Sungje Bock1, Hyung-Mo Kim1, Jaehi Kim1, Jaehyun An1,2, Yun-Sik Choi3, Xuan-Hung Pham1, Ahla Jo1, Kyeong-Min Ham1, Hobeom Song2, Jung-Won Kim2, Eunil Hahm1, Won-Yeop Rho4, Sang Hun Lee5, Seung-Min Park6, Sangchul Lee7, Dae Hong Jeong3, Ho-Young Lee8, Bong-Hyun Jun1.
Abstract
Prostate cancer can be detected early by testing the presence of prostate-specific antigen (PSA) in the blood. Lateral flow immunoassay (LFIA) has been used because it is cost effective and easy to use and also has a rapid sample-to-answer process. Quantum dots (QDs) with very bright fluorescence have been previously used to improve the detection sensitivity of LFIAs. In the current study, a highly sensitive LFIA kit was devised using QD-embedded silica nanoparticles. In the present study, only a smartphone and a computer software program, ImageJ, were used, because the developed system had high sensitivity by using very bright nanoprobes. The limit of PSA detection of the developed LFIA system was 0.138 ng/mL. The area under the curve of this system was calculated as 0.852. The system did not show any false-negative result when 47 human serum samples were analyzed; it only detected PSA and did not detect alpha-fetoprotein and newborn calf serum in the samples. Additionally, fluorescence was maintained on the strip for 10 d after the test. With its high sensitivity and convenience, the devised LFIA kit can be used for the diagnosis of prostate cancer.Entities:
Keywords: lateral flow immunoassay; prostate cancer; prostate-specific antigen; quantum dot; quantum-dot-embedded silica nanoparticles
Year: 2021 PMID: 35009984 PMCID: PMC8746978 DOI: 10.3390/nano12010033
Source DB: PubMed Journal: Nanomaterials (Basel) ISSN: 2079-4991 Impact factor: 5.076
Figure 1(a) Schematic illustration of QD2-PSA Ab fabrication (this figure is not drawn to scale); (b) transmission electron microscopy (TEM) images of QD2. The inset depicts individual QD2; (c) UV–Vis absorbance of SiO2 NPs, QD2, and QD2-PSA Ab; (d) PL intensity of SiO2 NPs, QD2, and QD2-PSA Ab; (e) PL intensity comparison between QDs and QD2.
Figure 2(a) Schematic illustration of lateral flow immunoassay (LFIA) process and precise analysis; (b) fluorescence image showing only the red channel of the test strip with PSA developed under a 365 nm UV lamp; (c) T/C value and fitting curve of the fluorescence intensity generated for each concentration of PSA.
PSA concentration in clinical samples.
| Clinical Sample No. | PSA Concentration | Clinical Sample No. | PSA Concentration | Clinical Sample No. | PSA Concentration |
|---|---|---|---|---|---|
| 1 | 0.001 | 17 | 0.954 | 33 | 4.557 |
| 2 | 0.159 | 18 | 1.146 | 34 | 4.655 |
| 3 | 0.164 | 19 | 1.309 | 35 | 4.815 |
| 4 | 0.300 | 20 | 1.412 | 36 | 4.888 |
| 5 | 0.323 | 21 | 1.455 | 37 | 4.931 |
| 6 | 0.429 | 22 | 1.488 | 38 | 5.182 |
| 7 | 0.479 | 23 | 1.514 | 39 | 5.607 |
| 8 | 0.514 | 24 | 1.677 | 40 | 5.880 |
| 9 | 0.577 | 25 | 1.689 | 41 | 6.418 |
| 10 | 0.619 | 26 | 2.093 | 42 | 7.551 |
| 11 | 0.677 | 27 | 2.233 | 43 | 7.729 |
| 12 | 0.797 | 28 | 2.788 | 44 | 7.900 |
| 13 | 0.897 | 29 | 3.637 | 45 | 8.125 |
| 14 | 0.919 | 30 | 3.847 | 46 | 12.843 |
| 15 | 0.921 | 31 | 4.043 | 47 | 12.950 |
| 16 | 0.945 | 32 | 4.398 | - | - |
Figure 3(a) Fluorescence image showing only the red channel of the test strip after the development of human serum samples under a 365 nm UV lamp; (b) T/C value and concentration of PSA of each human serum sample; (c) ROC curve.
Figure 4(a) Fluorescence image showing only the red channel of the test strip after the development of PSA, alpha-fetoprotein (AFP), and newborn calf serum (NCS) under a 365 nm UV lamp for the selectivity test; (b) T/C value of PSA, AFP, and NCS in strip tests.
Figure 5(a) Fluorescence image showing only the red channel of the test strip (stored for 10 d) after the development of 4.557 ng/mL of human serum sample under a 365 nm UV lamp for the stability test; (b) normalized T/C value of the developed human serum sample.