Abstract
Complex intraarticular distal radius fractures (DRFs), commonly managed with volar locking plates, are challenging. Combined volar and dorsal plating is frequently applied for treatment, however, biomechanical investigations are scant. The aim of this biomechanical study was to investigate volar plating versus double plating in DRFs with different degrees of lunate facet comminution.Thirty artificial radii with simulated AO/OTA 23-C2.1 and C3.1 DRFs, including dorsal defect and lunate facet comminution, were assigned to 3 groups: Group 1 with two equally-sized lunate facet fragments; Group 2 with small dorsal and large volar fragment; Group 3 with three equally-sized fragments. The specimens underwent volar and double locked plating and non-destructive ramped loading in 0° neutral position, 40° flexion and 40° extension.In each tested position, stiffness: (1) did not significantly differ among groups with same fixation method (p ≥ 0.15); (2) increased significantly after supplemental dorsal plating in Group 2 and Group 3 (p ≤ 0.02).Interfragmentary displacements between styloid process and lunate facet in neutral position were below 0.5 mm, being not significantly different among groups and plating techniques (p ≥ 0.63).Following volar plating, angular displacement of the lunate facet to radius shaft was significantly lower in Group 1 versus both Group 2 and Group 3 (p < 0.01). It decreased significantly after supplemental dorsal plating in Group 2 and Group 3 (p < 0.01), but not in Group 1 (p ≥ 0.13), and did not differ significantly among the three groups after double plating (p ≥ 0.74).Comminution of the lunate facet within its dorsal third significantly affected the biomechanical outcomes related to complex intraarticular DRFs treated with volar and double locked plates.Double plating demonstrates superior stability versus volar plating only for lunate facet comminution within its dorsal third. In contrast, volar plating could achieve stability comparable with double plating when the dorsal third of the lunate facet is not separated by the fracture pattern. Both fixation methods indicated achievable absolute stability between the articular fragments.Entities:
Keywords: Biomechanical testing; Complex intraarticular distal radius fracture; Dorsoulnar fragment fixation; Double plating; Volar plate
Mesh:
Year: 2022 PMID: 34986819 PMCID: PMC8734044 DOI: 10.1186/s12891-021-04989-w
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Coronal (top) and axial (bottom) computed tomography scans of clinical cases with intraarticular distal radius fractures with different degree of lunate facet comminution comprising two fragments of a comparable size (left), a smaller dorsal and a larger volar fragment (middle), and three fragments of a comparable size (right)
Fig. 2Schematic illustration of the distal radius and ulna in axial view (left), together with the group assignment according to the intraarticular fracture patterns at the lunate facet (right) with two equally-sized fragments (Group 1, top right), with a small dorsal and a large volar fragment at a ratio of 1:2 (Group 2, middle right), and with three equally-sized fragments (Group 3, bottom right)
Fig. 3Exemplified specimens from Group 2 after volar plating using a 2.4 mm Variable Angle LCP Two-Column Volar Distal Radius Plate (DePuy Synthes, Zuchwil, Switzerland) (left), and double plating using a supplemental Variable Angle LCP Dorsal Distal Radius Plate (DePuy Synthes, Zuchwil, Switzerland) (right)
Fig. 4Test setup with a specimen mounted for biomechanical testing in extension, with vertical arrow denoting the loading direction, and indicated inclination angle of the radial shaft
Stiffness (ST), angular displacement (AD) and interfragmentary displacement (ID) in the three study groups (GR) for volar (V) and double (D) plating (PL) under neutral (N) loading condition, flexion (F) and extension (E), in terms of mean and standard deviation. P-value (P) indicates statistical differences between volar and double plating for the respective outcome. Bold values indicate significant differences
| GR | PL | ST [N/mm] | AD [°] | ID [mm] | ||||
|---|---|---|---|---|---|---|---|---|
| N | F | E | N | F | E | N | ||
| 187.5 (48.4) | 51.4 (14.2) | 41.5 (11.3) | 0.38 (0.18) | 0.75 (0.26) | 0.88 (0.32) | 0.19 (0.07) | ||
| 197.5 (51.6) | 55.6 (15.9) | 48.9 (15.4) | 0.28 (0.10) | 0.67 (0.19) | 0.72 (0.24) | 0.17 (0.06) | ||
| 158.5 (44.4) | 41.5 (13.7) | 35.7 (9.9) | 1.23 0.35) | 1.95 (0.53) | 2.42 (0.75) | 0.22 (0.10) | ||
| 196.5 (50.5) | 54.4 (15.3) | 47.5 (14.9) | 0.31 (0.13) | 0.71 (0.22) | 0.76 (0.27) | 0.20 (0.08) | ||
| 155.9 (44.2) | 40.7 (13.1) | 34.7 (9.2) | 1.32 (0.47) | 2.11 (0.62) | 2.66 (0.79) | 0.23 (0.10) | ||
| 189.3 (49.7) | 53.8 (14.9) | 46.9 (13.5) | 0.34 (0.15) | 0.74 (0.23) | 0.82 (0.29) | 0.22 (0.09) | ||