| Literature DB >> 34977099 |
Kazutaka Kamiya1, Kimiya Shimizu2, Masahide Takahashi2, Wakako Ando3, Hideki Hayakawa3, Nobuyuki Shoji3.
Abstract
Purpose: To assess the 8-year clinical outcomes of implantation of an implantable collamer lens (ICL) with a central port (KS-Aquaport; EVO-ICL) for moderate to high myopia and myopic astigmatism.Entities:
Keywords: EVO-ICL; efficacy; long-term prognosis; myopia; phakic IOL; predictability; safety; stability
Year: 2021 PMID: 34977099 PMCID: PMC8716586 DOI: 10.3389/fmed.2021.799078
Source DB: PubMed Journal: Front Med (Lausanne) ISSN: 2296-858X
Preoperative patient demographics of the study population.
|
|
|
|---|---|
| Number of eyes | 177 |
| Age | 35.9 ± 7.9 years (21.2 to 50.6 years) |
| Male: Female | 79: 98 |
| Log UDVA | 1.32 ± 0.31 (0.72 to 1.92) |
| LogMAR CDVA | −0.15 ± 0.09 (−0.34 to 0.03) |
| Manifest spherical equivalent | −7.99 ± 3.33 D (−14.51 to −1.46 D) |
| Manifest cylinder | −1.13± 1.20 D (−3.48 to 1.22 D) |
| Intraocular pressure | 13.0 ± 2.5 mmHg (8.0 to 17.9 mmHg) |
| Mean keratometric readings | 43.51 ± 1.83 D (39.92 to 47.11 D) |
| Anterior chamber depth | 3.11 ± 0.29 mm (2.55 to 3.68 mm) |
| White-to-white distance | 11.6 ± 0.4 mm (10.8 to 12.3 mm) |
| Central corneal thickness | 528.7 ± 33.8 μm (462.4 to 594.9 μm) |
| Axial length | 26.97 ± 1.64 mm (23.76 to 30.18 mm) |
| Toric: Non-toric | 76: 101 |
| ICL size 12.1: 12.6: 13.2: 13.7 | 29: 78: 66: 4 |
SD, standard deviation; CI, confidence interval; logMAR, logarithm of the minimal angle of resolution; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; D, diopter; ICL, implantable collamer lens.
Figure 1Changes in corrected distance visual acuity (CDVA) after hole implantable collamer lens (EVO-ICL) implantation.
Figure 2Changes in uncorrected distance visual acuity (UDVA) after hole implantable collamer lens (EVO-ICL) implantation.
Figure 3Scatter plot of attempted vs. achieved correction (spherical equivalent) after hole implantable collamer lens (EVO-ICL) implantation.
Figure 4Time course of manifest spherical equivalent (MSE) after hole implantable collamer lens (EVO-ICL) implantation.
Figure 5Time course of intraocular pressure (IOP) after hole implantable collamer lens (EVO-ICL) implantation.
Figure 6Distribution of eyes according to the vault at 4, 6, and 8 years after hole implantable collamer lens (EVO-ICL) implantation.
Postoperative complications and secondary surgical interventions after hole implantable collamer lens (EVO-ICL) implantation.
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|
|
| |||||||
| Case 1 | 29 | Male | −6.00 | Non–toric | 1 week | Excessive high vault | ICL rotation (horizontal to vertical plane) |
| Case 2 | 27 | Female | −5.63 | Toric | 1 week | ICL rotation | ICL repositioning |
| Case 3 | 53 | Female | −6.88 | Non-toric | 2 years | Anterior subcapsular cataract | ICL extraction and cataract surgery |
| Case 4 | 39 | Female | −15.5 | Non-toric | 4 years | Anterior subcapsular cataract | ICL extraction and cataract surgery |
| Case 5 | 30 | Male | −10.25 | Non-toric | 8 years | Myopic regression | Photorefractive keratectomy |
| 30 | Male | −11.75 | Non-toric | 8 years | Myopic regression | Photorefractive keratectomy | |
| Case 6 | 35 | Male | −17.63 | Toric | 8 years | Anterior subcapsular cataract | ICL size change (1 size up) |
|
| |||||||
| Case 7 | 44 | Male | −5.13 | Non-toric | 1 years | Retinal tear | Photocoagulation |
| Case 8 | 43 | Male | −6.63 | Toric | 4 years | Epiretinal membrane | Observation |
MSE, manifest spherical equivalent; D, Diopter; ICL, implantable collamer lens.
Summary for long-term outcomes of conventional implantable collamer lens implantation (spanning 10 years or more) and hole implantable collamer lens implantation (5 years or more) for myopia.
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|
| Moya et al. ( | 2015 | ICL V3, V4 (non-hole) | 12 | 110 | 30.69 ± 5.59 | −16.90 ± 4.26 | 0.49 ± 0.25 | 0.22 ± 0.25 | 34.3% | 13.88% |
| Guber et al. ( | 2016 | ICL V4 (non-hole) | 10 | 75 | 38.8 ± 9.2 | −11.4 ± 2.9 | 0.6, decimal | 1.0, decimal | 65.7% | 54.8% |
| Kocová et al. ( | 2017 | ICL V4 (non-hole) | 10 | 40 | 28.28 ± 6.25 | −11.0 ± 4.45 | 1.0, decimal | 1.18, decimal | 75.0% | 12.5% |
| Nakamura et al. ( | 2019 | ICL V4 (non-hole) | 10 | 70 | 36.2 ± 7.7 | −9.97 ± 2.29 | −0.01 ± 0.24 | −0.18 ± 0.07 | 87.1% | 10.5% |
| Choi et al. ( | 2019 | ICL V4 (non-hole) | 10 | 110 | 30.3 ± 8.3 | −12.01 ± 3.70 | 0.13 ± 0.20 | 0.03 ± 0.07 | N.A. | 12.1% |
| Shimizu et al. ( | 2016 | ICL V4c (hole) | 5 | 32 | 31.9 ± 7.5 | −7.54 ± 2.40 | −0.17 ± 0.14 | −0.24 ± 0.08 | 100% | 0% |
| Alfonso et al. ( | 2019 | ICL V4c (hole) | 5 | 147 | 31.24 ± 5.4 | −9.20 ± 3.02 | 0.02 ± 0.08 | 0.02 ± 0.09 | 90.1% | 0% |
| Fernández-Vega-Cueto et al. ( | 2021 | ICL V4c (hole) | 7 | 84 | 31.04 ± 4.89 | −9.35 ± 2.90 | 0.17 ± 0.23 | 0.02 ± 0.08 | 100% | 0% |
| Current | 2021 | ICL V4c, V5 (hole) | 8 | 177 | 35.9 ± 7.9 | −7.99 ± 3.33 | −0.07 ± 0.17 | −0.20 ± 0.09 | 93% | 1.7% |
ICL, implantable collamer lens; MSE, manifest spherical equivalent; D, diopter; logMAR, logarithm of the minimal angle of resolution; UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity.