Laryssa Kurjewicz1, Anita Berndt.
Abstract
The relative dose rate for the different Gamma Knife helmets (4, 8, 14, and 18 mm) is characterized by their respective helmet factors. Since the plateau of the dose profile for the 4 mm helmet is at most 1 mm wide, detector choices are limited. Traditionally helmet factors have been measured using 1 x 1 x 1 mm3 thermoluminescent dosimeters (TLDs). However, these are time-consuming, cumbersome measurements. This article investigates the use of metal-oxide-semiconductor field effect transistors (MOSFETs) (active area of 0.2 x 0.2 mm2) as a more accurate and convenient dosimeter. Their suitability for these measurements was confirmed by basic characterization measurements. Helmet factors were measured using both MOSFETs and the established TLD approach. A custom MOSFET cassette was designed in analogy to the Elekta TLD cassette (Elekta Instruments AB) for use with the Elekta dosimetry sphere. Although both dosimeters provided values within 3% of the manufacturer's suggestion, MOSFETs provided superior accuracy and precision, in a fraction of the time required for the TLD measurements. Thus, MOSFETs proved to be a reasonable alternative to TLDs for performing helmet factor measurements.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17441247 DOI: 10.1118/1.2437282