, ENSEAL | Desiccation, coaptation, tissue transection |
| Ultrasonic technology | Ultracision harmonic scalpel, Harmonic ACE, Harmonic focus, SonoSurg, AutoSonix | Desiccation, coaptation, mechanical tissue transection |
| Hybrid device | Thunderbeat | |
| Laser energy | Nd: YAG laser, Argon laser, CO2 laser | |
| Argon beam coagulator | System 7550TM ABC, Cardioblate | |
| Radiofrequency (RF) energy | RF 3000, starburst, cardioblate |