Performance Comparisons of Jet and Mesh Nebulizers Using Different Interfaces in Simulated Spontaneously Breathing Adults and Children.

Arzu Ari1, Armele Dornelas de Andrade2, Meryl Sheard1, Bshayer AlHamad1, James B Fink1.   

Abstract

BACKGROUND: Different types of nebulizers and interfaces are used for the treatment of adults and children with pulmonary diseases. The purpose of this study was to determine the efficiency of a mesh nebulizer (MN) with a proprietary adapter and a jet nebulizer (JN) under different configurations in adult and pediatric models of spontaneous breathing. We hypothesize that delivery efficiency of JN and MN will differ depending on the interface used during aerosol therapy in simulated spontaneously breathing adult and pediatric models. While we expect that aerosol delivery with JN will be less efficient than MN, we also hypothesize that lung deposition obtained with the adult lung model will be more than that with the pediatric lung model in all conditions tested in this study.
METHODS: A lung model using a teaching manikin connected to a sinusoidal pump via a collecting filter at the level of the bronchi simulating a spontaneously breathing adult (Vt 500 mL, RR 15 bpm, I:E ratio 1:2) or pediatric patient (Vt 150 mL, RR 25 bpm, I:E ratio 1:2). Albuterol sulfate (2.5 mg/3 mL) was aerosolized with JN (Mistymax 10, Airlife) or MN (Aerogen Solo(®), Aerogen) with the Adapter (Aerogen Solo(®) Adapter, Aerogen Ltd, Galway, Ireland) using mouthpiece, aerosol mask, and valved-mask in adults and the dragon mask, aerosol mask, and valved-mask in pediatrics (n=3). The Adapter, specifically designed for MN, was attached to all the interfaces used in this study with supplemental oxygen of 2 lpm, and in addition, the MP was tested with no additional flow in the adult model. The JN was driven with 10 lpm based on the manufacturer's label. Drug was eluted from the filter and analyzed via spectrophotometry. Descriptive statistics, dependent t-test and one-way analysis of variance were used for data analysis. Significant level was set at 0.05.
RESULTS: In adults, delivery efficiency of JN with the valved mask was significantly greater than that with the aerosol mask (p=0.01). Aerosol delivery of JN with the mouthpiece was not statistically significant from the valved mask (p=0.123) and the aerosol mask (p=0.193). Drug delivery with MN with mouthpiece (15.42±1.4%) and valved-mask (15.15±1.1%) was greater than the open aerosol mask (7.54±0.39%; p=0.0001) in the adult lung model. With no flow mouthpiece delivery increased>2 fold (34.9±3.1%; p=.0001) compared to use of 2 lpm of flow. Using the JN with the pediatric model deposition with valved-mask (5.3±0.8%), dragon mask (4.7±0.9%), and aerosol mask (4.1±0.3%) were similar (p>0.05); while drug delivery with MN via valved-mask (11.1±0.7%) was greater than the dragon mask (6.44±0.3%; p=0.002) and aerosol mask (4.6±0.4%; p=0.002), and the dragon mask was more efficient than the open aerosol mask (p=0.009)
CONCLUSION: The type of nebulizer and interface used for aerosol therapy affects delivery efficiency in these simulated spontaneously breathing adult and pediatric models. Drug delivery was greatest with the valved-mouthpiece and mask with JN and MN, while the standard aerosol mask was least efficient in these simulated spontaneously breathing adult and pediatric lung models. Delivery efficiency of JN was less than MN in all conditions tested in this study except in the aerosol mask. Lung deposition obtained with the adult lung model was more than that with the pediatric lung model.

Entities:  

Keywords:  adults; aerosols; drug delivery; face mask; inhalation therapy; mouthpiece; nebulizers; pediatrics

Mesh:

Substances:

Year:  2014        PMID: 25493535     DOI: 10.1089/jamp.2014.1149

Source DB:  PubMed          Journal:  J Aerosol Med Pulm Drug Deliv        ISSN: 1941-2711            Impact factor:   2.849


  20 in total

Review 1.  A path to successful patient outcomes through aerosol drug delivery to children: a narrative review.

Authors:  Arzu Ari
Journal:  Ann Transl Med       Date:  2021-04

2.  Electronic Cigarette Vapor with Nicotine Causes Airway Mucociliary Dysfunction Preferentially via TRPA1 Receptors.

Authors:  Samuel Chung; Nathalie Baumlin; John S Dennis; Robert Moore; Sebastian F Salathe; Phillip L Whitney; Juan Sabater; William M Abraham; Michael D Kim; Matthias Salathe
Journal:  Am J Respir Crit Care Med       Date:  2019-11-01       Impact factor: 21.405

Review 3.  Devices for Improved Delivery of Nebulized Pharmaceutical Aerosols to the Lungs.

Authors:  Worth Longest; Benjamin Spence; Michael Hindle
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2019-07-09       Impact factor: 2.849

4.  SPECT-CT Comparison of Lung Deposition using a System combining a Vibrating-mesh Nebulizer with a Valved Holding Chamber and a Conventional Jet Nebulizer: a Randomized Cross-over Study.

Authors:  Jonathan Dugernier; Michel Hesse; Rita Vanbever; Virginie Depoortere; Jean Roeseler; Jean-Bernard Michotte; Pierre-François Laterre; François Jamar; Gregory Reychler
Journal:  Pharm Res       Date:  2016-11-07       Impact factor: 4.200

5.  Numerical Comparison of Nasal Aerosol Administration Systems for Efficient Nose-to-Brain Drug Delivery.

Authors:  Jingliang Dong; Yidan Shang; Kiao Inthavong; Hak-Kim Chan; Jiyuan Tu
Journal:  Pharm Res       Date:  2017-12-29       Impact factor: 4.200

Review 6.  Drug delivery interfaces: A way to optimize inhalation therapy in spontaneously breathing children.

Authors:  Arzu Ari
Journal:  World J Clin Pediatr       Date:  2016-08-08

Review 7.  A Narrative Review of a Pulmonary Aerosolized Formulation or a Nasal Drop Using Sera Containing Neutralizing Antibodies Collected from COVID-19-Recovered Patients as a Probable Therapy for COVID-19.

Authors:  Nishat Fatima; Vichitra Kaushik; Amjad Ayoub
Journal:  Iran J Med Sci       Date:  2021-05

8.  Effects of ventilator settings, nebulizer and exhalation port position on albuterol delivery during non-invasive ventilation: an in-vitro study.

Authors:  Yuda Sutherasan; Lorenzo Ball; Pasquale Raimondo; Valentina Caratto; Elisa Sanguineti; Federico Costantino; Maurizio Ferretti; Robert M Kacmarek; Paolo Pelosi
Journal:  BMC Pulm Med       Date:  2017-01-10       Impact factor: 3.317

9.  Comparison of the Clinical Efficacy of Salbutamol with Jet and Mesh Nebulizers in Asthmatic Children.

Authors:  Norihide Murayama; Kikuno Murayama
Journal:  Pulm Med       Date:  2018-03-13

10.  In vitro delivery efficiencies of nebulizers for different breathing patterns.

Authors:  Hyun Mok Park; Kyung Hwa Chang; Sang-Hyub Moon; Bong Joo Park; Sun Kook Yoo; Ki Chang Nam
Journal:  Biomed Eng Online       Date:  2021-06-10       Impact factor: 2.819

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.