Literature DB >> 19059923

Comparison of the Glidescope, the Pentax AWS, and the Truview EVO2 with the Macintosh laryngoscope in experienced anaesthetists: a manikin study.

M A Malik1, C O'Donoghue, J Carney, C H Maharaj, B H Harte, J G Laffey.   

Abstract

BACKGROUND: The Pentax Airwayscope, the Glidescope, and the Truview EVO2 constitute three novel laryngoscopes that facilitate visualization of the vocal cords without alignment of the oral, pharyngeal, and tracheal axes. We compared these devices with the Macintosh laryngoscope in a simulated easy and difficult laryngoscopy.
METHODS: Thirty-five experienced anaesthetists were allowed up to three attempts to intubate in each of four laryngoscopy scenarios in a Laerdal SimMan manikin. The time required to perform tracheal intubation, the success rate, number of intubation attempts and of optimization manoeuvres, and the severity of dental compression were recorded.
RESULTS: In the simulated easy laryngoscopy scenarios, there was no difference between the study devices and the Macintosh in success of tracheal intubation. In more difficult tracheal intubation scenarios, the Glidescope and Pentax AWS, and to a lesser extent the Truview EVO2 laryngoscope demonstrated advantages over the Macintosh laryngoscope including a better view of the glottis, greater success of tracheal intubation, and ease of device use. The Pentax AWS was more successful in achieving tracheal intubation, required less time to successfully perform tracheal intubation, caused less dental trauma, and was considered by the anaesthetists to be easier to use.
CONCLUSIONS: The Pentax AWS laryngoscope demonstrated more advantages over the Macintosh laryngoscope than either the Truview EVO2 or the Glidescope laryngoscope, when used by experienced anaesthetists in difficult tracheal intubation scenarios.

Entities:  

Mesh:

Year:  2009        PMID: 19059923     DOI: 10.1093/bja/aen342

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  34 in total

1.  A comparison of video laryngoscopy to direct laryngoscopy for the emergency intubation of trauma patients.

Authors:  Maria Michailidou; Terence O'Keeffe; Jarrod M Mosier; Randall S Friese; Bellal Joseph; Peter Rhee; John C Sakles
Journal:  World J Surg       Date:  2015-03       Impact factor: 3.352

2.  A Comparison of the Effects of Different Types of Laryngoscope on the Cervical Motions: Randomized Clinical Trial.

Authors:  Alkin Çolak; Elif Çopuroğlu; Ali Yılmaz; Sevtap Hekimoğlu Şahin; Nesrin Turan
Journal:  Balkan Med J       Date:  2015-04-01       Impact factor: 2.021

3.  Comparison of Macintosh, McCoy and C-MAC D-Blade video laryngoscope intubation by prehospital emergency health workers: a simulation study.

Authors:  Ahmet Yildirim; Hasan A Kiraz; İbrahim Ağaoğlu; Okhan Akdur
Journal:  Intern Emerg Med       Date:  2016-03-21       Impact factor: 3.397

4.  Recent advance in patient monitoring.

Authors:  Tomoki Nishiyama
Journal:  Korean J Anesthesiol       Date:  2010-09-20

5.  A comparison of the GlideScope video laryngoscope to the C-MAC video laryngoscope for intubation in the emergency department.

Authors:  Jarrod Mosier; Stephen Chiu; Asad E Patanwala; John C Sakles
Journal:  Ann Emerg Med       Date:  2013-01-30       Impact factor: 5.721

6.  Anesthetic management for a patient with severe mento-sternal contracture: difficult airway and scarce venous access -A case report-.

Authors:  Chong-Doo Park; Hye-Kyoung Lee; Ji-Yeon Yim; Im-Hong Kang
Journal:  Korean J Anesthesiol       Date:  2013-01-21

7.  Comparison of the C-MAC D-Blade, Conventional C-MAC, and Macintosh Laryngoscopes in Simulated Easy and Difficult Airways.

Authors:  Alper Kılıçaslan; Ahmet Topal; Atilla Erol; Sema Tuncer Uzun
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-08-01

8.  A comparison of the C-MAC video laryngoscope to the Macintosh direct laryngoscope for intubation in the emergency department.

Authors:  John C Sakles; Jarrod Mosier; Stephen Chiu; Mari Cosentino; Leah Kalin
Journal:  Ann Emerg Med       Date:  2012-05-05       Impact factor: 5.721

9.  A comparison of the TruView EVO2 and macintosh laryngoscope blades.

Authors:  Ayse Cigdem Tutuncu; Guner Kaya; Yusuf Tunali; Fatis Altintas; Ozlem Korkmaz Dilmen
Journal:  Clinics (Sao Paulo)       Date:  2011       Impact factor: 2.365

10.  A systematic review of the role of videolaryngoscopy in successful orotracheal intubation.

Authors:  David W Healy; Oana Maties; David Hovord; Sachin Kheterpal
Journal:  BMC Anesthesiol       Date:  2012-12-14       Impact factor: 2.217

View more