Literature DB >> 12563644

Comparison of different distal radius dorsal and volar fracture fixation plates: a biomechanical study.

Denju Osada1, Steven F Viegas, Munir A Shah, Randal P Morris, Rita M Patterson.   

Abstract

PURPOSE: To compare the biomechanical properties of 6 dorsal and volar fracture fixation plate designs in a cadaver model.
METHOD: Six different plating techniques were used on surgically simulated, unstable, extra-articular distal radius fractures in fresh-frozen cadavers. Specimens were tested to failure in axial compression with the Materials Testing System machine, and were analyzed with a motion analysis system. The 6 different fixation systems studied included an AO stainless steel Pi plate (group 1), an AO titanium Pi plate (group 2), a Forte plate (group 3), a dorsally placed Symmetry plate (group 4), a volarly placed Symmetry plate (group 5), and a volarly placed SCS/V plate (group 6).
RESULTS: All dorsal plates (groups 1, 2, 3, 4) failed in apex dorsal angulation and all volar plates (groups 5, 6) failed in apex volar angulation. No group developed an average angular deformity greater than 5 degrees with a load of 100 N, which compares with the physiologic loads expected with active wrist motion. Only the volarly placed SCS/V plated specimens (group 6) resisted deformation of 5 degrees or more at loads up to 250 N, which compares with the physiologic loads expected with active finger motion, and was significantly stronger and more rigid than the other 5 plate groups.
CONCLUSIONS: The SCS/V plate fixation system is the most rigid of the systems tested and may offer adequate stability for the treatment of the distal radius fracture in which the anterior and/or posterior metaphyseal cortex is comminuted severely.

Entities:  

Mesh:

Year:  2003        PMID: 12563644     DOI: 10.1053/jhsu.2003.50016

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  31 in total

1.  Comparison of external fixation, locking and non-locking palmar plating for unstable distal radius fractures in the elderly.

Authors:  N Schmelzer-Schmied; P Wieloch; A K Martini; W Daecke
Journal:  Int Orthop       Date:  2008-01-12       Impact factor: 3.075

2.  Complex fractures of the distal radius treated with angular stability plates.

Authors:  Marco Frattini; Giovanni Soncini; Maurizio Corradi; Bruno Panno; Silvio Tocco; Francesco Pogliacomi
Journal:  Chir Organi Mov       Date:  2009-07-21

3.  Volar locking plate fixation of distal radius fractures: use of an intra-operative 'carpal shoot through' view to identify dorsal compartment and distal radioulnar joint screw penetration.

Authors:  Daniel Marsland; Chris M Hobbs; Philip S Sauvé
Journal:  Hand (N Y)       Date:  2014-12

4.  Early Rehabilitation of Distal Radius Fractures Stabilized by Volar Locking Plate: A Prospective Randomized Pilot Study.

Authors:  Stefan Quadlbauer; Christoph Pezzei; Josef Jurkowitsch; Brigitta Kolmayr; Tina Keuchel; Daniel Simon; Thomas Hausner; Martin Leixnering
Journal:  J Wrist Surg       Date:  2016-08-05

Review 5.  MEN 2A-related cutaneous lichen amyloidosis: report of three kindred and systematic literature review of clinical, biochemical and molecular characteristics.

Authors:  Jessica Oliboni Scapineli; Lucieli Ceolin; Márcia Khaled Puñales; José Miguel Dora; Ana Luiza Maia
Journal:  Fam Cancer       Date:  2016-10       Impact factor: 2.375

6.  The treatment of distal radius articular fractures of C1-C2 type with DVR plate: analysis of 40 cases.

Authors:  L Tarallo; R Adani; R Mugnai; F Catani
Journal:  Musculoskelet Surg       Date:  2011-05-26

7.  Unstable Distal Radius Fractures Treated by Volar Locking Anatomical Plates.

Authors:  Anto Jose; Shishir Murugharaj Suranigi; Pascal Noel Deniese; Abey Thomas Babu; Kanagasabai Rengasamy; Syed Najimudeen
Journal:  J Clin Diagn Res       Date:  2017-01-01

8.  Biomechanical comparison of different volar fracture fixation plates for distal radius fractures.

Authors:  Kareem Sobky; Todd Baldini; Kenneth Thomas; Joel Bach; Allison Williams; Jennifer Moriatis Wolf
Journal:  Hand (N Y)       Date:  2007-09-07

9.  Biomechanical evaluation of volar locking plates for distal radius fractures.

Authors:  Scott M Levin; Cory O Nelson; Jonathan D Botts; Glenn A Teplitz; Yong Kwon; Fred Serra-Hsu
Journal:  Hand (N Y)       Date:  2007-08-07

10.  Radiographic outcomes of volar locked plating for distal radius fractures.

Authors:  Megan E Mignemi; Ian R Byram; Carmen C Wolfe; Kang-Hsien Fan; Elizabeth A Koehler; John J Block; Martin I Jordanov; Jeffry T Watson; Douglas R Weikert; Donald H Lee
Journal:  J Hand Surg Am       Date:  2012-12-04       Impact factor: 2.230

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.