| Literature DB >> 34967585 |
Ponnusamy Natesan1,2, Senthil Manikandan Palaniappan3, Muthuvinayagam M2, K M Ganesh3, Malla Bhaskara Rao1.
Abstract
BACKGROUND: Measurement of Collimator helmet factors (CHF) is an important quality assurance procedure to be performed on Leksell Gamma Knife unit at regular interval to make sure that the interchangeable collimator helmet fit into the source channels without any positional inaccuracy which leads to major treatment error. The primary aim of this study is to measure the CHFs for Elekta Leksell Gamma knife 4C helmets using GafChromic EBT3 film and Image J software.Entities:
Keywords: GAF chromic EBT3 film Dosimetry; Gamma Knife output factor; Gamma knife Collimator factor; ImageJ Software
Mesh:
Year: 2021 PMID: 34967585 PMCID: PMC9080369 DOI: 10.31557/APJCP.2021.22.12.4031
Source DB: PubMed Journal: Asian Pac J Cancer Prev ISSN: 1513-7368
Figure 14mm Helmet of Gamma Knife Treatment Unit
Figure 2Structure of GAF Chromic EBT3 Dosimetry Film
Figure 3ABS Phantom Settings in Leksell Gamma Knife 4C Treatment Unit
Figure 4Elekta ABS 160mm Phantom with GAFChromic EBT3 Film Inserts
Figure 5Calibration curve of GAFChromic EBT3 Film with 14mm Helmet
Elekta Reference, Measured and Published Values of Gamma Knife Helmet Factors for 4, 8, 14 and 18mm Collimator Helmets
| S.no | Helmet Size | Reference helmet factor by Elekta | Measured helmet factor ± std deviation | Mean values of helmet factors ± std deviation (Mack et al) | Percentage of deviation between Measured value and Elekta reference values | Percentage of deviation between Mean values and Elekta reference values |
|---|---|---|---|---|---|---|
| 1 | 4mm | 0.881 | 0.896 ±0.020 | 0.865±0.020 | 1.70% | -1.80% |
| 2 | 8mm | 0.956 | 0.958±0.016 | 0.950 ±0.005 | 0.21% | -0.62% |
| 3 | 14mm | 0.985 | 0.986±0.013 | 0.983±0.008 | 0.10% | 0.20% |
| 4 | 18mm | 1 | 1 | 1 | - | - |
Figure 6Comparison of Collimator Helmet Factors of Elekta, Measured and Published Values