Albuterol Delivery Efficiency in a Pediatric Model of Noninvasive Ventilation With Double-Limb Circuit.

Jeanne Velasco1, Ariel Berlinski2.   

Abstract

BACKGROUND: Noninvasive ventilation (NIV) is used to treat respiratory failure in patients with concomitant need for aerosol delivery. Limited pediatric data are available on aerosol delivery efficiency, and none at all regarding aerosol delivery efficiency with a double-limb circuit. We compared the effect of position in the double-limb ventilator circuit, types of nebulizer, and ventilator settings on aerosol delivery efficiency in a pediatric model of NIV. We hypothesized that placing a vibrating mesh nebulizer at the ventilator and using the highest inspiratory pressures would increase aerosol delivery efficiency.
METHODS: A breathing simulator was connected in series to a low dead-space filter holder (lung dose) and to an anatomically correct head/airway model of a 5-year-old child. A non-vented mask connected the model to a ventilator operated on noninvasive bi-level mode and assembled with a double-limb, heated-wired adult circuit. Inspiratory pressures of either 15 or 20 cm H2O and an expiratory pressure of 5 cm H2O were used. Two different vibrating mesh nebulizers and 2 different jet nebulizers loaded with albuterol solution were studied. Albuterol was measured with spectrophotometry. Aerosol delivery efficiency was calculated as lung dose expressed as a percentage of the nominal dose.
RESULTS: A vibrating mesh nebulizer before the mask or Y-piece provided the highest delivery efficiency and outperformed a vibrating mesh nebulizer integrated into the mask. Vibrating mesh nebulizers delivered more drug than jet nebulizers, regardless of their position in the circuit. Increasing the inspiratory pressure only improved aerosol delivery efficiency with a jet nebulizer placed at the ventilator.
CONCLUSIONS: In a pediatric model of NIV, the effect of nebulizer position on aerosol delivery efficiency depends on the type of device and its placement in the ventilator circuit. A vibrating mesh nebulizer placed at the mask or before the Y-piece of the double-limb circuit provided the highest aerosol drug delivery during NIV. Data generated with invasive ventilation models should not be generalized to NIV models.
Copyright © 2018 by Daedalus Enterprises.

Entities:  

Keywords:  aerosol delivery; airway model; albuterol; lung dose; nebulizer; noninvasive ventilation; pediatrics

Mesh:

Substances:

Year:  2017        PMID: 29114011     DOI: 10.4187/respcare.05833

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  3 in total

1.  The utilization of aerosol therapy in mechanical ventilation patients: a prospective multicenter observational cohort study and a review of the current evidence.

Authors:  Shan Lyu; Jie Li; Limin Yang; Xiaoliang Du; Xiaoyi Liu; Libo Chuan; Guoqiang Jing; Zhenyan Wang; Weiwei Shu; Chunjuan Ye; Qionglan Dong; Jun Duan; James B Fink; Zhancheng Gao; Zongan Liang
Journal:  Ann Transl Med       Date:  2020-09

2.  Effect of pressures and type of ventilation on aerosol delivery to chronic obstructive pulmonary disease patients.

Authors:  Marina E Boules; Nabila Ibrahim Laz; Ahmed A Elberry; Raghda R S Hussein; Mohamed E A Abdelrahim
Journal:  Beni Suef Univ J Basic Appl Sci       Date:  2022-04-15

3.  The Impact of Head Model Choice on the In Vitro Evaluation of Aerosol Drug Delivery.

Authors:  Lauren Gallagher; Mary Joyce; Barry Murphy; Marc Mac Giolla Eain; Ronan MacLoughlin
Journal:  Pharmaceutics       Date:  2021-12-23       Impact factor: 6.321

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.