| Literature DB >> 34262965 |
Arzu Ari1, James B Fink1.
Abstract
What is the delivered dose with jet and mesh nebulisers during spontaneous breathing (SB), noninvasive ventilation (NIV), and mechanical ventilation (MV) using an adult lung model with exhaled humidity (EH)? The delivery of salbutamol sulfate (2.5 mg per 3 mL) with jet (Mistymax10) and mesh nebulisers (Aerogen Solo) was compared during SB, NIV, and MV using breathing parameters (tidal volume 450 mL, respiratory rate 20 breaths per min, inspiratory:expiratory ratio 1:3) with three lung models simulating exhaled humidity. A manikin was attached to a sinusoidal pump via a filter at the bronchi to simulate an adult with SB. A ventilator (V60) was attached via a facemask to a manikin with a filter at the bronchi connected to a test lung to simulate an adult receiving NIV. A ventilator-dependent adult was simulated through a ventilator (Servo-i) operated with a heated humidifier (Fisher & Paykel) attached to an endotracheal tube (ETT) with a heated-wire circuit. The ETT was inserted into a filter (Respirgard II). A heated humidifier was placed between the filter and test lung to simulate exhaled humidity (35±2°C, 100% relative humidity). Nebulisers were placed at the Y-piece of the inspiratory limb during MV and positioned between the facemask and the leak-port during NIV. A mouthpiece was used during SB. The delivered dose was collected in an absolute filter that was attached to the bronchi of the mannequin during each aerosol treatment and measured with spectrophotometry. Drug delivery during MV was significantly greater than during NIV and SB with a mesh nebuliser (p=0.0001) but not with a jet nebuliser (p=0.384). Delivery efficiency of the mesh nebuliser was greater than the jet nebuliser during MV (p=0.0001), NIV (p=0.0001), and SB (p=0.0001). Aerosol deposition obtained with a mesh nebuliser was greater and differed between MV, NIV, and SB, while deposition was low with a jet nebuliser and similar between the modes of ventilation tested.Entities:
Year: 2021 PMID: 34262965 PMCID: PMC8273293 DOI: 10.1183/23120541.00027-2021
Source DB: PubMed Journal: ERJ Open Res ISSN: 2312-0541
FIGURE 1Schematic of the study design, including each nebuliser and modes of ventilation.
FIGURE 2Lung models used in this study. a) Mechanical ventilation lung model, b) noninvasive ventilation lung model, and c) spontaneous breathing lung model. ETT: endotracheal tube.
Mean±sd percentage of dose delivered with jet and mesh nebulisers during mechanical ventilation, noninvasive ventilation, and spontaneous breathing in adults
| 6.80±1.55 | 6.10±0.34 | 6.12±0.26 | 0.384 | |
| 23.16±0.67 | 18.36±1.06 | 17.29±1.69 | 0.0001 | |
| 0.0001 | 0.0001 | 0.0001 |