Lauren Gallagher1, Mary Joyce1, Barry Murphy1, Marc Mac Giolla Eain1, Ronan MacLoughlin1,2,3.
Abstract
There are variations in the values reported for aerosol drug delivery across in vitro experiments throughout the published literature, and often with the same devices or similar experimental setups. Factors contributing to this variability include, but are not limited to device type, equipment settings, drug type and quantification methods. This study assessed the impact of head model choice on aerosol drug delivery using six different adults and three different paediatric head models in combination with a facemask, mouthpiece, and high-flow nasal cannula. Under controlled test conditions, the quantity of drug collected varied depending on the choice of head model. Head models vary depending on a combination of structural design differences, facial features (size and structure), internal volume measurements and airway geometries and these variations result in the differences in aerosol delivery. Of the widely available head models used in this study, only three were seen to closely predict in vivo aerosol delivery performance in adults compared with published scintigraphy data. Further, this testing identified the limited utility of some head models under certain test conditions, for example, the range reported across head models was aerosol drug delivery of 2.62 ± 2.86% to 37.79 ± 1.55% when used with a facemask. For the first time, this study highlights the impact of head model choice on reported aerosol drug delivery within a laboratory setting and contributes to explaining the differences in values reported within the literature.Entities:
Keywords: adult; aerosol drug delivery; facemask; head model; high flow nasal cannula; inhalation; mouthpiece; paediatric; vibrating mesh nebuliser
Year: 2021 PMID: 35056920 PMCID: PMC8777612 DOI: 10.3390/pharmaceutics14010024
Source DB: PubMed Journal: Pharmaceutics ISSN: 1999-4923 Impact factor: 6.321
Figure 1(A) TruCorp AirSim Advance X (AA91100X, TruCorp®, Craigavon, Ireland); (B) LUCY (Life/form® Adult Airway Management Trainer, Head Only, SKU: LF03603, Nasco Healthcare, Saugerties, NY, USA); (C) 3D Printed anatomically correct head model generated from an individual CT scan; (D) ANNE (Resusci Anne® QCPR, Laerdal medical, Stavanger, Norway); (E) FRANK (Model HSM-A 1550, Michigan Instruments, Kentwood, MI, USA); (F) ADAM (Adult #2000, Simulaids, INC, Saugerties, NY, USA); (G) TruCorp AirSim Child X (AC 10006X, TruCorp®, Craigavon, Ireland); (H) 3D Printed anatomically correct oropharyngeal (Male 5 years) generated from an individual CT scan; (I) 3D printed anatomically correct nasopharyngeal (1688) head model generated from an individual CT scan.
Head Model Characteristics.
| Head Model | Design | Internal | Internal | Dorsum Nasi (L) 1 | Nasal Orifices (w) 1 | Oral Opening (w) 1 | |
|---|---|---|---|---|---|---|---|
| Adult | LUCY | Airway | Nasal and oral anatomical landmarks | 212 cm3 | 3.5 cm | 2.2 cm | 4.5 cm |
| TruCorp AirSim Advance X | Airway | Nasal and oral anatomical landmarks | 318 cm3 | 4.5 cm | 3.5 cm | 5.0 cm | |
| ANNE | CPR Trainer Model | Simple single airway | 52 cm3 | 4.3 cm | 2.0 cm | 5.0 cm | |
| ADAM | CPR Adult | Simple single airway | 77 cm3 | 4.3 cm | N/A | 4.5 cm | |
| FRANK | Head Simulation Model | Simple single airway | 27 cm3 | 4.3 cm | N/A | 4.5 cm | |
| 3D printed Adult Head Model | 3D printed | Anatomically correct | 84 cm3 | 3.5 cm | 2.5 cm | N/A | |
| Paediatric | TruCorp AirSim Child X | Airway | Anatomically correct | 76 cm3 | 3.4 cm | 2.2 cm | 4.0 cm |
| 3D printed | 3D printed | Single anatomically correct oropharyngeal airway | 25 cm3 | N/A | N/A | 3.5 cm | |
| 3D printed | 3D printed | Single anatomically correct nasopharyngeal airway | 21 cm3 | 2.0 cm | 2.3 cm | N/A |
1 Nasal surface measurements represent the length of the Dorsum Nasi (L) and width of the lateral openings (W) of the nasal and oral orifices.
Figure 2Experimental Set-up: Head model set-up in combination with the ASL 5000 breathing simulator. Aerosol administration provided with the Aerogen Ultra with a facemask or mouthpiece and the Aerogen Solo/Airvo 2 with Optiflow Nasal cannula at supplemental flow rates of 0, 2 and 6 LPM and high flow gas rates of 10 and 50 LPM respectively.
Mean ± SD % tracheal dose for Adult head models in combination with the Aerogen Ultra with a facemask or mouthpiece and the Aerogen Solo/Airvo 2 with Optiflow Nasal cannula at supplemental flow rates of 0, 2 and 6 LPM and high flow gas rates of 10 and 50 LPM respectively.
| Interface | Supplemental Gas Flow Rate | TruCorp AirSim | LUCY | ANNE | ADAM | FRANK | 3D Printed Adult Head Model | |
|---|---|---|---|---|---|---|---|---|
| Facemask | 0 | 6.35 ± 0.87 | 22.79 ± 1.84 | 2.62 ± 2.86 | 13.02 ± 1.60 | 37.79 ± 1.55 | NT 1 | 0.000 |
| 2 | 17.48 ± 1.82 | 29.30 ± 0.95 | 24.89 ± 3.65 | 16.61 ± 2.04 | 35.80 ± 1.26 | NT 1 | 0.000 | |
| 6 | 16.75 ± 2.86 | 23.61 ± 1.73 | 24.18 ± 3.07 | 17.55 ± 1.82 | 31.25 ± 2.54 | NT 1 | 0.000 | |
| Mouthpiece | 0 | 32.40 ± 4.42 | 33.24 ± 1.64 | 1.93 ± 2.79 | NT 1 | 28.20 ± 1.33 | 30.53 ± 1.32 | 0.000 |
| 2 | 36.56 ± 4.13 | 32.90 ± 2.95 | 27.94 ± 2.49 | NT 1 | 30.10 ± 1.37 | 30.54 ± 3.37 | 0.003 | |
| 6 | 22.29 ± 1.42 | 21.92 ± 1.92 | 28.11 ± 2.13 | NT 1 | 24.97 ± 1.19 | 24.96 ± 3.74 | 0.002 | |
| Nasal | 10 | 11.85 ± 1.07 | 15.97 ± 1.84 | NT 1 | NT 1 | NT 1 | 15.66 ± 0.87 | 0.001 |
| 50 | 4.13 ± 1.17 | 2.60 ± 0.21 | NT 1 | NT 1 | NT 1 | 2.32 ± 0.23 | 0.003 |
1 NT = not tested.
Mean ± SD % tracheal dose for Paediatric Head Models in combination with the Aerogen Ultra with a facemask or mouthpiece and the Aerogen Solo/Airvo 2 with Optiflow Nasal cannula at supplemental flow rates of 0, 2 and 6 LPM and high flow gas rates of 3, 5 and 10 LPM respectively.
| Interface | Supplemental Gas Flow Rate (LPM) | TruCorp AirSim Child X | 3D Printed | 3D Printed | |
|---|---|---|---|---|---|
| Facemask | 0 | 5.24 ± 1.53 | 31.86 ± 3.67 | NT 1 | 0.000 |
| 2 | 21.51 ± 4.14 | 24.05 ± 5.88 | NT 1 | 0.453 | |
| 6 | 25.50 ± 4.94 | 26.94 ± 2.73 | NT 1 | 0.583 | |
| Mouthpiece | 0 | 20.42 ± 3.58 | 3.88 ± 0.70 | NT 1 | 0.000 |
| 2 | 23.43 ± 3.99 | 26.48 ± 7.76 | NT 1 | 0.457 | |
| 6 | 21.18 ± 0.56 | 24.03 ± 3.62 | NT 1 | 0.120 | |
| Nasal | 3 | 9.08 ± 2.67 | NT 1 | 6.49 ± 2.46 | 0.149 |
| 5 | 7.04 ± 1.46 | NT 1 | 5.55 ± 0.98 | 0.094 | |
| 10 | 4.00 ± 0.63 | NT 1 | 6.28 ± 0.22 | 0.000 |
1 NT = not tested.