| Literature DB >> 34974645 |
Ji Won Bak1, Yeonji Noh1, Juyoun Kim2, Byeongmun Hwang1, Seongsik Kang1, Heejeong Son1, Minsoo Kim1.
Abstract
BACKGROUND: The GlideScope® videolaryngoscope (GVL) is widely used in patients with difficult airways and provides a good glottic view. However, the acute angle of the blade can make insertion and advancement of an endotracheal tube (ETT) more difficult than direct laryngoscopy, and the use of a stylet is recommended. This randomized controlled trial compared Parker Flex-It™ stylet (PFS) with GlideRite® rigid stylet (GRS) to facilitate intubation with the GVL in simulated difficult intubations.Entities:
Keywords: Airway management; General anesthesia; Intubation; Laryngoscopes
Year: 2021 PMID: 34974645 PMCID: PMC8841254 DOI: 10.17085/apm.21095
Source DB: PubMed Journal: Anesth Pain Med (Seoul) ISSN: 1975-5171
Fig. 1.The GlideRite® rigid stylet (GRS, Verathon, USA), a dedicated stylet provided by the manufacturer.
Fig. 2.The Parker Flex-It™ stylet (PFS, Parker Medical, USA), shown individually and inserted into an endotracheal tube.
Fig. 3.CONSORT flow chart for the study patients. GRS group: GlideRite® rigid stylet (Verathon, USA) group, PFS group: Parker-Flex-It™ stylet (Parker Medical, USA) group.
Patient Demographics and Airway Assessment Data
| Characteristic | GRS group (n = 27) | PFS group (n = 27) |
|---|---|---|
| Age (yr) | 47 (37, 54) | 51 (44, 56) |
| Sex (M/F) | 11/16 | 11/16 |
| BMI (kg/m2) | 25.1 ± 3.3 | 24.0 ± 2.4 |
| ASA (I/II) | 17/10 | 17/10 |
| Thyromental distance (cm) | 9.0 (8.5, 10.0) | 9.5 (8.5, 10.0) |
| Neck circumference (cm) | 38.0 ± 4.0 | 37.1 ± 2.8 |
| Mouth opening without collar (cm) | 4.8 (4.3, 5.0) | 4.6 (3.9, 4.8) |
| Mouth opening with collar (cm) | 1.9 ± 0.4 | 1.9 ± 0.4 |
| Mallampati class (I/II) | 10/17 | 12/15 |
| Cormack-Lehane grade (I/II/III/IV) | 10/14/3/0 | 10/17/0/0 |
| Cormack-Lehane grade with collar by a Macintosh laryngoscope (I/II/III/IV) | 0/0/3/24 | 0/0/2/25 |
| Cormack-Lehane grade with collar by the GVL (I/II/III/IV) | 19/6/2/0 | 22/4/1/0 |
Values are presented as median (1Q, 3Q), number of patients, or mean ± SD. GRS: GlideRite® rigid stylet (Verathon, USA), PFS: Parker Flex-It™ stylet (Parker Medical, USA), BMI: body mass index, ASA: American Society of Anesthesiologists physical status classification, GVL: GlideScope® videolaryngoscope.
Intubation Data of 54 Patients Compared Using Either the GRS or PFS
| Intubation data | GRS group (n = 27) | PFS group (n = 27) | MD (95% CI) | OR (95% CI) | P value |
|---|---|---|---|---|---|
| TIT (s) | 45.9 (41.8, 52.8) | 49.5 (46.2, 64.5) | –7.5 (–16.1 to 1.1) | 0.108 | |
| EIT (s) | 21.2 (18.3, 26.3) | 23.2 (17.0, 40.3) | –7.4 (–15.1 to 0.2) | 0.257 | |
| EAT (s) | 26.7 (22.7, 30.0) | 26.2 (22.7, 30.6) | –0.1 (–4.1 to 4.0) | 0.863 | |
| Ease of intubation (100-mm VAS) | 40.8 ± 12.7 | 52.1 ± 18.3 | –11.3 (–19.9 to –2.8) | 0.011[ | |
| Intubation attempts 1/2/3/fail | 25/2/0/0 | 20/6/1/0 | 0.142 | ||
| Success rate for the first attempt (%) | 92.6 | 74.1 | 0.229 (0.043 to 1.224) | 0.085 | |
| ETT advanced from stylet by assistant | 11 (40.7) | 1 (3.7) | 0.056 (0.007 to 0.475) | 0.008[ | |
| Laryngeal manipulation | 0 (0) | 1 (3.7) | 1.000 |
Values are presented as median (1Q, 3Q), mean ± SD, or number of patients (%). GRS: GlideRite® rigid stylet (Verathon, USA), PFS: Parker Flex-It™ stylet (Parker Medical, USA), MD: mean difference, 95% CI: 95% confidence interval, OR: odds ratio, TIT: total intubation time, EIT: endotracheal tube insertion time, EAT: endotracheal tube advancement time, VAS: visual analog scale, ETT: endotracheal tube.
Significantly different between the groups (P < 0.05).
Incidence of Complications
| Parameters | GRS group (n = 27) | PFS group (n = 27) | P value |
|---|---|---|---|
| Sore throat at 1 h | 12 (44.4) | 15 (55.6) | 0.587 |
| Sore throat at 24 h | 3 (11.1) | 7 (25.9) | 0.293 |
| Hoarseness at 1 h | 6 (22.2) | 5 (18.5) | 1.000 |
| Hoarseness at 24 h | 1 (3.7) | 2 (7.4) | 1.000 |
| Dysphagia at 1 h | 3 (11.1) | 2 (7.4) | 1.000 |
| Dysphagia at 24 h | 1 (3.7) | 0 (0) | 1.000 |
| Hypoxemia (SpO2) < 90% | 0 (0) | 0 (0) | |
| Dental damage | 0 (0) | 0 (0) | |
| Oropharyngeal trauma | 8 (29.6) | 4 (14.8) | 0.327 |
| Blood on the tube | 5 (18.5) | 3 (11.1) | 0.704 |
Values are presented as number of patients (%). GRS: GlideRite® rigid stylet (Verathon, USA), PFS: Parker Flex-It™ stylet (Parker Medical, USA), SpO2: peripheral oxygen saturation.