| Literature DB >> 32026245 |
Gavin Bennett1, Mary Joyce2, Louise Sweeney2, Ronan MacLoughlin2.
Abstract
INTRODUCTION: The use of concurrent aerosol delivery during high-flow nasal therapy (HFNT) may be exploited to facilitate the delivery of a variety of prescribed medications for inhalation. Until now, a systematic approach to determine the conditions required to yield an optimal emitted dose has not been reported. The aim of this study was to establish the effects of gas flow rate, input droplet size, and nebulizer position on the amount of aerosol exiting the nasal cannula during HFNT and thus becoming available for inhalation.Entities:
Keywords: Aerosol; Design of experiments; Droplet size; Emitted dose; Flow rate; High-flow nasal therapy; Nebulizer
Year: 2018 PMID: 32026245 PMCID: PMC6967237 DOI: 10.1007/s41030-018-0054-x
Source DB: PubMed Journal: Pulm Ther ISSN: 2364-1754
Fig. 1Illustration of test setup employed, indicating the two nebulizer placement positions within the HFNT circuit
Input droplet size and output rate data with 0.9% saline (N = 3), direct from the nebulizer
| Input droplet size (µm VMD) | Aerosol output rate (mL/min) |
|---|---|
| 3.22 ± 0.02 | 0.27 ± 0.02 |
| 4.05 ± 0.02 | 0.34 ± 0.03 |
| 4.88 ± 0.03 | 0.51 ± 0.02 |
Study design obtained using a design of experiments approach
| Base design | 3 Factors, 8 runs |
|---|---|
| Replicates | 4 |
| Center points | 20 |
| Axial points | 8 |
| Total runs | 60 |
| Blocks | 5 |
Summary of testing completed to determine the effects of input droplet size, delivery gas flow rate, and nebulizer position on emitted dose using an adult cannula
| Input droplet size (µm VMD) | Delivery gas flow rate (LPM) | Nebulizer position | Sample number ( |
|---|---|---|---|
| 3.22 | 10 | Proximal | |
| Distal | |||
| 25 | Proximal | ||
| Distal | |||
| 40 | Proximal | ||
| Distal | |||
| 4.05 | 10 | Proximal | |
| Distal | |||
| 25 | Proximal | ||
| Distal | |||
| 40 | Proximal | ||
| Distal | |||
| 4.88 | 10 | Proximal | |
| Distal | |||
| 25 | Proximal | ||
| Distal | |||
| 40 | Proximal | ||
| Distal |
Summary of testing completed to determine the effects of input droplet size, delivery gas flow rate, and nebulizer position on emitted dose using a pediatric cannula
| Input droplet size (µm VMD) | Delivery gas flow rate (LPM) | Nebulizer position | Sample number ( |
|---|---|---|---|
| 3.22 | 5 | Proximal | |
| Distal | |||
| 20 | Proximal | ||
| Distal | |||
| 4.05 | 12.5 | Proximal | |
| Distal | |||
| 4.88 | 20 | Proximal | |
| Distal |
Summary of testing completed to determine the effects of input droplet size, delivery gas flow rate, and nebulizer position on emitted dose using an infant cannula
| Input droplet size (µm VMD) | Delivery gas flow rate (LPM) | Nebulizer position | Sample number ( |
|---|---|---|---|
| 3.22 | 2 | Proximal | |
| Distal | |||
| 8 | Proximal | ||
| Distal | |||
| 4.05 | 5 | Proximal | |
| Distal | |||
| 4.88 | 8 | Proximal | |
| Distal |
Mean ± standard deviation values of emitted dose (%) for testing completed according to the DOE using an adult nasal cannula (Comfort Flo P/N: 2411-01, Teleflex Medical)
| Input droplet size (µm VMD) | Flow rate (LPM) | Nebulizer position | Sample number ( | Emitted dose (%) |
|---|---|---|---|---|
| 3.22 | 10 | Proximal | 11.63 ± 0.39 | |
| Distal | 8.75 ± 0.69 | |||
| 25 | Proximal | n/a | ||
| Distal | 2.79 | |||
| 40 | Proximal | 2.68 ± 0.81 | ||
| Distal | 0.70 ± 0.34 | |||
| 4.05 | 10 | Proximal | 6.62 | |
| Distal | 4.04 | |||
| 25 | Proximal | 2.62 ± 0.77 | ||
| Distal | 2.61 ± 0.66 | |||
| 40 | Proximal | 0.77 | ||
| Distal | 1.25 | |||
| 4.88 | 10 | Proximal | 2.75 ± 0.52 | |
| Distal | 5.04 ± 0.52 | |||
| 25 | Proximal | 1.80 | ||
| Distal | 1.91 | |||
| 40 | Proximal | 1.11 ± 0.40 | ||
| Distal | 0.88 ± 0.19 |
Fig. 2Effects of flow rate, input droplet size, and nebulizer position on emitted dose. Blue bars in the graph show the factors that had significant effects on the emitted dose (at the 0.1 level of significance). Gray bars represent nonsignificant terms that were removed from the model. Flow rate had the greatest influence on emitted dose, followed by input droplet size and then nebulizer position
Top five alternative solutions to maximize emitted dose with an adult nasal cannula (Comfort Flo P/N: 2411-01, Teleflex Medical)
| Input droplet size (µm VMD) | Flow rate (LPM) | Nebulizer position | Predicted emitted dose (%) |
|---|---|---|---|
| 3.22 | 10 | Distal | 8.91 |
| 4.05 | 10 | Proximal | 7.07 |
| 4.05 | 10 | Distal | 6.64 |
| 3.22 | 25 | Proximal | 5.38 |
| 4.88 | 10 | Distal | 4.37 |
Fig. 3Effect of gas flow rate on emitted dose
Fig. 4Effect of input droplet size on emitted dose
Fig. 5Effect of nebulizer position on emitted dose
Effect of nasal cannula design (delivery gas flow rate and cannula type) on emitted dose
| Delivery gas flow (LPM) | Comfort Flo | Comfort Flo Plus |
|---|---|---|
| 10 | 6.62 | 55.65 ± 1.19 |
| 25 | 2.62 ± 0.77 | 49.89 ± 1.93 |
| 40 | 0.77 | 40.40 ± 0.98 |
Mean ± standard deviation values of emitted dose (%) for testing completed using a pediatric cannula (Comfort Flo P/N: 2411-02, Teleflex Medical)
| Input droplet size (µm VMD) | Delivery gas flow rate (LPM) | Nebulizer position | Sample number ( | Emitted dose (%) |
|---|---|---|---|---|
| 3.22 | 5 | Proximal | 18.73 ± 1.50 | |
| Distal | 8.46 ± 0.64 | |||
| 20 | Proximal | 2.35 ± 0.35 | ||
| Distal | 1.21 ± 0.36 | |||
| 4.05 | 12.5 | Proximal | 4.31 ± 0.79 | |
| Distal | 2.02 ± 0.19 | |||
| 4.88 | 20 | Proximal | 1.05 ± 0.26 | |
| Distal | 1.07 ± 0.07 |
Mean ± standard deviation values of emitted dose (%) for testing completed using an infant cannula (Comfort Flo P/N: 2411-03, Teleflex Medical)
| Input droplet size (µm VMD) | Delivery gas flow rate (LPM) | Nebulizer position | Sample number ( | Emitted dose (%) |
|---|---|---|---|---|
| 3.22 | 2 | Proximal | 18.03 ± 1.05 | |
| Distal | 6.81 ± 0.78 | |||
| 8 | Proximal | 4.55 ± 0.61 | ||
| Distal | 1.36 ± 0.31 | |||
| 4.05 | 5 | Proximal | 7.27 ± 1.12 | |
| Distal | 1.86 ± 0.12 | |||
| 4.88 | 8 | Proximal | 2.14 ± 0.38 | |
| Distal | 1.10 ± 0.15 |
Mean ± standard deviation values of MMAD and GSD for each input droplet size at the nasal cannula with two different gas flow rates
| Input droplet size | Adult cannula at 10 LPM | Adult cannula at 40 LPM | ||||
|---|---|---|---|---|---|---|
| VMD (µm) | MMAD (µm) | GSD (µm) | MMAD (µm) | GSD (µm) | MMAD (µm) | GSD (µm) |
| 3.22 | 3.23 ± 0.04 | 2.36 ± 0.02 | 1.82 ± 0.18 | 1.45 ± 0.06 | 1.72 ± 0.08 | 1.64 ± 0.40 |
| 4.05 | 3.38 ± 0.27 | 2.49 ± 0.06 | 1.90 ± 0.09 | 1.46 ± 0.06 | 1.72 ± 0.14 | 3.22 ± 1.76 |
| 4.88 | 4.62 ± 0.11 | 2.31 ± 0.07 | 1.68 ± 0.09 | 1.72 ± 0.19 | 1.31 ± 0.10 | 1.73 ± 0.43 |