Virtual injections using 4D flow MRI with displacement corrections and constrained probabilistic streamlines.

Grant S Roberts1, Michael W Loecher2, Alma Spahic1, Kevin M Johnson1,3, Patrick A Turski3, Laura B Eisenmenger3, Oliver Wieben1,3.   

Abstract

PURPOSE: Streamlines from 4D-flow MRI have been used clinically for intracranial blood-flow tracking. However, deterministic and stochastic errors degrade streamline quality. The purpose of this study is to integrate displacement corrections, probabilistic streamlines, and novel fluid constraints to improve selective blood-flow tracking and emulate "virtual bolus injections."
METHODS: Both displacement artifacts (deterministic) and velocity noise (stochastic) inherently occur during phase-contrast MRI acquisitions. Here, two displacement correction methods, single-step and iterative, were tested in silico with simulated displacements and were compared with ground-truth velocity fields. Next, the effects of combining displacement corrections and constrained probabilistic streamlines were performed in 10 healthy volunteers using time-averaged 4D-flow data. Measures of streamline length and depth into vasculature were then compared with streamlines generated with no corrections and displacement correction alone using one-way repeated-measures analysis of variance and Friedman's tests. Finally, virtual injections with improved streamlines were generated for three intracranial pathology cases.
RESULTS: Iterative displacement correction outperformed the single-step method in silico. In volunteers, the combination of displacement corrections and constrained probabilistic streamlines allowed for significant improvements in streamline length and increased the number of streamlines entering the circle of Willis relative to streamlines with no corrections and displacement correction alone. In the pathology cases, virtual injections with improved streamlines were qualitatively similar to dynamic arterial spin labeling images and allowed for forward/reverse selective flow tracking to characterize cerebrovascular malformations.
CONCLUSION: Virtual injections with improved streamlines from 4D-flow MRI allow for flexible, robust, intracranial flow tracking.
© 2021 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  4D flow MRI; angiography; blood flow tracking; displacement correction; intracranial; probabilistic streamlines

Mesh:

Substances:

Year:  2021        PMID: 34971458      PMCID: PMC8884720          DOI: 10.1002/mrm.29134

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  48 in total

1.  Correction for acceleration-induced displacement artifacts in phase contrast imaging.

Authors:  P Thunberg; L Wigström; B Wranne; J Engvall; M Karlsson
Journal:  Magn Reson Med       Date:  2000-05       Impact factor: 4.668

2.  Quantification of intravoxel velocity standard deviation and turbulence intensity by generalizing phase-contrast MRI.

Authors:  Petter Dyverfeldt; Andreas Sigfridsson; John-Peder Escobar Kvitting; Tino Ebbers
Journal:  Magn Reson Med       Date:  2006-10       Impact factor: 4.668

3.  4D radial contrast-enhanced MR angiography with sliding subtraction.

Authors:  Ty A Cashen; Hyun Jeong; Maulin K Shah; Hem M Bhatt; Wanyong Shin; James C Carr; Matthew T Walker; H Hunt Batjer; Timothy J Carroll
Journal:  Magn Reson Med       Date:  2007-11       Impact factor: 4.668

4.  Combined analysis of spatial and velocity displacement artifacts in phase contrast measurements of complex flows.

Authors:  D A Steinman; C R Ethier; B K Rutt
Journal:  J Magn Reson Imaging       Date:  1997 Mar-Apr       Impact factor: 4.813

5.  Analysis of noise in phase contrast MR imaging.

Authors:  A H Andersen; J E Kirsch
Journal:  Med Phys       Date:  1996-06       Impact factor: 4.071

6.  Accelerating 4D flow MRI by exploiting vector field divergence regularization.

Authors:  Claudio Santelli; Michael Loecher; Julia Busch; Oliver Wieben; Tobias Schaeffter; Sebastian Kozerke
Journal:  Magn Reson Med       Date:  2015-02-13       Impact factor: 4.668

7.  Understanding acceleration-induced displacement artifacts in phase-contrast MR velocity measurements.

Authors:  R Frayne; B K Rutt
Journal:  J Magn Reson Imaging       Date:  1995 Mar-Apr       Impact factor: 4.813

8.  Time-of-Arrival Parametric Maps and Virtual Bolus Images Derived From Contrast-Enhanced Time-Resolved Radial Magnetic Resonance Angiography Improve the Display of Brain Arteriovenous Malformation Vascular Anatomy.

Authors:  Tilman Schubert; Yijing Wu; Kevin M Johnson; Oliver Wieben; Jane Maksimovic; Charles Mistretta; Patrick Turski
Journal:  Invest Radiol       Date:  2016-11       Impact factor: 6.016

9.  Time-resolved vessel-selective digital subtraction MR angiography of the cerebral vasculature with arterial spin labeling.

Authors:  Philip M Robson; Weiying Dai; Ajit Shankaranarayanan; Neil M Rofsky; David C Alsop
Journal:  Radiology       Date:  2010-11       Impact factor: 11.105

10.  Robust 4D flow denoising using divergence-free wavelet transform.

Authors:  Frank Ong; Martin Uecker; Umar Tariq; Albert Hsiao; Marcus T Alley; Shreyas S Vasanawala; Michael Lustig
Journal:  Magn Reson Med       Date:  2014-02-18       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.