Literature DB >> 34971464

Twenty-four-channel high-impedance glove array for hand and wrist MRI at 3T.

Bei Zhang1,2, Bili Wang1, Justin Ho1, Shota Hodono3, Christopher Burke4, Riccardo Lattanzi1,5, Markus Vester6, Robert Rehner6, Daniel Sodickson1, Ryan Brown1, Martijn Cloos1,3.   

Abstract

PURPOSE: To present a novel 3T 24-channel glove array that enables hand and wrist imaging in varying postures.
METHODS: The glove array consists of an inner glove holding the electronics and an outer glove protecting the components. The inner glove consists of four main structures: palm, fingers, wrist, and a flap that rolls over on top. Each structure was constructed out of three layers: a layer of electrostatic discharge flame-resistant fabric, a layer of scuba neoprene, and a layer of mesh fabric. Lightweight and flexible high impedance coil (HIC) elements were inserted into dedicated tubes sewn into the fabric. Coil elements were deliberately shortened to minimize the matching interface. Siemens Tim 4G technology was used to connect all 24 HIC elements to the scanner with only one plug.
RESULTS: The 24-channel glove array allows large motion of both wrist and hand while maintaining the SNR needed for high-resolution imaging.
CONCLUSION: In this work, a purpose-built 3T glove array that embeds 24 HIC elements is demonstrated for both hand and wrist imaging. The 24-channel glove array allows a great range of motion of both the wrist and hand while maintaining a high SNR and providing good theoretical acceleration performance, thus enabling hand and wrist imaging at different postures to extract kinematic information.
© 2022 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  MRI RF array; dynamic imaging; hand imaging; high impedance coil; wrist imaging

Mesh:

Year:  2021        PMID: 34971464      PMCID: PMC8847333          DOI: 10.1002/mrm.29147

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  33 in total

1.  Resolution enhancement in single-shot imaging using simultaneous acquisition of spatial harmonics (SMASH).

Authors:  M A Griswold; P M Jakob; Q Chen; J W Goldfarb; W J Manning; R R Edelman; D K Sodickson
Journal:  Magn Reson Med       Date:  1999-06       Impact factor: 4.668

2.  Scapholunate kinematics of asymptomatic wrists in comparison with symptomatic contralateral wrists using four-dimensional CT examinations: initial clinical experience.

Authors:  Shadpour Demehri; Nima Hafezi-Nejad; John N Morelli; Uma Thakur; Scott D Lifchez; Kenneth R Means; John Eng; Jaimie T Shores
Journal:  Skeletal Radiol       Date:  2015-12-14       Impact factor: 2.199

3.  Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays.

Authors:  D K Sodickson; W J Manning
Journal:  Magn Reson Med       Date:  1997-10       Impact factor: 4.668

4.  Correction for intensity falloff in surface coil magnetic resonance imaging.

Authors:  W W Brey; P A Narayana
Journal:  Med Phys       Date:  1988 Mar-Apr       Impact factor: 4.071

5.  Cine MRI: a new approach to the diagnosis of scapholunate dissociation.

Authors:  I Langner; S Fischer; A Eisenschenk; S Langner
Journal:  Skeletal Radiol       Date:  2015-03-12       Impact factor: 2.199

6.  Dynamic CT Scan of the Normal Scapholunate Joint in a Clenched Fist and Radial and Ulnar Deviation.

Authors:  Paul M Kelly; John G Hopkins; Andrew J Furey; Daniel S Squire
Journal:  Hand (N Y)       Date:  2017-08-29

7.  Materials and methods for higher performance screen-printed flexible MRI receive coils.

Authors:  Joseph R Corea; P Balthazar Lechene; Michael Lustig; Ana C Arias
Journal:  Magn Reson Med       Date:  2016-09-09       Impact factor: 4.668

8.  Dynamic motion analysis of dart throwers motion visualized through computerized tomography and calculation of the axis of rotation.

Authors:  Y Edirisinghe; J M Troupis; M Patel; J Smith; M Crossett
Journal:  J Hand Surg Eur Vol       Date:  2013-10-25

9.  Conductive Thread-Based Stretchable and Flexible Radiofrequency Coils for Magnetic Resonance Imaging.

Authors:  Jana M Vincent; Joseph V Rispoli
Journal:  IEEE Trans Biomed Eng       Date:  2019-11-28       Impact factor: 4.538

10.  A high-impedance detector-array glove for magnetic resonance imaging of the hand.

Authors:  Bei Zhang; Daniel K Sodickson; Martijn A Cloos
Journal:  Nat Biomed Eng       Date:  2018-05-07       Impact factor: 25.671

View more
  2 in total

1.  Twenty-four-channel high-impedance glove array for hand and wrist MRI at 3T.

Authors:  Bei Zhang; Bili Wang; Justin Ho; Shota Hodono; Christopher Burke; Riccardo Lattanzi; Markus Vester; Robert Rehner; Daniel Sodickson; Ryan Brown; Martijn Cloos
Journal:  Magn Reson Med       Date:  2021-12-31       Impact factor: 4.668

2.  A flexible MRI coil based on a cable conductor and applied to knee imaging.

Authors:  Bili Wang; Syed S Siddiq; Jerzy Walczyk; Mary Bruno; Iman Khodarahmi; Inge M Brinkmann; Robert Rehner; Karthik Lakshmanan; Jan Fritz; Ryan Brown
Journal:  Sci Rep       Date:  2022-09-02       Impact factor: 4.996

  2 in total