Combined analysis of spatial and velocity displacement artifacts in phase contrast measurements of complex flows.

D A Steinman1, C R Ethier, B K Rutt.   

Abstract

MR phase contrast (PC) velocity imaging is a promising tool for quantifying blood flow velocity in vivo. PC velocity imaging is, however, susceptible to artifacts that result from the displacement of spins during the finite duration pulse sequences. Such displacement artifacts can lead to errors in velocity measurements, especially in the presence of oblique and accelerating flows, which are common throughout the cardiovascular system. By tracking particles (representing spins) through a computed velocity field, and assuming that spatial and velocity encodings occur at discrete times during the pulse sequence, we simulate the separate and combined effects of oblique and acceleration artifacts on PC velocity images. We demonstrate, both by simulation and MR measurement, the errors associated with such artifacts in PC velocity measurements in a representative flow geometry. Using example particle trajectories, we provide a fluid dynamic basis for characteristic phase-velocity image distortions that can arise when imaging complex, physiologically relevant flows.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9090588     DOI: 10.1002/jmri.1880070214

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  8 in total

1.  In vivo validation of numerical prediction for turbulence intensity in an aortic coarctation.

Authors:  Amirhossein Arzani; Petter Dyverfeldt; Tino Ebbers; Shawn C Shadden
Journal:  Ann Biomed Eng       Date:  2011-10-21       Impact factor: 3.934

2.  A methodology to detect abnormal relative wall shear stress on the full surface of the thoracic aorta using four-dimensional flow MRI.

Authors:  Pim van Ooij; Wouter V Potters; Aart J Nederveen; Bradley D Allen; Jeremy Collins; James Carr; S Chris Malaisrie; Michael Markl; Alex J Barker
Journal:  Magn Reson Med       Date:  2014-04-18       Impact factor: 4.668

3.  Volumetric velocity measurements in restricted geometries using spiral sampling: a phantom study.

Authors:  Anders Nilsson; Johan Revstedt; Einar Heiberg; Freddy Ståhlberg; Karin Markenroth Bloch
Journal:  MAGMA       Date:  2014-05-18       Impact factor: 2.310

4.  Virtual injections using 4D flow MRI with displacement corrections and constrained probabilistic streamlines.

Authors:  Grant S Roberts; Michael W Loecher; Alma Spahic; Kevin M Johnson; Patrick A Turski; Laura B Eisenmenger; Oliver Wieben
Journal:  Magn Reson Med       Date:  2021-12-31       Impact factor: 4.668

5.  Four-Dimensional Flow Magnetic Resonance Imaging for Assessment of Velocity Magnitudes and Flow Patterns in The Human Carotid Artery Bifurcation: Comparison with Computational Fluid Dynamics.

Authors:  Minh Tri Ngo; Chul In Kim; Jinmu Jung; Gyung Ho Chung; Dong Hwan Lee; Hyo Sung Kwak
Journal:  Diagnostics (Basel)       Date:  2019-12-13

6.  Numerical simulation of time-resolved 3D phase-contrast magnetic resonance imaging.

Authors:  Thomas Puiseux; Anou Sewonu; Ramiro Moreno; Simon Mendez; Franck Nicoud
Journal:  PLoS One       Date:  2021-03-26       Impact factor: 3.240

7.  Synthesis of patient-specific multipoint 4D flow MRI data of turbulent aortic flow downstream of stenotic valves.

Authors:  Pietro Dirix; Stefano Buoso; Eva S Peper; Sebastian Kozerke
Journal:  Sci Rep       Date:  2022-09-26       Impact factor: 4.996

8.  Quantifying coronary sinus flow and global LV perfusion at 3T.

Authors:  Karin Markenroth Bloch; Marcus Carlsson; Håkan Arheden; Freddy Ståhlberg
Journal:  BMC Med Imaging       Date:  2009-06-11       Impact factor: 1.930