Literature DB >> 28527605

Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit.

Mohsen Najafi1, Ghazale Geraily1, Alireza Shirazi2, Mahbod Esfahani3, Javad Teimouri4.   

Abstract

In recent years, Gafchromic films are used as an advanced instrument for dosimetry systems. The EBT3 films are a new generation of Gafchromic films. Our main interest is to compare the response of the EBT3 films exposed to gamma rays provided by the Theratron 780C as a conventional radiotherapy system and the Leksell Gamma Knife as a stereotactic radiotherapy system (SRS). Both systems use Cobalt-60 sources, thus using the same energy. However, other factors such as source-to-axis distance, number of sources, dose rate, direction of irradiation, shape of phantom, the field shape of radiation, and different scatter contribution may influence the calibration curve. Calibration curves for the 2 systems were measured and plotted for doses ranging from 0 to 40 Gy at the red and green channels. The best fitting curve was obtained with the Levenberg-Marquardt algorithm. Also, the component of dose uncertainty was obtained for any calibration curve. With the best fitting curve for the EBT3 films, we can use the calibration curve to measure the absolute dose in radiation therapy. Although there is a small deviation between the 2 curves, the p-value at any channel shows no significant difference between the 2 calibration curves. Therefore, the calibration curve for each system can be the same because of minor differences. The results show that with the best fitting curve from measured data, while considering the measurement uncertainties related to them, the EBT3 calibration curve can be used to measure the unknown dose both in SRS and in conventional radiotherapy.
Copyright © 2017. Published by Elsevier Inc.

Entities:  

Keywords:  Calibration; Cobalt-60; EBT3 film; Gamma Knife

Mesh:

Year:  2017        PMID: 28527605     DOI: 10.1016/j.meddos.2017.01.003

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  7 in total

1.  Evaluation of Dose Distribution in Optimized Stanford Total Skin Electron Therapy (TSET) Technique in Rando Anthropomorphic Phantom using EBT3 Gafchromatic Films.

Authors:  Shaghayegh Fahimi Monzari; Ghazale Geraily; Mehdi Aghili; Heydar Toolee
Journal:  J Biomed Phys Eng       Date:  2021-08-01

2.  Radiation dose verification of an X-ray based blood irradiator using EBT3 radiochromic films calibrated using Gamma Knife machine.

Authors:  Khaled Soliman; Marouf Adili; Abdullah Alrushoud
Journal:  Rep Pract Oncol Radiother       Date:  2019-06-07

3.  Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique.

Authors:  Beong Ik Hur; Seong Jin Jin; Gyeong Rip Kim; Jong Hyeok Kwak; Young Ha Kim; Sang Weon Lee; Soon Ki Sung
Journal:  J Korean Neurosurg Soc       Date:  2021-01-01

4.  Determination of Collimator Helmet Factors for Leksell Gamma Knife 4C Unit Using GAF Chromic EBT3 Film and ImageJ Software.

Authors:  Ponnusamy Natesan; Senthil Manikandan Palaniappan; Muthuvinayagam M; K M Ganesh; Malla Bhaskara Rao
Journal:  Asian Pac J Cancer Prev       Date:  2021-12-01

5.  Development of custom lead shield and strainer for targeted irradiation for mice in the gamma cell chamber.

Authors:  Nurhaslina Hasan; Nur Fatihah Ronny Sham; Muhammad Khalis Abdul Karim; Syed Baharom Syed Ahmad Fuad; Narimah Abdul Hamid Hasani; Effat Omar; Mohammad Johari Ibahim
Journal:  Sci Rep       Date:  2021-07-15       Impact factor: 4.379

6.  Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams.

Authors:  Nashrulhaq Tagiling; Raizulnasuha Ab Rashid; Siti Nur Amirah Azhan; Norhayati Dollah; Moshi Geso; Wan Nordiana Rahman
Journal:  Heliyon       Date:  2018-10-17

7.  A method for time-independent film dosimetry: Can we obtain accurate patient-specific QA results at any time postirradiation?

Authors:  Leon Dunn; Guy Godwin; James Hellyer; Xiaolei Xu
Journal:  J Appl Clin Med Phys       Date:  2022-01-20       Impact factor: 2.102